Surgically induced astigmatism assessment: comparison between three corneal measuring devices.

Shay Ofir, Adi Abulafia, Guy Kleinmann, Olga Reitblat, Ehud I Assia.   

Abstract

PURPOSE: To compare the findings of three corneal measuring devices for calculating surgically induced astigmatism (SIA) during cataract extraction.
METHODS: Patients' records were retrospectively reviewed to identify patients who had corneal astigmatism measurements before and at least 1 month after cataract surgery through 2.4-mm corneal incisions by all three tested devices (Lenstar LS900, Haag-Streit, Koeniz, Switzerland; IOLMaster 500, Carl Zeiss Meditec, Dublin, CA; and Atlas topographer, Carl Zeiss Meditec). Vector analysis was used to calculate the SIA for each measuring device.
RESULTS: Seventy eyes of 49 patients were included. All three measuring devices had similar SIA results: the median SIAs were 0.45 diopters (D) for the Lenstar, 0.41 D for the IOLMaster, and 0.47 D for the Atlas topographer (P = .884).
CONCLUSIONS: The three evaluated anterior corneal measuring devices produced similar results in measuring SIA. [J Refract Surg. 2015;31(4):244-247.]. Copyright 2015, SLACK Incorporated.

Entities:  

Mesh:

Year:  2015        PMID: 25884579     DOI: 10.3928/1081597X-20150319-04

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Clinical evaluation of surgery-induced astigmatism in cataract surgery using 2.2 mm or 1.8 mm clear corneal micro-incisions.

Authors:  Jun Yang; Xiu Wang; Hong Zhang; Yi Pang; Rui-Hua Wei
Journal:  Int J Ophthalmol       Date:  2017-01-18       Impact factor: 1.779

2.  Outcomes of 1.8-3.0 mm incision phacoemulsification combined with trabeculectomy for primary angle-closure glaucoma with cataract.

Authors:  Qing Wang; Zheng-Xuan Jiang; Rong-Feng Liao
Journal:  Int J Ophthalmol       Date:  2020-02-18       Impact factor: 1.779

3.  Analysis of cataract surgery induced astigmatism: Two polar methods comparison.

Authors:  Veronica García-López; Claudia García-López; Victoria de Juan; Raul Martin
Journal:  J Optom       Date:  2016-12-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.