AIM OF THE STUDY: To use different beam arrangements and numbers to plan intensity-modulated radiation therapy (IMRT) and investigate their effects on low and high radiation doses delivered to the functional lung, in order to reduce radiation-induced lung damage.Entities:
Keywords: dosiometry beam angle; intensity-modulated radiation therapy; non-small lung carcinoma; number of beams
Year: 2014 PMID: 25784844 PMCID: PMC4355654 DOI: 10.5114/wo.2014.46237
Source DB: PubMed Journal: Contemp Oncol (Pozn) ISSN: 1428-2526
Fig. 1Example of a seven-beam radiotherapy plan for a patient with left lung cancer
Fig. 3Example of a five-beam radiotherapy plan for a patient with left lung cancer
Functional Lung volume receiving ≥ 20 Gy (FV20) as a percentage of the total functional lung volume for the seven-beam computed tomography (CT-7) plan and the seven-beam single-photon emission CT (SPECT-7) plan
| Patient | CT-7 (%) | SPECT-7 (%) | Change (%) | % Reduction |
|---|---|---|---|---|
| 1 | 7.64 | 6.26 | 1.38 | 18.06 |
| 2 | 6.6 | 5.32 | 1.28 | 19.39 |
| 3 | 14.86 | 7.91 | 6.95 | 46.77 |
| 4 | 12.44 | 12.51 | –0.07 | –0.56 |
| 5 | 19 | 13.69 | 5.31 | 27.95 |
| 6 | 18.28 | 14.06 | 4.22 | 23.09 |
| 7 | 8.61 | 7.55 | 1.06 | 12.31 |
| 8 | 13.33 | 6.98 | 6.35 | 47.64 |
| 9 | 6.63 | 5.02 | 1.61 | 24.28 |
| 10 | 8.07 | 4.74 | 3.33 | 41.26 |
| Average | 3.14 ±2.46 | 26.02 ±15.45 |
Reduction in FV20 in the SPECT-7 plan as compared to FV20 in the CT-7 plan
Functional Lung volume receiving ≥ 30 Gy (FV30) as a percentage of the total functional lung volume for the seven-beam computed tomography (CT-7) Plan and the seven-beam single-photon emission CT (SPECT-7) plan
| Patient | CT-7 (%) | SPECT-7 (%) | Change (%) | % Reduction |
|---|---|---|---|---|
| 1 | 2.41 | 2.2 | 0.21 | 8.71 |
| 2 | 3.89 | 2.56 | 1.33 | 34.19 |
| 3 | 3.31 | 3.11 | 0.2 | 6.04 |
| 4 | 7.07 | 6.62 | 0.45 | 6.37 |
| 5 | 8.64 | 6.28 | 2.36 | 27.32 |
| 6 | 4.89 | 5.63 | –0.74 | –15.13 |
| 7 | 3.19 | 3.23 | –0.04 | –1.25 |
| 8 | 6.5 | 4.88 | 1.62 | 24.92 |
| 9 | 3.23 | 2.74 | 0.49 | 15.17 |
| 10 | 3.02 | 1.88 | 1.14 | 37.75 |
| Average | 0.70 ±0.91 | 14.41 ±16.66 |
Reduction in FV30 in the SPECT-7 plan as compared to FV30 in the CT-7 plan
Fig. 4Dose-volume histograms of seven-beam computed tomography (CT-7) and single-photon emission CT (SPECT-7) plans for the planning target volume (PTV; green lines) and the total lung volume (blue lines) of a patient with stage III non-small cell lung cancer
Fig. 5Dose-volume histograms of four-, five- and seven-beam single- photon emission computed tomography (SPECT-4, SPECT-5 and SPECT-7, respectively) for the planning target volume (PTV; green lines) and total lung volume (blue lines) of a patient with stage III non-small cell lung cancer
Comparison of dose-related parameters between different radiotherapy plans
| Parameter | SPECT-4 | SPECT-5 | SPECT-7 |
|---|---|---|---|
| FV5 (%) | 26.56 ±10.25 | 36.84 ±11.31 | 41.80 ±14.29 |
| FV13 (%) | 17.52 ±6.81 | 18.40 ±6.26 | 18.06 ±10.27 |
| FV20 (%) | 10.60 ±3.40 | 10.14 ±3.63 | 8.40 ±3.63 |
| FV30 (%) | 5.61 ±2.38 | 4.39 ±1.95 | 3.91 ±1.77 |
| MLD (Gy) | 9.02 ±2.55 | 9.55 ±2.63 | 9.64 ±2.87 |
| Cord max (Gy) | 39.42 ±9.73 | 39.32 ±8.66 | 38.51 ±7.09 |
| Heart V40 (%) | 6.47 ±10.24 | 7.04 ±11.55 | 5.85 ±10.19 |
| CI | 0.77 ±0.61 | 0.86 ±0.03 | 0.87 ±0.05 |
| HI | 1.74 ±O.33 | 1.56 ±0.24 | 1.44 ±0.11 |
All values are expressed as mean and standard deviation.
Compared to the five-beam single-photon emission computed tomography (SPECT-5) plan: t1,2,3,4 = –4.403, 2.659, –3.422, 4.307; p < 0.05.
Compared to the seven-beam SPECT (SPECT-7) plan: t1,2,3,4 = –3.087, 3.681, 3.408, 2.498; p < 0.05.
Compared to the four-beam SPECT (SPECT-4) plan: t1,2,3,4,5 = –4.409, 2.322, 3.433, –2.543, 6.503, 3.317, p < 0.05.