David P Piñero1, Alberto López-Navarro2, Inmaculada Cabezos1, Dolores de Fez1, María T Caballero1, Vicent J Camps1.
Abstract
PURPOSE: To evaluate the intrasession repeatability of refractive and ocular aberrometric measurements obtained using a new multidiagnostic device in healthy eyes. PATIENTS AND METHODS: A total of 107 eyes of 107 patients, age ranging from 23 to 65 years, were enrolled in this study. A complete eye examination was performed in all eyes, including an ocular examination using the VX120 system. Three consecutive measurements were obtained using this device to assess the intrasession repeatability of different refractive and ocular aber-rometric parameters. The within-subject standard deviation (Sw), intrasubject precision (1.96×Sw), and intraclass correlation coefficient (ICC) were calculated.Entities:
Keywords: Scheimpflug imaging; VX120 system; intrasession repeatability; ocular aberrations; refraction
Year: 2017 PMID: 30214365 PMCID: PMC6095565 DOI: 10.2147/OPTO.S132572
Source DB: PubMed Journal: Clin Optom (Auckl) ISSN: 1179-2752
Summary of the intrasession repeatability outcomes for the refractive measurements obtained by means of the VX120 system
| Refractive parameter | Overall mean (SD) | Sw (°) | Pr (°) | ICC (range 95% CI) |
|---|---|---|---|---|
| Overall median (range) | ||||
| Sphere (D) | −0.43 (2.02) | 0.05 | 0.11 | 0.997 (0.996 to 0.998) |
| 0.00 (−7.00 to 6.00) | ||||
| Cylinder (D) | −0.40 (0.75) | 0.10 | 0.19 | 0.959 (0.944 to 0.971) |
| 0.00 (−4.00 to 0.00) | ||||
| J0 (D) | 0.06 (0.35) | 0.04 | 0.07 | 0.975 (0.965 to 0.982) |
| 0.00 (−0.99 to 1.85) | ||||
| J45 (D) | 0.01 (0.23) | 0.03 | 0.07 | 0.947 (0.926 to 0.962) |
| 0.00 (−1.00 to 0.81) | ||||
| B (D) | 1.40 (1.65) | 0.09 | 0.18 | 0.995 (0.993 to 0.996) |
| 1.00 (0.00 to 7.34) |
Abbreviations: Sw, within-subject standard deviation; J0 and J45, power vector components; B, overall blurring strength; Pr, intrasession precision 1.96×Sw; SD, standard deviation; ICC, intraclass correlation coefficient.
Summary of the intrasession repeatability outcomes for the aberrometric measurements obtained by means of the VX120 system
| Aberrometric parameter | Overall mean (SD) | Sw (°) | Pr (°) | ICC (range 95% CI) |
|---|---|---|---|---|
| HOA RMS (µm) | 0.23 (0.13) | 0.04 | 0.09 | 0.922 (0.893 to 0.945) |
| Primary coma RMS (µm) | 0.12 (0.09) | 0.03 | 0.06 | 0.939 (0.916 to 0.957) |
| Trefoil RMS (µm) | 0.13 (0.09) | 0.03 | 0.06 | 0.940 (0.918 to 0.957) |
| Z40 (µm) | 0.06 (0.09) | 0.03 | 0.05 | 0.954 (0.936 to 0.967) |
| Secondary astigmatism RMS (µm) | 0.04 (0.04) | 0.02 | 0.03 | 0.861 (0.808 to 0.901) |
| Quadrafoil RMS (µm) | 0.04 (0.03) | 0.01 | 0.03 | 0.805 (0.731 to 0.861) |
Abbreviations: Sw, within-subject standard deviation; Pr, intrasession precision 1.96×Sw; Z40, Zernike term corresponding to the primary spherical aberration; SD, standard deviation; ICC, intraclass correlation coefficient; HOA, higher order aberrations; RMS, root mean square.
Summary of correlations between different refractive and aberrometric parameters and their within-subject standard deviation (Sw,) associated
| Correlation Pearson coefficient | ||
|---|---|---|
| Sphere with Sw | 0.053 | 0.589 |
| Cylinder with Sw | −0.194 | 0.045 |
| J0 with Sw | −0.033 | 0.739 |
| J45 with Sw | 0.024 | 0.807 |
| B with S w | 0.180 | 0.064 |
| HOA RMS with Sw | 0.469 | <0.001 |
| Primary coma RMS with Sw | 0.588 | <0.001 |
| Trefoil RMS with Sw | 0.449 | <0.001 |
| Z40 with Sw | 0.506 | <0.001 |
| Secondary astigmatism RMS with Sw | 0.446 | <0.001 |
| Quadrafoil RMS with Sw | 0.554 | <0.001 |
Abbreviations: Sw, within-subject standard deviation; J0 and J45, power vector components; B, overall blurring strength; Z40, Zernike term corresponding to the primary spherical aberration; HOA, higher order aberrations; RMS, root mean square.