Literature DB >> 25855463

Reliability of Oscillometric Blood Pressure Monitoring in Atrial Fibrillation Patients Admitted for Electric Cardioversion.

Monica Maselli1, Valter Giantin1, Domenico Corrado2, Alessandro Franchin1, Francesca Attanasio1, Valentina Pengo1, Alessandra Tramontano1, Pietro De Toni1, Egle Perissinotto3, Enzo Manzato4.   

Abstract

The reliability of automated oscillometric blood pressure (BP) monitors in atrial fibrillation (AF) has been poorly investigated, only comparing different patients with AF and sinus rhythm (SR), and is a method influenced by individual characteristics. This study compared the reliability of the oscillometric device A&D TM-2430 (A&D Company, Tokyo, Japan) with that of a mercury sphygmomanometer in AF patients whose SR was restored after electric cardioversion (ECV). Three consecutive BP measurements were obtained on the day before and about 30 days after ECV in stable treatment conditions. Of the 100 patients studied, 63 reported an SR at follow-up, with a significant increase in systolic BP and a significant decrease in diastolic BP according to both devices. There were no significant differences between the systolic and diastolic biases before and after ECV using Bland Altman analysis (P > .05 each). The oscillometric device analyzed, using three repeated measurements, is reliable in measuring BP in AF patients.
© 2015 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2015        PMID: 25855463      PMCID: PMC8031867          DOI: 10.1111/jch.12547

Source DB:  PubMed          Journal:  J Clin Hypertens (Greenwich)        ISSN: 1524-6175            Impact factor:   3.738


  24 in total

1.  Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects' characteristics.

Authors: 
Journal:  Blood Press Monit       Date:  1998-08       Impact factor: 1.444

2.  European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults.

Authors:  Eoin O'Brien; Neil Atkins; George Stergiou; Nikos Karpettas; Gianfranco Parati; Roland Asmar; Yutaka Imai; Jiguang Wang; Thomas Mengden; Andrew Shennan
Journal:  Blood Press Monit       Date:  2010-02       Impact factor: 1.444

3.  Atrial fibrillation--the growing epidemic.

Authors:  Gregory Y H Lip; Puneet Kakar; Timothy Watson
Journal:  Heart       Date:  2007-05       Impact factor: 5.994

Review 4.  Hypertension and atrial fibrillation: diagnostic approach, prevention and treatment. Position paper of the Working Group 'Hypertension Arrhythmias and Thrombosis' of the European Society of Hypertension.

Authors:  Athanasios J Manolis; Enrico Agabiti Rosei; Antonio Coca; Renata Cifkova; Serap E Erdine; Sverre Kjeldsen; Gregory Y H Lip; Krzysztof Narkiewicz; Gianfranco Parati; Josep Redon; Roland Schmieder; Costas Tsioufis; Giuseppe Mancia
Journal:  J Hypertens       Date:  2012-02       Impact factor: 4.844

5.  Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings.

Authors:  S N Hunyor; J M Flynn; C Cochineas
Journal:  Br Med J       Date:  1978-07-15

6.  Electrical cardioversion for atrial fibrillation improves microvascular flow independent of blood pressure changes.

Authors:  Paul W G Elbers; Wilhelmina B Prins; Herbert W M Plokker; Eric P A van Dongen; Mat van Iterson; Can Ince
Journal:  J Cardiothorac Vasc Anesth       Date:  2012-06-09       Impact factor: 2.628

7.  Ambulatory blood pressure monitoring in atrial fibrillation.

Authors:  G Y Lip; J Zarifis; M Beevers; D G Beevers
Journal:  Am J Cardiol       Date:  1996-08-01       Impact factor: 2.778

8.  Effects of sinus rhythm restoration in patients with persistent atrial fibrillation: a clinical, echocardiographic and hormonal study.

Authors:  Beata Wozakowska-Kapłon; Grzegorz Opolski
Journal:  Int J Cardiol       Date:  2004-08       Impact factor: 4.164

9.  2013 ESH/ESC Guidelines for the management of arterial hypertension: the Task Force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC).

Authors:  Giuseppe Mancia; Robert Fagard; Krzysztof Narkiewicz; Josep Redón; Alberto Zanchetti; Michael Böhm; Thierry Christiaens; Renata Cifkova; Guy De Backer; Anna Dominiczak; Maurizio Galderisi; Diederick E Grobbee; Tiny Jaarsma; Paulus Kirchhof; Sverre E Kjeldsen; Stéphane Laurent; Athanasios J Manolis; Peter M Nilsson; Luis Miguel Ruilope; Roland E Schmieder; Per Anton Sirnes; Peter Sleight; Margus Viigimaa; Bernard Waeber; Faiez Zannad
Journal:  J Hypertens       Date:  2013-07       Impact factor: 4.844

10.  Blood pressure measurement in atrial fibrillation: goodbye mercury?

Authors:  T Watson; G Y H Lip
Journal:  J Hum Hypertens       Date:  2006-04-13       Impact factor: 3.012

View more
  8 in total

1.  How to evaluate BP measurements using the oscillometric method in atrial fibrillation: the value of pulse rate variation.

Authors:  Xi-Xing Wang; Wei Shuai; Kui Hong; Jinsong Xu; Ju-Xiang Li; Ping Li; Xiao-Shu Cheng; Hai Su
Journal:  Hypertens Res       Date:  2016-03-31       Impact factor: 3.872

2.  Systolic blood pressure increases in patients with atrial fibrillation regaining sinus rhythm after electrical cardioversion.

Authors:  Joakim Olbers; Ellen Jacobson; Fredrik Viberg; Nils Witt; Petter Ljungman; Mårten Rosenqvist; Jan Östergren
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-14       Impact factor: 3.738

3.  How does blood pressure change in hypertensive patients with atrial fibrillation after successful electrical cardioversion?

Authors:  Cesare Cuspidi; Marijana Tadic; Guido Grassi
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-14       Impact factor: 3.738

4.  Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis.

Authors:  Christopher E Clark; Sinead T J McDonagh; Richard J McManus
Journal:  J Hum Hypertens       Date:  2019-01-10       Impact factor: 3.012

5.  Reliability of Oscillometric Blood Pressure Monitoring in Atrial Fibrillation Patients Admitted for Electric Cardioversion.

Authors:  Monica Maselli; Valter Giantin; Domenico Corrado; Alessandro Franchin; Francesca Attanasio; Valentina Pengo; Alessandra Tramontano; Pietro De Toni; Egle Perissinotto; Enzo Manzato
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-04-09       Impact factor: 3.738

6.  Should Oscillometric Blood Pressure Monitors Be Used in Patients With Atrial Fibrillation?

Authors:  Martin G Myers; George S Stergiou
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-04-09       Impact factor: 3.738

7.  Auscultatory versus oscillometric blood pressure measurement in patients with atrial fibrillation and arterial hypertension.

Authors:  Aistėja Šelmytė-Besusparė; Jūratė Barysienė; Dovilė Petrikonytė; Audrius Aidietis; Germanas Marinskis; Aleksandras Laucevičius
Journal:  BMC Cardiovasc Disord       Date:  2017-03-23       Impact factor: 2.298

8.  Intrasession Reliability Analysis for Oscillometric Blood Pressure Method Using a Digital Blood Pressure Monitor in Peruvian Population.

Authors:  Sabina Barrios-Fernandez; Eduardo Manuel Sosa-Sánchez; Jorge Carlos-Vivas; Laura Muñoz-Bermejo; Jesús Morenas-Martín; María Dolores Apolo-Arenas; Jose Carmelo Adsuar; Francisco Javier Domínguez-Muñoz
Journal:  Healthcare (Basel)       Date:  2022-01-21
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.