Arnold Bainomugisa1,2, Eddie Wampande3,4, Chris Muchwa5,6, Joseph Akol7,8, Paul Mubiri9, Henry Ssenyungule10,11, Enock Matovu12, Sam Ogwang13,14, Moses Joloba15,16.
Abstract
BACKGROUND: The incidence of M. tuberculosis (MTB) and non tuberculous Mycobacterium species (NTMs) like M. avium and M. kansasii has increased due to Human Immunodeficiency Virus (HIV) epidemic. Therefore accurate, rapid and cost effective methods for the identification of these NTMs and MTB are greatly needed for appropriate TB management. Thus in this study we evaluated the performance of Lightcycler(®) Mycobacterium detection assay to detect MTB, M. avium and M. kansasii in sputum specimens.Entities:
Mesh:
Year: 2015 PMID: 25884439 PMCID: PMC4424587 DOI: 10.1186/s12879-015-0921-0
Source DB: PubMed Journal: BMC Infect Dis ISSN: 1471-2334 Impact factor: 3.090
Figure 1GenoType Mycobacterium CM line probe assay results used to differentiate M. avium and M. kansasii from Mycobacteria other than Tuberculosis (MOTTs).
Figure 2Sigmoid amplification curves captured during anealing and elongation step by the Roche LightCycler 480 II indicating the detection of M. tuberculosis, M. avium or M. kansasii.
Figure 3Melting temperature determination using Tm calling on LightCycler 480® software determining the different mycobacteria species using temperature ranges; 54.4°C to 57.4°C for M. tuberculosis, 47.5°C to 50.5°C for M. avium and 57.5°C to 60.5°C for M. kansasii.
Analysis of the sputum specimens by Mycobacterium detection assay and liquid culture
|
|
|
| ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
| |
| Baseline |
|
|
| 132 |
|
|
| 14 | 131 |
|
2x2 performance analysis of Real time PCR and Liquid culture
|
| |||||||
|---|---|---|---|---|---|---|---|
|
|
|
|
|
| |||
| Liquid culture | Baseline | MTB |
| - | - | 94 | |
|
| - |
| - | - | 2 | ||
| Negative | 1 | 11 |
|
| 145 | ||
Performance of Real time PCR while using Liquid culture as the reference standard
|
| ||||
|---|---|---|---|---|
|
|
| |||
|
|
|
|
| |
| Sensitivity (%) | 100 | 96–100 | 100 | 19–100 |
| Specificity (%) | 99 | 96–99 | 95 | 91–97 |
| Prevalence (%) | 39 | 33–45 | 0.8 | 0.1–2.9 |
| PPV (%) | 98 | 94–99 | 15 | 2–45 |
| NPV (%) | 100 | 97–100 | 100 | 98–100 |
| Likelihood (+) | 144 | 20–1018 | 21 | 12–38 |
| Likelihood (−) | 0.00 | - | 0.00 | - |
| Kappa stat | 0.9 | 0.9–1.0 | 0.3 | -0.03–0.5 |
Costs comparison between Real time PCR assay and liquid culture
|
|
|
| ||
|---|---|---|---|---|
|
|
|
|
| |
| Bactec 960 MGIT culture media | 3 | 1,458 | ||
| Bactec 960 MGIT supplement | 2 | 232 | ||
| ZN carbolfuschin stain (BD) | 1 | 79 | ||
| ZN Decolorizer stain (BD) | 1 | 60 | ||
| Methylene Blue stain (BD) | 1 | 70 | ||
| Capilia TB Neo | 1 | 1,200 | ||
| Genotype mycobacterium CM kit | 1 | 1,800 | ||
| Light cycler Mycobacterium detection kit | 1 | 2,500 | ||
| PCR plates | 3 | 450 | ||
|
|
|
| ||
| US Unit cost ($) | 20.0 | 12.2 | ||