Literature DB >> 25826278

Withdrawal of a novel-design duodenoscope ends outbreak of a VIM-2-producing Pseudomonas aeruginosa.

Charlotte J Verfaillie1, Marco J Bruno2, Anne F Voor in 't Holt3, Jolanda G Buijs4, Jan-Werner Poley2, Arjo J Loeve5, Juliette A Severin3, Leo F Abel2, Bert J Smit6, Inge de Goeij3, Margreet C Vos3.   

Abstract

BACKGROUND AND STUDY AIMS: Infections are a recognized risk of endoscopic retrograde cholangiopancreatography (ERCP). This paper reports on a large outbreak of VIM-2-producing Pseudomonas aeruginosa that was linked to the use of a recently introduced duodenoscope with a specific modified design (Olympus TJF-Q180V).
METHODS: Epidemiological investigations and molecular typing were executed in order to identify the source of the outbreak. Audits on implementation of infection control measures were performed. Additional infection control strategies were implemented to prevent further transmission. The design and the ability to clean and disinfect the duodenoscope were evaluated, and the distal tip was dismantled.
RESULTS: From January to April 2012, 30 patients with a VIM-2-positive P. aeruginosa were identified, of whom 22 had undergone an ERCP using a specific duodenoscope, the TJF-Q180V. This was a significant increase compared with the hospital-wide baseline level of 2 - 3 cases per month. Clonal relatedness of the VIM-2 P. aeruginosa was confirmed for all 22 cases and for the VIM-2 strain isolated from the recess under the forceps elevator of the duodenoscope. An investigational study of the new modified design, including the dismantling of the duodenoscope tip, revealed that the fixed distal cap hampered cleaning and disinfection, and that the O-ring might not seal the forceps elevator axis sufficiently. The high monthly number of cases decreased below the pre-existing baseline level following withdrawal of the TJF-Q180V device from clinical use.
CONCLUSIONS: Duodenoscope design modifications may compromise microbiological safety as illustrated by this outbreak. Extensive pre-marketing validation of the reprocessability of any new endoscope design and stringent post-marketing surveillance are therefore mandatory. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25826278     DOI: 10.1055/s-0034-1391886

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  31 in total

Review 1.  Duodenoscope-Associated Infections: Update on an Emerging Problem.

Authors:  M Rubayat Rahman; Abhilash Perisetti; Roxana Coman; Pardeep Bansal; Rajiv Chhabra; Hemant Goyal
Journal:  Dig Dis Sci       Date:  2018-12-19       Impact factor: 3.199

Review 2.  What every endoscopist should know about decontamination.

Authors:  Helen Griffiths; Laura Dwyer
Journal:  Frontline Gastroenterol       Date:  2019-01-18

3.  Outbreak Column 22: How to manage an outbreak.

Authors:  Evonne T Curran
Journal:  J Infect Prev       Date:  2017-11-24

Review 4.  Duodenoscope-Associated Bacterial Infections: A Review and Update.

Authors:  Jennifer T Higa; Michael Gluck; Andrew S Ross
Journal:  Curr Treat Options Gastroenterol       Date:  2016-06

Review 5.  Hygiene: The Looming Achilles Heel in Endoscopy.

Authors:  Michael Jung; Ulrike Beilenhoff
Journal:  Visc Med       Date:  2016-02-10

Review 6.  Software Tools in Endoscopy - Nice to Have or Essential?

Authors:  Oliver Möschler
Journal:  Visc Med       Date:  2016-01-29

Review 7.  Duodenoscope-associated infections: a review.

Authors:  Gheorghe G Balan; Catalin Victor Sfarti; Stefan Andrei Chiriac; Carol Stanciu; Anca Trifan
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2019-09-03       Impact factor: 3.267

8.  A Novel Protocol Obviates Endoscope Sampling for Carbapenem-Resistant Enterobacteriaceae: Experience of a Center with a Prior Outbreak.

Authors:  Zachary L Smith; Arshish Dua; Kia Saeian; Nathan A Ledeboer; Mary Beth Graham; Murad Aburajab; Darren D Ballard; Abdul H Khan; Kulwinder S Dua
Journal:  Dig Dis Sci       Date:  2017-07-05       Impact factor: 3.199

9.  Performance and applicability of a first generation single-use duodenoscope: a single-center cohort study.

Authors:  Diederik Persyn; Hannah Van Malenstein; Emma Vanderschueren; Annette Schuermans; Mira Dreesen; Wouter Meert; Kristof Buysschaert; Schalk Van Der Merwe; Wim Laleman
Journal:  Ann Gastroenterol       Date:  2022-06-02

10.  Turbulent fluid flow is a novel closed-system sample extraction method for flexible endoscope channels of various inner diameters.

Authors:  Seo Yean Sohn; Michelle J Alfa; Richard Lai; Yacoob Tabani; Mohamed E Labib
Journal:  J Microbiol Methods       Date:  2019-11-20       Impact factor: 2.363

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.