Influence of water and fat heterogeneity on fat-referenced MR thermometry.

Paul Baron1, Roel Deckers1, Job G Bouwman1, Chris J G Bakker1, Martijn de Greef1, Max A Viergever1, Chrit T W Moonen1, Lambertus W Bartels1.   

Abstract

PURPOSE: To investigate the effect of the aqueous and fatty tissue magnetic susceptibility distribution on absolute and relative temperature measurements as obtained directly from the water/fat (w/f) frequency difference.
METHODS: Absolute thermometry was investigated using spherical phantoms filled with pork and margarine, which were scanned in three orthogonal orientations. To evaluate relative fat referencing, multigradient echo scans were acquired before and after heating pork tissue via high-intensity focused ultrasound (HIFU). Simulations were performed to estimate the errors that can be expected in human breast tissue.
RESULTS: The sphere experiment showed susceptibility-related errors of 8.4 °C and 0.2 °C for pork and margarine, respectively. For relative fat referencing measurements, fat showed pronounced phase changes of opposite polarity to aqueous tissue. The apparent mean temperature for a numerical breast model assumed to be 37 °C was 47.2 ± 21.6 °C. Simulations of relative fat referencing for a HIFU sonication (ΔT = 29.7 °C) yielded a maximum temperature error of 6.6 °C compared with 2.5 °C without fat referencing.
CONCLUSION: Variations in the observed frequency difference between water and fat are largely due to variations in the w/f spatial distribution. This effect may lead to considerable errors in absolute MR thermometry. Additionally, fat referencing may exacerbate rather than correct for proton resonance frequency shift-temperature measurement errors.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  MR thermometry; MR-HIFU; absolute thermometry; fat referenced thermometry; magnetic susceptibility

Mesh:

Substances:

Year:  2015        PMID: 25940426     DOI: 10.1002/mrm.25727

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  6 in total

1.  Calibration of methylene-referenced lipid-dissolved xenon frequency for absolute MR temperature measurements.

Authors:  Michael A Antonacci; Le Zhang; Simone Degan; Detlev Erdmann; Rosa T Branca
Journal:  Magn Reson Med       Date:  2018-09-14       Impact factor: 4.668

2.  Correlation distance dependence of the resonance frequency of intermolecular zero quantum coherences and its implication for MR thermometry.

Authors:  Le Zhang; Andrew McCallister; Karl M Koshlap; Rosa Tamara Branca
Journal:  Magn Reson Med       Date:  2017-06-27       Impact factor: 4.668

3.  Accurate MR thermometry by hyperpolarized 129 Xe.

Authors:  Le Zhang; Alex Burant; Andrew McCallister; Victor Zhao; Karl M Koshlap; Simone Degan; Michael Antonacci; Rosa Tamara Branca
Journal:  Magn Reson Med       Date:  2016-10-19       Impact factor: 4.668

4.  Direct detection of brown adipose tissue thermogenesis in UCP1-/- mice by hyperpolarized 129Xe MR thermometry.

Authors:  Michael A Antonacci; Christian McHugh; Michele Kelley; Andrew McCallister; Simone Degan; Rosa T Branca
Journal:  Sci Rep       Date:  2019-10-16       Impact factor: 4.379

5.  Anatomical Phase Extraction (APE) Method: A Novel Method to Correct Detrimental Effects of Tissue-Inhomogeneity in Referenceless MR Thermometry-Preliminary Ex Vivo Investigation.

Authors:  Chien-Feng Judith Huang; Win-Li Lin; San-Chao Hwang; Ching Yao; Hsu Chang; Li-Wei Kuo
Journal:  Comput Math Methods Med       Date:  2021-08-10       Impact factor: 2.238

Review 6.  Magnetic Resonance Imaging Techniques for Brown Adipose Tissue Detection.

Authors:  Mingming Wu; Daniela Junker; Rosa Tamara Branca; Dimitrios C Karampinos
Journal:  Front Endocrinol (Lausanne)       Date:  2020-08-07       Impact factor: 5.555

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.