| Literature DB >> 25853025 |
Katsutoshi Sugimoto1, Fuminori Moriyasu1, Hirohito Takeuchi1, Mayumi Ando1, Takatomo Sano1, Toshifumi Mori1, Yoshihiro Furuichi1, Yoshiyuki Kobayashi1, Ikuo Nakamura1.
Abstract
INTRODUCTION: Irreversible electroporation (IRE) is a promising technique for the focal treatment of soft tissue tumors. Even though the local application of an excessive electric field is a potential cause of cardiac arrhythmias, initial clinical studies have shown that IRE is generally safe when cardiac gating is employed. CASE DESCRIPTION: In this case report, we observed an episode of ventricular extrasystoles without hemodynamic changes during which time the synchronization device failed to operate properly, with pulses delivered not in the absolute refractory period but in the relative refractory period. DISCUSSION AND EVALUATION: At present, persons performing IRE must keep in mind that there is a small but real risk of synchronization failure even when a cardiac synchronization device is used.Entities:
Keywords: Arrhythmia; Complication; Irreversible electroporation; Liver; Metastasis; Tumor
Year: 2015 PMID: 25853025 PMCID: PMC4326365 DOI: 10.1186/s40064-015-0828-7
Source DB: PubMed Journal: Springerplus ISSN: 2193-1801
Figure 1Illustration of the IRE electrode placed near the heart. A: Sagittal B-mode US image demonstrates that the IRE electrode (arrow) is placed in the periphery of the hypoechoic lesion (arrowheads) in segment 4. The distance from the electrode to the inferior border of the heart (asterisk) is 1.7 cm based on US measurement. B: Sagittal B-mode US image demonstrates that the IRE electrode (arrow) is placed in the periphery of the hypoechoic lesion (arrowheads) in segment 2. The distance from the electrode to the inferior border of the heart (asterisk) is 1.0 cm based on US measurement.
Figure 2Electrocardiogram during IRE treatment. A: Tracing obtained during IRE treatment of the hepatic mass in segment 4 with the synchronization device. The distance from the electrode to the inferior border of the heart is 1.7 cm based on US measurement. IRE pulses are incorrectly delivered on the T wave (arrows), resulting in an episode of ventricular extrasystoles (asterisks). B: The synchronization device failed to operate properly during treatment.
Figure 3Electrocardiogram during IRE treatment. A: Tracing obtained during IRE treatment of the hepatic mass in segment 2 with the synchronization device. The distance from the electrode to the inferior border of the heart is 1.0 cm based on US measurement. IRE pulses are correctly delivered on the R wave (arrows). B: The synchronization device operated properly during treatment.