Literature DB >> 2592088

Implant failures in patients with proximal fractures of the femur treated with a sliding screw device.

C C Mainds1, R J Newman.   

Abstract

A review of 385 proximal fractures of the femur treated with a telescopic screw system revealed that in nine cases the device had cut out of the head of the femur necessitating its removal. In two cases the head of the femur had been penetrated at the time of insertion. In all but one of the remaining seven cases the screw had been inserted in the superior half of the head and neck of the femur. The overall cut-out rate was 2.25 per cent but was 11.9 per cent when the screw was placed in the superior half of the head. When the screw was placed in the inferior half of the head, the cut-out rate was only 0.29 per cent. There was no correlation between this mechanical failure and age, sex, type of fracture or the presence of metabolic bone disease. These results indicate that if this otherwise effective form of internal fixation is to be used in the treatment of proximal fractures of the femur, care must be taken to avoid placement of the device in the superior half of the head and neck of the femur.

Entities:  

Mesh:

Year:  1989        PMID: 2592088     DOI: 10.1016/0020-1383(89)90151-4

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  14 in total

1.  A new technique for lag screw placement in the dynamic hip screw fixation of intertrochanteric fractures: decreasing radiation time dramatically.

Authors:  Wei-Chao Sheng; Jia-Zhen Li; Sheng-Hua Chen; Shi-Zhen Zhong
Journal:  Int Orthop       Date:  2008-02-12       Impact factor: 3.075

2.  Cephalomedullary fixation for intertrochanteric fractures: an operative technical tip.

Authors:  Lin Jin; Liping Zhang; Zhiyong Hou; Wei Chen; Pengcheng Wang; Yingze Zhang
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-03-22

Review 3.  Is rotation the mode of failure in pertrochanteric fractures fixed with nails? Theoretical approach and illustrative cases.

Authors:  C Kokoroghiannis; D Vasilakos; K Zisis; G Dimitriou; E Pappa; D Evangelopoulos
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-09-20

4.  Avoiding malalignment in proximal femur fractures treated with newer generation cephalomedullary nails: Some technical tips.

Authors:  Arvind Kumar; Samarth Mittal; Aditya Jain; Vivek Trikha
Journal:  J Clin Orthop Trauma       Date:  2018-11-23

5.  [Clinical evaluation of PFNA® and relationship between the tip-apex distance and mechanical failure].

Authors:  M Kraus; G Krischak; K Wiedmann; C Riepl; F Gebhard; J A Jöckel; A Scola
Journal:  Unfallchirurg       Date:  2011-06       Impact factor: 1.000

6.  Risk factors in cutout of sliding hip screw in intertrochanteric fractures: an evaluation of 937 patients.

Authors:  Kuang-Kai Hsueh; Chi-Kuang Fang; Chuan-Mu Chen; Yu-Ping Su; Heng-Fei Wu; Fang-Yao Chiu
Journal:  Int Orthop       Date:  2009-09-26       Impact factor: 3.075

7.  The value of a special surgical team in preventing complications in the treatment of hip fractures.

Authors:  M J Parker; G A Pryor; J W Myles
Journal:  Int Orthop       Date:  1994-06       Impact factor: 3.075

8.  A biomechanical study comparing helical blade with screw design for sliding hip fixations of unstable intertrochanteric fractures.

Authors:  Qiang Luo; Grace Yuen; Tak-Wing Lau; Kelvin Yeung; Frankie Leung
Journal:  ScientificWorldJournal       Date:  2013-02-20

9.  Critical factors in cut-out complication after Gamma Nail treatment of proximal femoral fractures.

Authors:  Alicja J Bojan; Claudia Beimel; Gilbert Taglang; David Collin; Carl Ekholm; Anders Jönsson
Journal:  BMC Musculoskelet Disord       Date:  2013-01-02       Impact factor: 2.362

10.  Outcome of intertrochanteric fractures treated by intramedullary nail with two integrated lag screws: A study in Asian population.

Authors:  Jong-Won Kim; Tae-Young Kim; Yong-Chan Ha; Young-Kyun Lee; Kyung-Hoi Koo
Journal:  Indian J Orthop       Date:  2015 Jul-Aug       Impact factor: 1.251

View more