OBJECTIVE: To evaluate the level of ambient radiation in a PET/CT center.Entities:
Keywords: Dosimetry; PET/CT; Thermoluminescence dosimetry
Year: 2015 PMID: 25798004 PMCID: PMC4366025 DOI: 10.1590/0100-3984.2013.1821
Source DB: PubMed Journal: Radiol Bras ISSN: 0100-3984
Figure 1Sketch of the first floor of Centro de Tecnologia em Imagem Molecular (CIMol).
Figure 2Sketch of the second floor of Centro de Tecnologia em Imagem Molecular (CIMol).
Estimate of annual equivalent dose for photons at each monitored spot.
| Spot | Shielding description | Shielding material | Thickness (cm) | T | Measured dose (mSv) | Hx(mSv/year) | Uncertainty (mSv/year) | Area classification | Conclusion |
|---|---|---|---|---|---|---|---|---|---|
| 2 | Slab | Reinforced concrete | 26.0 | 1/4 | 0.12 | 0.12 | ± 0.02 | Free | Appropriate |
| 3 | Slab | Reinforced concrete | 26.0 | 1/2 | 0.15 | 0.30 | ± 0.06 | Free | Appropriate |
| 4 | Slab | Reinforced concrete | 26.0 | 1/6 | 0.09 | 0.06 | ± 0.01 | Free | Appropriate |
| 5 | Slab | Reinforced concrete | 26.0 | 1 | 0.09 | 0.36 | ± 0.07 | Free | Appropriate |
| 6 | Slab | Reinforced concrete | 12.0 | 1/6 | 0.09 | 0.06 | ± 0.01 | Free | Appropriate |
| 7 | Wall | Reinforced concrete | 16.0 | 1 | 0.09 | 0.36 | ± 0.07 | Free | Appropriate |
| 8 | Wall (several) | Reinforced concrete + masonry | 4 X 15.0 + 14.0 + 16.0 | 1/2 | 0.18 | 0.36 | ± 0.07 | Free | Appropriate |
| 9 | Wall (several) | Reinforced concrete + masonry | 16.0 + 25.0 | 1/4 | 0.18 | 0.18 | ± 0.04 | Free | Appropriate |
| 10 | Wall (several) | Reinforced concrete | 15.0 | 1 | 0.09 | 0.36 | ± 0.07 | Free | Appropriate |
| 11 | Wall (several) | Concrete + masonry | 15.0 + 15.0 | 1/16 | 0.15 | 0.04 | ± 0.01 | Free | Appropriate |
| 12 | Wall | Reinforced concrete | 16.0 | 1 | 0.15 | 0.60 | ± 0.11 | Controlled | Appropriate |
| 13 | Wall | Plumbiferous glass + reinforced concrete | 3.0 + 16.0 | 1 | 0.12 | 0.48 | ± 0.10 | Controlled | Appropriate |
| 14 | None | — | — | 1 | 0.30 | 1.2 | ± 0.24 | Controlled | Appropriate |
| 15 | None | — | — | 1 | 0.27 | 1.08 | ± 0.22 | Controlled | Appropriate |
| 16 | Fixed screen | Lead + plumbiferous glass + reinforced concrete | 2.0 + 3.0 + 16.0 | 1 | 0.18 | 0.72 | ± 0.14 | Controlled | Appropriate |
| 17 | Wall | Reinforced concrete + masonry | 15.0 + 16.0 | 1/2 | 0.18 | 0.36 | ± 0.07 | Free | Appropriate |
| 18 | Wall (several) | Reinforced concrete + masonry | 15.0 + 16.0 | 1/2 | 0.15 | 0.30 | ± 0.06 | Free | Appropriate |
| 19 | Wall | Reinforced concrete | 16.0 | 1 | 0.15 | 0.60 | ± 0.12 | Controlled | Appropriate |