| Literature DB >> 25895503 |
Anton Eklund1,2, Martin Slettengren3,4, Jan van der Linden5,6.
Abstract
INTRODUCTION: In the intensive care setting, most physiologic parameters are monitored automatically. However, urine output (UO) is still monitored hourly by manually handled urinometers. In this study, we evaluated an automatic urinometer (AU) and compared it with a manual urinometer (MU).Entities:
Mesh:
Year: 2015 PMID: 25895503 PMCID: PMC4416393 DOI: 10.1186/s13054-015-0899-4
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Staff opinion of an automatic urinometer compared with a manual urinometer
|
|
| ||||
|---|---|---|---|---|---|
|
|
|
|
|
| |
|
|
|
|
|
| |
| 1. How easy was it to learn to use the automatic urinometer? | 39% | 54% | 7% | 0% | 0% |
| 2. Was the collection of urine output data from the automatic urinometer easier compared with the manual urinometer? | 32% | 43% | 14% | 11% | 0% |
|
|
|
|
|
| |
| 3. Did you feel that you had less contact with the urine bags with the automatic urinometer compared with the manual urinometer? | 36% | 25% | 39% | 0% | 0% |
|
|
|
|
|
| |
| 4. Do you think the reliability of the urine output data is higher with the automatic urinometer than with the manual urinometer? | 21% | 64% | 7% | 0% | 7% |
| 5. Does using the automatic urinometer give you more time for other activities? | 0% | 32% | 68% | 0% | 0% |
Clinical variables of the patients in the automatic urinometer group and the manual urinometer group
|
|
|
|
|
|---|---|---|---|
| Female sex, % | 28 | 50 | 0.29 |
| Age, yr | 68.0 (64.3 to 75.5) | 66.5 (63.5 to 73.5) | 0.75 |
| Weight, kg | 79.2 ± 15.3 | 80.3 ± 7.9 | 0.83 |
| Height, cm | 174 ± 5.8 | 172 ± 9.8 | 0.48 |
| BMI, kg/m2 | 1.9 ± 0.2 | 2.0 ± 0.2 | 0.74 |
| EuroSCORE II, % | 1.7 (0.9 to 2.6) | 1.6 (0.9 to 3.1) | 0.85 |
| Preoperative albumin, g/L | 39.0 (36.0 to 40.3) | 38.0 (36.0 to 39.8) | 0.60 |
| Preoperative creatinine, μmol/L | 82.5 (67.8 to 97.8) | 86.5 (62.5 to 100) | 0.83 |
| Preoperative eGFR, ml/min | 78.0 (62.4 to 104) | 75.4 (67.8 to 89.4) | 0.91 |
| IDDM, % | 6 | 6 | 1.00 |
| COPD, % | 6 | 6 | 1.00 |
| LVEF <50%, % | 33 | 31 | 1.00 |
| Surgical procedures | |||
| CABG, % | 28 | 25 | 1.00 |
| Single-valve replacement, % | 44 | 44 | 1.00 |
| Valve + CABG, % | 11 | 19 | 0.65 |
| Others, % | 17 | 13 | 1.00 |
| ECC, min | 84.5 (71.5 to 133) | 79.5 (67.0 to 127) | 0.56 |
| ICU stay, hr | 22.5 (18.8 to 25.0) | 23.0 (18.5 to 25.3) | 0.77 |
| Ventilation time in ICU, hr | 3.0 (1.0 to 4.3) | 2.5 (1.0 to 4.8) | 0.96 |
| Inotropes in ICU, % | 0 | 6 | 0.47 |
aAU, Automatic urinometer; BMI, Body mass index; CABG, Coronary artery bypass graft; COPD, Chronic obstructive pulmonary disease; ECC, Extracorporeal circulation; eGFR, Estimated glomerular filtration rate (Cockcroft-Gault equation); EuroSCORE, European System for Cardiac Operative Risk Evaluation [8]; ICU, Intensive care unit; IDDM, Insulin-dependent diabetes mellitus; LVEF, Left ventricular ejection fraction; MU, Manual urinometer. Data are presented as percentages, medians (25th to 75th percentile range) or means ± standard deviation.
Performance parameters of the automatic urinometer and the manual urinometer
|
|
|
| |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| AU | All | 220 | 7.7 | 0.5 | +18.9 | +17.1 | +15.4 | +2.9 | +1.9 | +0.9 | −11.6 | −13.3 | −15.1 |
| <100 ml | 176 | 6.4 | 0.5 | +15.4 | +13.8 | +12.2 | +2.2 | +1.3 | +0.3 | −9.7 | −11.3 | −12.9 | |
| ≥100 ml | 44 | 11.4 | 1.7 | +33.5 | +27.6 | +21.6 | +8.0 | +4.5 | +1.0 | −12.6 | −18.6 | −24.5 | |
| MU | All | 188 | 8.4 | 0.6 | +23.9 | +21.8 | +19.8 | +6.5 | +5.3 | +4.1 | −9.2 | −11.3 | −13.4 |
| <100 ml | 124 | 4.5 | 0.4 | +13.1 | +11.6 | +10.4 | +3.6 | +2.8 | +2.0 | −4.8 | −6.2 | −7.6 | |
| ≥100 ml | 64 | 11.5 | 1.4 | +38.1 | +33.1 | +28.2 | +13.0 | +10.1 | +7.2 | −8.0 | −12.9 | −17.9 | |
aAU, Automatic urinometer; CI, Confidence interval; LOA, Limit of agreement; MU, Manual urinometer; n, Number of measurements; SD, Standard deviation; SE, Standard error; , Mean. For each urinometer, data are shown for all measurements combined, as well as subdivided at a volume of 100 ml. All parameters are in milliliters, except n.
Figure 1Bland-Altman plots of the agreement of each urinometer. Volumes are given in milliliters. In each plot, the red lines represent (from above): upper 95% limit of agreement; mean bias; lower 95% limit of agreement. Dotted lines represent confidence intervals of each parameter. Three measurements with x-axis volumes >400 ml were omitted for the purpose of visibility.
Figure 2Histograms of the agreement of each urinometer. Volumes are given in milliliters. Each bar represents an increment of 2 ml. A positive value represents an overestimation by the urinometer compared with a reference measurement by measuring cylinder.
Evaluation of diagnostics of urine output <40 ml/hr
|
|
|
|
|---|---|---|
| Sensitivity (%) | 90.4 | 98.2 |
| Specificity (%) | 98.5 | 96.2 |
| Positive predictive value (%) | 97.4 | 91.7 |
| Negative predictive value (%) | 94.4 | 99.2 |
aThe automatic urinometer (AU) and manual urinometer (MU) are compared with a measuring cylinder (gold standard).