OBJECTIVES: The principal aim of this study was to assess the safety and effectiveness of the middle ear implant Vibrant Soundbridge (VSB) in patients with moderate-to-severe sensorineural hearing loss up to a mean (± standard deviation) duration of 11.1 ± 2.1 years (min. = 8.2, max. = 13.9, n = 16) after the intervention.
DESIGN: This was a retrospective, single-subject repeated-measurements study over a long-term period. A total of 104 German-speaking adults (for 122 implants) were included in this study (54 male, 50 female). The mean age at implantation was 54.5 years (min. = 19.0, max. = 80.4). Audiological outcome and speech intelligibility were assessed in all VSB patients at different time points in non-overlapping groups.
RESULTS: Bone conduction (BC) thresholds were preserved after the implantation and no indication was found of an increase over time of the small air-bone gaps introduced by the implantation. BC and air conduction thresholds worsened similarly in both implanted and non-implanted ears over time. The decrease in audiological benefit provided by the VSB was moderate and the Word Recognition Score in quiet conditions at 65 dB SPL was still largely improved with the VSB in the longest observed group.
CONCLUSIONS: These results confirm that the VSB does not affect the integrity of the inner/middle ear and is still beneficial in long-term follow-up.
© 2015 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2015        PMID: 25824716     DOI: 10.1159/000368387

Source DB:  PubMed          Journal:  Audiol Neurootol        ISSN: 1420-3030            Impact factor:   1.854


  5 in total

1.  Clinical predictors for satisfaction with incus vibroplasty: a preliminary study.

Authors:  Jae Joon Han; Jihye Rhee; Jae-Jin Song; Ja-Won Koo; Byung Yoon Choi
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-12-05       Impact factor: 2.503

2.  Importance of adhesiolysis in revision surgery for vibrant soundbridge device failures at the short incus process.

Authors:  Yoon Ah Park; Tae Hoon Kong; Jin Soon Chang; Young Joon Seo
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-08-22       Impact factor: 2.503

3.  The Hannover Coupler V2: Audiological outcomes of a round window coupler for the Vibrant Soundbridge.

Authors:  Nicole Knölke; Dawid Murawski; Nina Wardenga; Susan Busch; Hannes Maier; Thomas Lenarz
Journal:  Laryngoscope Investig Otolaryngol       Date:  2022-05-28

4.  High-frequency hearing, tinnitus, and patient satisfaction with stapedotomy: A randomized prospective study.

Authors:  Dan Bagger-Sjöbäck; Karin Strömbäck; Malou Hultcrantz; Georgios Papatziamos; Henrik Smeds; Niklas Danckwardt-Lillieström; Bo Tideholm; Ann Johansson; Sten Hellström; Pierre Hakizimana; Anders Fridberger
Journal:  Sci Rep       Date:  2015-08-21       Impact factor: 4.379

5.  Progressive Sensorineural Hearing Loss in Vibrant Soundbridge Users Requiring Cochlear Implantation.

Authors:  Faris F Brkic; Wolf-Dieter Baumgartner; Dominik Riss; Thomas Thurner; David T Liu; Wolfgang Gstöttner; Erich Vyskocil
Journal:  J Pers Med       Date:  2022-01-31
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.