| Literature DB >> 25763090 |
Jaeuk U Kim1, Yu Jung Lee1, Jeon Lee2, Jong Yeol Kim1.
Abstract
Aim of the Study. The three conventional pulse-diagnostic palpation locations (PLs) on both wrists are Cun, Guan, and Chi, and each location reveals different clinical information. To identify anatomical or hemodynamic specificity, we used ultrasonographic imaging to determine the arterial diameter, radial artery depth, and arterial blood flow velocity at the three PLs and at nearby non-PL segments. Methods. We applied an ultrasound scanner to 44 subjects and studied the changes in the arterial diameter and depth as well as in the average/maximum blood flow velocities along the radial artery at three PLs and three non-PLs located more proximally than Chi. Results. All of the measurements at all of the PLs were significantly different (P < 0.01). Artery depth was significantly different among the non-PLs; however, this difference became insignificant after normalization to the arm circumference. Conclusions. Substantial changes in the hemodynamic and anatomical properties of the radial artery around the three PLs were insignificant at the nearby non-PLs segments. This finding may provide a partial explanation for the diagnostic use of "Cun, Guan, and Chi.".Entities:
Year: 2015 PMID: 25763090 PMCID: PMC4339866 DOI: 10.1155/2015/381634
Source DB: PubMed Journal: Evid Based Complement Alternat Med ISSN: 1741-427X Impact factor: 2.629
Subjects' information.
| Subject information | Mean ± SD |
|---|---|
| Number ( | 23 males and 21 females |
| Age (yr) | 23.17 ± 2.16 |
| Height (cm) | 167.26 ± 8.04 |
| Weight (kg) | 59.04 ± 7.84 |
| BMI (kg/m2) | 21.01 ± 1.27 |
| Systolic/diastolic blood pressure (mmHg) | 115.80/70.19 ± 15.15/8.29 |
| Total cholesterol (mg/dL) | 165.40 ± 30.91 |
Figure 1(a) Six measurement locations composed of the three pulse-diagnostic locations (Cun, Guan, and Chi) and non-pulse-diagnostic locations (P1, P2, and P3) on the radial artery and (b) marking of the measurement locations by thin wires (left) and a measurement scene (right).
Artery depth and diameter and blood flow velocities.
| Variable | Measurement location [mean (SD)] | |||||
|---|---|---|---|---|---|---|
|
|
|
| P1 | P2 | P3 | |
| Left | ||||||
| Artery depth (mm) | 3.26 (0.71) | 2.74 (0.66) | 3.32 (0.85) | 3.79 (1.09) | 3.90 (1.11) | 4.17 (1.26) |
| Artery diameter (mm) | 2.52 (0.36) | 2.42 (0.35) | 2.52 (0.30) | 2.54 (0.32) | 2.51 (0.31) | 2.51 (0.31) |
| Maximum blood flow velocity (cm/sec) | 41.68 (12.39) | 55.34 (14.70) | 56.26 (11.82) | 54.98 (12.23) | 56.36 (12.21) | 57.66 (13.57) |
| Average blood flow velocity (cm/sec) | 5.75 (3.25) | 9.24 (4.86) | 9.51 (4.44) | 9.81 (5.24) | 9.66 (4.51) | 10.14 (5.04) |
| Right | ||||||
| Artery depth (mm) | 3.60 (0.82) | 2.74 (0.72) | 3.36 (1.14) | 3.95 (1.36) | 4.32 (1.42) | 4.64 (1.41) |
| Artery diameter (mm) | 2.47 (0.39) | 2.37 (0.32) | 2.46 (0.33) | 2.53 (0.41) | 2.55 (0.34) | 2.55 (0.33) |
| Maximum blood flow velocity (cm/sec) | 35.50 (4.64) | 49.28 (11.95) | 50.91 (11.55) | 51.43 (12.53) | 52.85 (13.15) | 53.94 (12.13) |
| Average blood flow velocity (cm/sec) | 4.64 (2.99) | 7.46 (4.24) | 7.98 (4.41) | 8.26 (4.44) | 8.45 (5.12) | 8.39 (4.43) |
Figure 3Diagram of the anatomical and hemodynamic characteristics at both arms.
Results of ANOVA and multiple comparison analysis.
| ANOVA | Scheffe's grouping | |||||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
| P1 | P2 | P3 | ||
| Left | Artery diameter | <0.0001 | A | B | A | A | A | A |
| Artery depth | <0.0001 | A | B | A | C | D | E | |
| Average blood flow velocity | <0.0001 | A | B | B | B | B | C | |
|
| ||||||||
| Right | Artery diameter | <0.0001 | A | C | A | A | B | B |
| Artery depth | <0.0001 | A | B | C | D | E | F | |
| Average blood flow velocity | <0.0001 | A | B | C | C | C | C | |
Each letter from A to F indicates distinct group by Scheffe's test.
Paired t-test between the left and right arms.
| Artery depth | Artery diameter | Average blood flow velocity | Maximum blood flow velocity | |
|---|---|---|---|---|
| L_ | 0.004** | 0.329 | 0.008** | 0.000** |
| L_ | 0.987 | 0.224 | 0.001** | 0.000** |
| L_ | 0.740 | 0.049* | 0.003** | 0.000** |
| L_P1-R_P1 | 0.169 | 0.852 | 0.005** | 0.000** |
| L_P2-R_P2 | 0.000** | 0.253 | 0.011* | 0.000** |
| L_P3-R_P3 | 0.000** | 0.299 | 0.001* | 0.017* |
L: left, R: right, * P < .05, and ** P < .01.
Figure 4(a) Ultrasonographic image which shows the changes in the vessel features around the pulse measuring locations and (b) it shows the detailed shape of the prominent bone beneath Guan.
Figure 2(a) The measurement of the arterial depth and diameter in B-mode and (b) the measurement of the blood flow velocity in the Color Doppler mode.