Evaluation of the Keeler Pulsair non-contact tonometer.

W E Sponsel1, P L Kaufman, T I Strinden, K L DePaul, H N Bowes, K W Olander, H S Barnebey.   

Abstract

The recently-introduced Pulsair non-contact tonometer (Keeler Instruments, Inc, Broomall, PA) was evaluated against the Goldmann applanation tonometer in cannulated post mortem human eyes, and in living subjects at three clinical centers. Statistical analysis of the findings revealed strong linear relationships between Goldmann or manometric and Pulsair intraocular pressure readings, with correlation coefficients ranging from 0.79 to 0.97. However, with the Pulsair reading as the dependent variable, the slope of the relationship in each case was significantly less than 1.0 (range 0.65 to 0.83), while the intercept was positive (range 1.31 to 2.33 mmHg). Thus, the Pulsair tended to read low at intraocular pressure above the normal range. Conversion factors and a calibration table allowing intraocular pressure estimates more comparable to Goldmann tonometry using the current (J-series) Pulsair are provided herein. A calibration algorithm could be integrated into the circuitry of future production models of the Pulsair instrument.

Entities:  

Mesh:

Year:  1989        PMID: 2589057     DOI: 10.1111/j.1755-3768.1989.tb04109.x

Source DB:  PubMed          Journal:  Acta Ophthalmol (Copenh)        ISSN: 0001-639X


  4 in total

1.  Evaluation of Keeler Pulsair non-contact tonometry: reliability and reproducibility.

Authors:  L Bonomi; S Baravelli; C Cobbe; L Tomazzoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1991       Impact factor: 3.117

2.  Clinical comparison of Pulsair non-contact tonometer and Goldmann applanation tonometer in Indian population.

Authors:  Shalini Mohan; Satyaprakash Tiwari; Arvind Jain; Jaya Gupta; Surendra Kumar Sachan
Journal:  J Optom       Date:  2013-07-26

3.  Reproducibility with the Keeler Pulsair 2000 non-contact tonometer.

Authors:  S A Vernon
Journal:  Br J Ophthalmol       Date:  1995-06       Impact factor: 4.638

4.  Intraocular pressure measurement after penetrating keratoplasty: minified Goldmann applanation tonometer, pneumatonometer, and Tono-Pen versus manometry.

Authors:  M J Ménage; P L Kaufman; M A Croft; S P Landay
Journal:  Br J Ophthalmol       Date:  1994-09       Impact factor: 4.638