Air compressor battery duration with mechanical ventilation in a field anesthesia machine.

Dale F Szpisjak1, Anthony A Giberman2.   

Abstract

Compressed air to power field anesthesia machine ventilators may be supplied by air compressor with battery backup. This study determined the battery duration when the compPAC ventilator's air compressor was powered by NiCd battery to ventilate the Vent Aid Training Test Lung modeling high (HC = 0.100 L/cm H2O) and low (LC = 0.020 L/cm H2O) pulmonary compliance. Target tidal volumes (VT) were 500, 750, and 1,000 mL. Respiratory rate = 10 bpm, inspiratory-to-expiratory time ratio = 1:2, and fresh gas flow = 1 L/min air. N = 5 in each group. Control limits were determined from the first 150 minutes of battery power for each run and lower control limit = mean VT - 3SD. Battery depletion occurred when VT was below the lower control limit. Battery duration ranged from 185.8 (±3.2) minutes in the LC-1000 group to 233.3 (±3.6) minutes in the HC-750 group. Battery duration of the LC-1000 group was less than all others (p = 0.027). The differences among the non-LC-1000 groups were not clinically significant. Reprint &
Copyright © 2015 Association of Military Surgeons of the U.S.

Entities:  

Mesh:

Year:  2015        PMID: 25939102     DOI: 10.7205/MILMED-D-14-00461

Source DB:  PubMed          Journal:  Mil Med        ISSN: 0026-4075            Impact factor:   1.437


  1 in total

1.  Evaluation of the Approaches of the Anaesthesiologists on Maintenance of Anaesthesia Machines.

Authors:  Uğur Altınışık; Hatice Betül Altınışık; Tuncer Şimşek; Hasan Ali Kiraz; Tuğba Doğu; Hüseyin Toman; Hasan Şahin; Mesut Erbaş
Journal:  Turk J Anaesthesiol Reanim       Date:  2016-06-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.