Performance characteristics of a novel clustered multi-pinhole technology for simultaneous high-resolution SPECT/PET.

Kenta Miwa1, Masayuki Inubushi, Yasuto Takeuchi, Tetsuro Katafuchi, Mitsuru Koizumi, Tsuneo Saga, Masayuki Sasaki.   

Abstract

OBJECTIVE: Versatile emission computed tomography (VECTor) for small-animal imaging enables fully simultaneous SPECT/PET image acquisition based on clustered multi-pinhole collimation. The present study experimentally evaluated the detailed performance characteristics of the clustered multi-pinhole system for simultaneous (99m)Tc and (18)F imaging from the user perspective.
METHODS: Spatial resolution, sensitivity, count rate linearity were determined for the VECTor system (MILabs). Two hot-rod micro-resolution phantoms with 6 sectors were created to test the resolution of (99m)Tc and (18)F. Sensitivity and count rate linearity were measured by scanning (99m)Tc and (18)F point sources positioned at the center of the field of view. Furthermore, we quantified the influence of (18)F on (99m)Tc SPECT images. The ratios of SPECT counts on the (99m)Tc-only and simultaneous (18)F-(99m)Tc at various time points were evaluated as a function of the (18)F-to-(99m)Tc activity concentration ratio.
RESULTS: The 0.5-mm hot-rods can be visually distinguished in the (99m)Tc image, and 0.8-mm rods for (18)F remained clearly visible. The point-source sensitivity was 2800 cps/MBq for (99m)Tc and 2899 cps/MBq for (18)F, respectively. Count rates up to 120,000 cps for one bed position were linear for the activity. Spill-over from (18)F into (99m)Tc SPECT images was negligible when the activity concentration of the administered (18)F solution exceeded that of the (99m)Tc solution by up to a factor of 2.
CONCLUSIONS: We evaluated the performance characteristics of the VECTor that lead to determination of the optimal administered doses of (99m)Tc and (18)F tracers. We found that the VECTor achieved high resolution and high sensitivity as well as good (99m)Tc and (18)F linearity. Simultaneous SPECT/PET imaging with (99m)Tc and (18)F tracers, and stand-alone (99m)Tc and (18)F imaging using clustered-pinhole collimators is feasible and practical for a wide range of research applications using small animals.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25894055     DOI: 10.1007/s12149-015-0966-6

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  4 in total

1.  Quantification of Myocardial Perfusion Defect Size in Rats: Comparison between Quantitative Perfusion SPECT and Autoradiography.

Authors:  Hiroshi Wakabayashi; Junichi Taki; Anri Inaki; Tomo Hiromasa; Koichi Okuda; Takayuki Shibutani; Kazuhiro Shiba; Seigo Kinuya
Journal:  Mol Imaging Biol       Date:  2018-01-16       Impact factor: 3.488

2.  Prognostic Value of Early Evaluation of Left Ventricular Dyssynchrony After Myocardial Infarction.

Authors:  Hiroshi Wakabayashi; Junichi Taki; Anri Inaki; Tomo Hiromasa; Takafumi Yamase; Norihito Akatani; Koichi Okuda; Takayuki Shibutani; Kazuhiro Shiba; Seigo Kinuya
Journal:  Mol Imaging Biol       Date:  2019-08       Impact factor: 3.488

3.  H2RSPET: a 0.5 mm resolution high-sensitivity small-animal PET scanner, a simulation study.

Authors:  Youfang Lai; Qian Wang; Shiwei Zhou; Zhaoheng Xie; Jinyi Qi; Simon R Cherry; Mingwu Jin; Yujie Chi; Junwei Du
Journal:  Phys Med Biol       Date:  2021-03-09       Impact factor: 3.609

4.  64Cu-ATSM internal radiotherapy to treat tumors with bevacizumab-induced vascular decrease and hypoxia in human colon carcinoma xenografts.

Authors:  Yukie Yoshii; Mitsuyoshi Yoshimoto; Hiroki Matsumoto; Takako Furukawa; Ming-Rong Zhang; Masayuki Inubushi; Atsushi B Tsuji; Yasuhisa Fujibayashi; Tatsuya Higashi; Tsuneo Saga
Journal:  Oncotarget       Date:  2017-09-28
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.