Quadrupole mass filters with added hexapole fields.

Zilan Xiao1, XianZhen Zhao, D J Douglas.   

Abstract

Conventional mass analysis has been investigated experimentally with six quadrupole mass filters with added hexapole fields; three with added hexapole fields of 4%, 8% and 12% with equal diameter rods, and three with added hexapole fields of 4%, 8% and 12% with unequal diameter rods to remove an added octopole field. Compared with conventional quadrupoles, these rod sets have very large field distortions. With the positive resolving dc applied to the y rods (Mathieu parameter a(x) < 0) only low resolution (10-100) and low transmission are seen. With the polarity reversed (a(x) > 0) much higher resolution (> or = 1000) and transmission are possible. Increasing the magnitude of the added hexapole field decreases the limiting resolution at m/z 609. Removing the added octopole field increases the limiting resolution. In some cases structure is formed on the peaks. For a given scan line slope, U/V(rf), the resolution decreases as the amplitude of the added hexapole field increases. These results are consistent with changes to the stability diagrams, calculated here. With a(x) > 0, adding a hexapole field causes the x stability boundary to move outward with all rod sets. With a(x) < 0, the boundaries become diffuse and the tip of the stability diagram becomes rounded, limiting the resolution to ca. 10-100. Where comparisons are possible, experiments show the rod sets with added hexapole fields have transmission 10-300 times less than a conventional quadrupole. Thus these quadrupoles are less useful for mass analysis than conventional quadrupoles. However, it is surprising, given the highly distorted fields, that some of the quadrupoles give resolution of 1000 or more. Copyright 2010 John Wiley & Sons, Ltd.

Entities:  

Year:  2010        PMID: 20552578     DOI: 10.1002/rcm.4596

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  2 in total

1.  Design and performance evaluation of a linear ion trap mass analyzer featuring half round rod electrodes.

Authors:  Xiaoxu Li; Xiaohua Zhang; Rujiao Yao; Yang He; Yongyong Zhu; Jie Qian
Journal:  J Am Soc Mass Spectrom       Date:  2015-03-10       Impact factor: 3.109

2.  Navigate Flying Molecular Elephants Safely to the Ground: Mass-Selective Soft Landing up to the Mega-Dalton Range by Electrospray Controlled Ion-Beam Deposition.

Authors:  Andreas Walz; Karolina Stoiber; Annette Huettig; Hartmut Schlichting; Johannes V Barth
Journal:  Anal Chem       Date:  2022-05-24       Impact factor: 8.008

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.