Literature DB >> 25820883

The effect of frictional torque and bending moment on corrosion at the taper interface : an in vitro study.

A Panagiotidou1, J Meswania1, K Osman2, B Bolland3, J Latham4, J Skinner5, F S Haddad6, A Hart5, G Blunn1.   

Abstract

The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper interface of a modular femoral component and to investigate whether different combinations of material also had an effect. The combinations we examined were 1) cobalt-chromium (CoCr) heads on CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14 Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending moment by offsetting the application of axial load from the midline in the mediolateral plane. Increments of offset equivalent to head + 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing torque for all combinations of material. However, Ti stems showed the highest corrosion currents. Increased bending moments associated with using larger offset heads produced more corrosion: Ti stems generally performed worse than CoCr stems. Using ceramic heads did not prevent corrosion, but reduced it significantly in all loading configurations. ©2015 The British Editorial Society of Bone & Joint Surgery.

Entities:  

Keywords:  Corrosion; Metal on Metal; Taper; Total Hip Replacement; Wear

Mesh:

Substances:

Year:  2015        PMID: 25820883     DOI: 10.1302/0301-620X.97B4.34800

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  14 in total

Review 1.  [Tribology in hip arthroplasty : Benefits of different materials].

Authors:  J Philippe Kretzer; Maximilian Uhler; Sebastian Jäger; Therese Bormann; Robert Sonntag; Mareike Schonhoff; Stefan Schröder
Journal:  Orthopade       Date:  2021-02-25       Impact factor: 1.087

Review 2.  Corrosion of the Head-Stem Taper Junction-Are We on the Verge of an Epidemic?: Review Article.

Authors:  Michael Morlock; Dennis Bünte; Julian Gührs; Nicholas Bishop
Journal:  HSS J       Date:  2016-09-20

Review 3.  Management of metal-on-metal hip implant patients: Who, when and how to revise?

Authors:  Reshid Berber; John A Skinner; Alister J Hart
Journal:  World J Orthop       Date:  2016-05-18

Review 4.  Personalized Hip Joint Replacement with Large Diameter Head: Current Concepts.

Authors:  Pascal-André Vendittoli; Sagi Martinov; Mina Wahba Morcos; Sivan Sivaloganathan; William G Blakeney
Journal:  J Clin Med       Date:  2022-03-30       Impact factor: 4.241

5.  Large-diameter total hip arthroplasty modular heads require greater assembly forces for initial stability.

Authors:  A R MacLeod; N P T Sullivan; M R Whitehouse; H S Gill
Journal:  Bone Joint Res       Date:  2016-08       Impact factor: 5.853

6.  A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement.

Authors:  Thom Bitter; Imran Khan; Tim Marriott; Elaine Lovelady; Nico Verdonschot; Dennis Janssen
Journal:  Proc Inst Mech Eng H       Date:  2017-06-09       Impact factor: 1.617

7.  An Analytical Calculation of Frictional and Bending Moments at the Head-Neck Interface of Hip Joint Implants during Different Physiological Activities.

Authors:  Hamidreza Farhoudi; Reza H Oskouei; Ali A Pasha Zanoosi; Claire F Jones; Mark Taylor
Journal:  Materials (Basel)       Date:  2016-12-05       Impact factor: 3.623

8.  Clinically insignificant trunnionosis in large-diameter metal-on-polyethylene total hip arthroplasty.

Authors:  H S Hothi; D Kendoff; C Lausmann; J Henckel; T Gehrke; J Skinner; A Hart
Journal:  Bone Joint Res       Date:  2017-01       Impact factor: 5.853

Review 9.  Trunnion corrosion: what surgeons need to know in 2018.

Authors:  J R Berstock; M R Whitehouse; C P Duncan
Journal:  Bone Joint J       Date:  2018-01       Impact factor: 5.082

10.  Three-dimensional friction measurement during hip simulation.

Authors:  Robert Sonntag; Steffen Braun; Loay Al-Salehi; Joern Reinders; Ulrike Mueller; J Philippe Kretzer
Journal:  PLoS One       Date:  2017-09-08       Impact factor: 3.240

View more