In vitro biomechanical comparison of six different fixation methods following 5-mm sagittal split advancement osteotomies.

Y Oguz1, E R Watanabe2, J M Reis3, R Spin-Neto4, M A Gabrielli2, V A Pereira-Filho2.   

Abstract

The sagittal split ramus osteotomy (SSRO) is a surgical technique used widely to treat many congenital and acquired mandibular discrepancies. Stabilization of the osteotomy site and the potential for skeletal relapse after the procedure are still major problems. The aim of this study was to compare the mechanical stability of six methods of rigid fixation in SSRO using a biomechanical test model. Sixty polyurethane replicas of human hemimandibles were divided into six groups. In group I, the osteotomies were fixed with two four-hole titanium miniplates; in group II, with one four-hole miniplate; in group III, with one four-hole miniplate+a bicortical screw; in group IV, with a grid miniplate; in group V, with a four-hole locking miniplate; and in group VI, with a six-hole miniplate. A linear load in the premolar region was applied to the hemimandibles. The resistance forces (N) needed to displace the distal segment by 1, 3, and 5mm were recorded and the data transmitted from the load cell to a computer. One-way analysis of variance with Tukey's post hoc test was performed to compare the means between groups. For the three displacement conditions, there was a strong tendency for the 2.0-mm plate+screw and the grid plate to have higher values.
Copyright © 2015 International Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  biomechanical test; fixation; polyurethane hemimandibles; sagittal split ramus osteotomy

Mesh:

Substances:

Year:  2015        PMID: 25840861     DOI: 10.1016/j.ijom.2014.11.019

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Surg        ISSN: 0901-5027            Impact factor:   2.789


  2 in total

1.  Influence of the mandibular plane and magnitude of the movement in sagittal split ramus osteotomy: an in vitro study.

Authors:  Gleysson Matias de Assis; Victor Diniz Borborema Dos Santos; Salomão Israel Monteiro Lourenço Queiroz; Adriano Rocha Germano
Journal:  Oral Maxillofac Surg       Date:  2022-01-10

2.  Biomechanical Evaluation of Seven Fixation Methods for Sagittal Split Ramus Osteotomy with Four Advancement Levels by Finite Element Analysis.

Authors:  Yu He; Henglei Zhang; Jia Qiao; Xi Fu; Shixing Xu; Qi Jin; Jianfeng Liu; Ying Chen; Bing Yu; Feng Niu
Journal:  Front Surg       Date:  2022-05-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.