| Literature DB >> 25969772 |
Wei Wang1, Nan Wang2, Yu Zhang1, Zheng Cai1, Qihe Chen3, Guoqing He3.
Abstract
A convenient and accurate reversed-phase high-performance liquid chromatography (RP-HPLC) method for angiotensin I-converting enzyme inhibitory peptides assay was described in this paper. The mobile phase consisted of 70% A (0.05% TFA and 0.05% triethylamine in water, pH = 2.9-3.3) and 30% B (100% acetonitrile) using an Isogradient program. The flow rate was 0.5 mL/min. The absorb wavelength was 226.5 nm; the column temperature was controlled at 25°C. This method for angiotensin I-converting enzyme inhibitory peptides assay was convenient for the Iso-gradient program. The accuracy of the RT-HPLC method was verified by analyzing ACE inhibitory activity of the hydrolysate peptides of silkworm pupae protein, and the results showed that the RT-HPLC method was available for exploring new source of angiotensin I-converting enzyme inhibitory peptides rapidly and veraciously.Entities:
Year: 2012 PMID: 25969772 PMCID: PMC4403573 DOI: 10.5402/2013/453910
Source DB: PubMed Journal: ISRN Biotechnol ISSN: 2090-9403
Figure 1The ultraviolet absorbance spectra of HHL and HA.
Figure 2The separation result between HHL and HA by the RP-HPLC conditions.
Figure 3The effect of HHL concentration on the conversion rate from HHL to HA.
Recoveries of HA in reaction with different concentrations of captopril (n = 9).
| Captopril | Original/( | Adding HA/( | Total ( | Recovery rate (%) |
|---|---|---|---|---|
| 5 mM | 2.210 ± 0.012 | 20 | 22.143 ± 0.011 | 99.70 |
| 2.5 mM | 3.713 ± 0.009 | 20 | 23.821 ± 0.014 | 100.46 |
| 1 mM | 6.427 ± 0.016 | 20 | 26.441 ± 0.007 | 100.05 |
*All the assayed results were expressed as means ± SEM.