Elif Hocaoglu1, Ercan Inci1, Sibel Aydin2, Dilek Hacer Cesme1, Nadir Kalfazade3.
Abstract
OBJECTIVE: The aim of this study was to evaluate the capability and the reliability of diffusion-weighted imaging (DWI) in the changes of kidneys occurring after extracorporeal shock wave lithotripsy (ESWL) treatment for renal stones.Entities:
Mesh:
Year: 2015 PMID: 25928520 PMCID: PMC4752067 DOI: 10.1590/S1677-5538.IBJU.2015.01.19
Source DB: PubMed Journal: Int Braz J Urol ISSN: 1677-5538 Impact factor: 1.541
Figure 156 years old male patient. Apparent diffusion coefficient (ADC) maps of treated and contralateral kidneys before (Figure-1a) and after (Figure-1b) treatment with ESWL. The ADC value is decreased from 2.29x10-3 mm2/sn to 1.92x10-3 mm2/sn in treated kidney. The ADC values of the contralateral kidney are 2.30x10-3 mm2/sn and 2.23x10-3 mm2/sn before and after ESWL treatment, respectively.
Figure 244 years old male patient. Apparent diffusion coefficient (ADC) maps and signal intensities (with b values of 1000 s/mm2) of treated (left) and contralateral kidneys before (Figures 2a and 2b) and after (Figures 2c and 2d) treatment with ESWL. The ADC value decreased from 2.17x10-3 mm2/sn (Figure-2a) to 1.64x10-3 mm2/sn (Figure-2c) in the treated kidney. Signal intensity of the treated kidney with b values of 1000 s/mm2 increased from 27.0 s/mm2 (Figure-2b) to 46.6 s/mm2 (Figure-2d). The ADC values of the contralateral kidney are 2.16x10-3 mm2/sn (Figure-2a) and 2.21x10-3 mm2/sn (Figure-2c) before and after ESWL treatment, respectively. Signal intensities of the contralateral kidney with b values of 1000 s/mm2 are 27.7 s/mm2 (Figure-2b) and 30.7 s/mm2 (Figure-2d) before and after ESWL treatment, respectively.
The RI values of treated kidneys before and after ESWL in certain regions.
| RI | p | ||
|---|---|---|---|
| Pre-ESWL | Post-ESWL | ||
| Mean±SD | Mean±SD | ||
| Upper Pole | 0.58±0.05 | 0.59±0.04 | 0.375 |
| Middle Zone | 0.58±0.05 | 0.59±0.04 | 0.204 |
| Lower Pole | 0.58±0.04 | 0.59±0.04 | 0.307 |
The mean ADC values of ipsilateral kidneys with b-0, b-500 and b-1000 values before and after ESWL treatment.
| Pre-ESWL | Post-ESWL | p | ||
|---|---|---|---|---|
| Mean±SD | Mean±SD | |||
|
| b0 | 244.47±44.99 | 254.03±43.41 | 0.107 |
| b500 | 67.37±7.01 | 72.15±13.98 | 0.075 | |
| b1000 | 26.87±5.84 | 31.18±11.55 | 0.014* | |
| ADC | 2.173±0.112 | 2.099±0.243 | 0.049* | |
|
| b0 | 255.06±43.95 | 266.34±40.70 | 0.006** |
| b500 | 68.47±9.38 | 75.37±13.72 | 0.005** | |
| b1000 | 27.15±5.81 | 33.40±11.60 | 0.001** | |
| ADC | 2.189±0.125 | 2.053±0.244 | 0.004** | |
|
| b0 | 241.12±47.75 | 254.97±46.31 | 0.003** |
| b500 | 65.43±9.28 | 69.72±11.64 | 0.015* | |
| b1000 | 25.56±6.95 | 30.56±11.26 | 0.012* | |
| ADC | 2.201±0.129 | 2.115±0.204 | 0.023* | |
ADC (x10-3 mm2/sn)
*p<0.05; **p<0.01
Figure 3The mean ADC values before and after ESWL treatment in upper pole, middle zone and lower pole of treated kidneys.
Figure 4Signal intensities with b=1000 s/mm2 before and after ESWL treatment in upper pole, middle zone and lower pole of treated kidneys.