Y Oguz1, E R Watanabe2, J M Reis3, R Spin-Neto4, M A Gabrielli2, V A Pereira-Filho2.
Abstract
The sagittal split ramus osteotomy (SSRO) is a surgical technique used widely to treat many congenital and acquired mandibular discrepancies. Stabilization of the osteotomy site and the potential for skeletal relapse after the procedure are still major problems. The aim of this study was to compare the mechanical stability of six methods of rigid fixation in SSRO using a biomechanical test model. Sixty polyurethane replicas of human hemimandibles were divided into six groups. In group I, the osteotomies were fixed with two four-hole titanium miniplates; in group II, with one four-hole miniplate; in group III, with one four-hole miniplate+a bicortical screw; in group IV, with a grid miniplate; in group V, with a four-hole locking miniplate; and in group VI, with a six-hole miniplate. A linear load in the premolar region was applied to the hemimandibles. The resistance forces (N) needed to displace the distal segment by 1, 3, and 5mm were recorded and the data transmitted from the load cell to a computer. One-way analysis of variance with Tukey's post hoc test was performed to compare the means between groups. For the three displacement conditions, there was a strong tendency for the 2.0-mm plate+screw and the grid plate to have higher values.Entities:
Keywords: biomechanical test; fixation; polyurethane hemimandibles; sagittal split ramus osteotomy
Mesh:
Substances:
Year: 2015 PMID: 25840861 DOI: 10.1016/j.ijom.2014.11.019
Source DB: PubMed Journal: Int J Oral Maxillofac Surg ISSN: 0901-5027 Impact factor: 2.789