New methods for optical distance indicator and gantry angle quality control tests in medical linear accelerators: image processing by using a 3D phantom.

Mahdi Heravian Shandiz1, Ghorban Safaeian Layen2, Kazem Anvari3, Mohammadmahdi Khalilzadeh1.   

Abstract

PURPOSE: In order to keep the acceptable level of the radiation oncology linear accelerators, it is necessary to apply a reliable quality assurance (QA) program.
MATERIALS AND METHODS: The QA protocols, published by authoritative organizations, such as the American Association of Physicists in Medicine (AAPM), determine the quality control (QC) tests which should be performed on the medical linear accelerators and the threshold levels for each test. The purpose of this study is to increase the accuracy and precision of the selected QC tests in order to increase the quality of treatment and also increase the speed of the tests to convince the crowded centers to start a reliable QA program. A new method has been developed for two of the QC tests; optical distance indicator (ODI) QC test as a daily test and gantry angle QC test as a monthly test. This method uses an image processing approach utilizing the snapshots taken by the CCD camera to measure the source to surface distance (SSD) and gantry angle.
RESULTS: The new method of ODI QC test has an accuracy of 99.95% with a standard deviation of 0.061 cm and the new method for gantry angle QC has a precision of 0.43°. The automated proposed method which is used for both ODI and gantry angle QC tests, contains highly accurate and precise results which are objective and the human-caused errors have no effect on the results.
CONCLUSION: The results show that they are in the acceptable range for both of the QC tests, according to AAPM task group 142.

Entities:  

Keywords:  Gantry angle; Linear accelerator; Optical distance indicator (ODI); Quality control; Source to surface distance (SSD)

Year:  2015        PMID: 25874177      PMCID: PMC4394068          DOI: 10.3857/roj.2015.33.1.42

Source DB:  PubMed          Journal:  Radiat Oncol J        ISSN: 2234-1900


  5 in total

1.  Distance measurement based on pixel variation of CCD images.

Authors:  Chen-Chien Hsu; Ming-Chih Lu; Wei-Yen Wang; Yin-Yu Lu
Journal:  ISA Trans       Date:  2009-06-21       Impact factor: 5.468

2.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

3.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40.

Authors:  G J Kutcher; L Coia; M Gillin; W F Hanson; S Leibel; R J Morton; J R Palta; J A Purdy; L E Reinstein; G K Svensson
Journal:  Med Phys       Date:  1994-04       Impact factor: 4.071

4.  Technical innovation to calibrate the gantry angle indicators of linear accelerators.

Authors:  L Chang; S Y Ho; J M Wu; C Y Yu; C C Sung
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

5.  A new geometric and mechanical verification device for medical LINACs.

Authors:  Keith T Welsh; Robert A Wlodarczyk; L E Reinstein
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

  5 in total
  3 in total

1.  Automating quality assurance of digital linear accelerators using a radioluminescent phosphor coated phantom and optical imaging.

Authors:  Cesare H Jenkins; Dominik J Naczynski; Shu-Jung S Yu; Yong Yang; Lei Xing
Journal:  Phys Med Biol       Date:  2016-08-12       Impact factor: 3.609

2.  A novel and independent method for time-resolved gantry angle quality assurance for VMAT.

Authors:  Todsaporn Fuangrod; Peter B Greer; Benjamin J Zwan; Michael P Barnes; Joerg Lehmann
Journal:  J Appl Clin Med Phys       Date:  2017-07-13       Impact factor: 2.102

3.  Correlation analysis between 2D and quasi-3D gamma evaluations for both intensity-modulated radiation therapy and volumetric modulated arc therapy.

Authors:  Jung-In Kim; Chang Heon Choi; Hong-Gyun Wu; Jin Ho Kim; Kyubo Kim; Jong Min Park
Journal:  Oncotarget       Date:  2017-01-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.