Literature DB >> 25840309

Explanted multifocal intraocular lenses.

Marrie van der Mooren1, Roger Steinert2, Farrell Tyson2, Michelle J M Langeslag2, Patricia A Piers2.   

Abstract

UNLABELLED: We report 2 cases in which single-piece multifocal acrylic intraocular lenses (IOLs) were explanted because of complications related to the presence of glistenings in the bulk of the IOL optic. In both cases, the patients complained about blurry or hazy vision. In vivo slitlamp examinations prior to IOL explantation confirmed the presence of severe glistenings in the IOL optic in 1 case and moderate glistenings in the second case. In the first case, the symptoms resolved and both corrected and uncorrected distance visual acuities improved by 4 lines following IOL exchange with a monofocal IOL. In the second case, the visual symptoms persisted with a hard contact lens. Symptoms resolved following an exchange with a monofocal IOL that was free of glistenings. These findings indicate that straylight caused by IOLs with glistenings may be clinically significant in cases in which multifocal IOLs are implanted and patients require optimized retinal sensitivity. FINANCIAL DISCLOSURE: Mr. van der Mooren, Ms. Langeslag, and Dr. Piers are employees of Abbott Medical Optics, Inc. Drs. Steinert and Tyson are consultants to Abbott Medical Optics Inc.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25840309     DOI: 10.1016/j.jcrs.2015.02.005

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Method for in vitro assessment of straylight from intraocular lenses.

Authors:  Grzegorz Łabuz; Fernando Vargas-Martín; Thomas J T P van den Berg; Norberto López-Gil
Journal:  Biomed Opt Express       Date:  2015-10-19       Impact factor: 3.732

2.  Validation of a spectral light scattering method to differentiate large from small particles in intraocular lenses.

Authors:  Grzegorz Łabuz; Eleni Papadatou; Fernando Vargas-Martín; Norberto López-Gil; Nicolaas J Reus; Thomas J T P van den Berg
Journal:  Biomed Opt Express       Date:  2017-02-27       Impact factor: 3.732

3.  Vision with different presbyopia corrections simulated with a portable binocular visual simulator.

Authors:  Aiswaryah Radhakrishnan; Daniel Pascual; Susana Marcos; Carlos Dorronsoro
Journal:  PLoS One       Date:  2019-08-20       Impact factor: 3.240

4.  Patients' dissatisfaction with multifocal intraocular lenses managed by exchange with other multifocal lenses of different optical profiles.

Authors:  Olena Al-Shymali; Colm McAlinden; Jorge L Alio Del Barrio; Mario Canto-Cerdan; Jorge L Alio
Journal:  Eye Vis (Lond)       Date:  2022-03-01

5.  The assessment of the impact of glistening on visual performance in relation to tear film quality.

Authors:  Amanda Argay; Peter Vamosi
Journal:  PLoS One       Date:  2020-10-12       Impact factor: 3.240

  5 in total