Peter J Crick1, T William Bentley, Yuqin Wang, William J Griffiths.
Abstract
Sterols, and specifically oxysterols, play important roles in the biosynthesis of bile acids and steroid hormones as well as possessing biological activities in their own right. Analysis of oxysterols is complicated due to their low abundance in biological systems and poor ionisation characteristics in mass spectrometry. Over the past decade, we have developed a liquid chromatography-mass spectrometry method termed enzyme-assisted derivatisation for sterol analysis (EADSA). Our derivatisation procedure relies on two solid-phase extraction steps to (i) separate cholesterol from oxysterols and (ii) remove excess derivatisation reagents. Recent inter-batch variation in C18 reversed-phase cartridges has led us to experiment with alternative columns. Here, we present our findings and report an improved sample preparation procedure using polymeric hydrophilic-lipophilic balanced reversed-phase cartridges.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25796527 PMCID: PMC4471317 DOI: 10.1007/s00216-015-8609-2
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142
Fig. 1Structures and representative chromatograms of synthetic standards after derivatisation by EADSA. Retention times: DHEAS, 1.15 min; [2H7]24R/S-HC, 7.82 min; cholest-4-en-3-one, 11.99 min. [2H7]24R/S-HC gives four peaks as each of the R and S isomers give syn and anti conformers upon EADSA. SPE-2 was from batch C
Fig. 2Comparison of SPE cartridges for the removal of cholesterol. Reconstructed ion chromatograms (RIC) for 518.4105 ± 10 ppm. (a) Waters Sep-Pak tC18 batch 011032331C, (b) Waters Sep-Pak tC18 batch 011133059D, and (c) Waters Oasis HLB. Cholesterol retention time: 11.99 min. Waters Oasis HLB was used for SPE-2 in all cases
Fig. 3Comparison of SPE cartridges for the recovery of side-chain hydroxycholesterols after EADSA. RICs for (a) m/z 541.4493 ± 10 ppm showing [2H7]24R/S-HC. Top panel SPE-2 is Waters Sep-Pak tC18 batch 011032331C; bottom panel SPE-2 is Waters Oasis HLB and (b) RIC for m/z 534.4054 ± 10 ppm showing endogenous side-chain hydroxycholesterols. Top panel SPE-2 is Waters Sep-Pak tC18 batch 011032331C; bottom panel SPE-2 is Waters Oasis HLB. Peaks in (a) and (b) are normalised to the most intense peak in each column