Repeatability of topographic and aberrometric measurements at different accommodative states using a combined topographer and open-view aberrometer.

Claudia Gabriel1, Oliver Klaus Klaproth1, Christine Titke1, Martin Baumeister1, Jens Bühren1, Thomas Kohnen2.   

Abstract

PURPOSE: To evaluate the repeatability of corneal first-surface aberrations and aberrometric measurements of a combined aberrometer-topographer (iTrace Visual Function Analyzer) for 5 accommodative stimuli.
SETTING: Department of Ophthalmology, Goethe-University, Frankfurt, Germany.
DESIGN: Prospective case series.
METHODS: Eyes of young adults with low to moderate myopia or hyperopia were examined in 2 measurement series comprising 3 corneal topography measurements and wavefront measurements focusing on targets at 5 distances (0.4, 0.5, 0.6, 1.0, and 4.0 m). Repeatability was analyzed using the standard deviation (SD) within 6 measurements and the coefficient of repeatability (CoR) by Bland-Altman analysis of the means of the 2 measurement series.
RESULTS: Fifty eyes (50 subjects) were evaluated. The lowest calculated mean SD/CoR measurements using the aberrometer function at 5 target distances were 0.07 μm/0.14 μm for total ocular aberrations root mean square (RMS), 0.07 μm/0.16 μm for lower-order aberrations (LOAs) RMS, 0.03 μm/0.66 μm for higher-order aberrations (HOAs) RMS, 0.10 diopter (D)/0.22 D for sphere, 0.08 D/0.13 D for cylinder, and 0.09 D/0.21 D for the spherical equivalent (SE). Repeatability of the total RMS, LOAs RMS, sphere, and SE measurements improved with increasing target distance (P < .01 for aberrometric values; P < .03 for refractive values). Measurement reproducibility using the corneal topographer showed mean SDs/CoRs of 0.19 μm/0.41 μm for total RMS, 0.19 μm/0.39 μm for LOAs RMS, and 0.08 μm/0.23 μm for HOAs RMS.
CONCLUSION: The aberrometer showed satisfactory results for measurement repeatability at all target distances and declining repeatability with increasing accommodation effort. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 25840305     DOI: 10.1016/j.jcrs.2014.07.037

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  4 in total

1.  Validation of refraction and anterior segment parameters by a new multi-diagnostic platform (VX120).

Authors:  Ariela Gordon-Shaag; David P Piñero; Cyril Kahloun; David Markov; Tzadok Parnes; Liat Gantz; Einat Shneor
Journal:  J Optom       Date:  2018-03-09

2.  Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Alberto López-Navarro; Inmaculada Cabezos; Dolores de Fez; María T Caballero; Vicent J Camps
Journal:  Clin Optom (Auckl)       Date:  2017-05-04

3.  Precision and agreement of higher order aberrations measured with ray tracing and Hartmann-Shack aberrometers.

Authors:  Zequan Xu; Yanjun Hua; Wei Qiu; Guoqiang Li; Qiang Wu
Journal:  BMC Ophthalmol       Date:  2018-01-27       Impact factor: 2.209

4.  Reliability of Ocular Aberration Measurements in Children with Moderate and Low Myopia under Scotopic Conditions.

Authors:  Zhouyue Li; Dongmei Cui; Sichun Ao; Weizhong Lan; Xiao Yang
Journal:  J Ophthalmol       Date:  2018-02-11       Impact factor: 1.909

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.