Sliding-slab three-dimensional TSE imaging with a spiral-In/Out readout.

Zhiqiang Li1, Dinghui Wang1, Ryan K Robison1, Nicholas R Zwart1, Michael Schär1,2,3, John P Karis4, James G Pipe1.   

Abstract

PURPOSE: T2 -weighted imaging is of great diagnostic value in neuroimaging. Three-dimensional (3D) Cartesian turbo spin echo (TSE) scans provide high signal-to-noise ratio (SNR) and contiguous slice coverage. The purpose of this preliminary work is to implement a novel 3D spiral TSE technique with image quality comparable to 2D/3D Cartesian TSE.
METHODS: The proposed technique uses multislab 3D TSE imaging. To mitigate the slice boundary artifacts, a sliding-slab method is extended to spiral imaging. A spiral-in/out readout is adopted to minimize the artifacts that may be present with the conventional spiral-out readout. Phase errors induced by B0 eddy currents are measured and compensated to allow for the combination of the spiral-in and spiral-out images. A nonuniform slice encoding scheme is used to reduce the truncation artifacts while preserving the SNR performance.
RESULTS: Preliminary results show that each of the individual measures contributes to the overall performance, and the image quality of the results obtained with the proposed technique is, in general, comparable to that of 2D or 3D Cartesian TSE.
CONCLUSION: 3D sliding-slab TSE with a spiral-in/out readout provides good-quality T2 -weighted images, and, therefore, may become a promising alternative to Cartesian TSE.
© 2015 Wiley Periodicals, Inc.

Keywords:  3D TSE; sliding-slab; spiral-in/out readout

Mesh:

Year:  2015        PMID: 25753219     DOI: 10.1002/mrm.25660

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

1.  Automated design of pulse sequences for magnetic resonance fingerprinting using physics-inspired optimization.

Authors:  Stephen P Jordan; Siyuan Hu; Ignacio Rozada; Debra F McGivney; Rasim Boyacioğlu; Darryl C Jacob; Sherry Huang; Michael Beverland; Helmut G Katzgraber; Matthias Troyer; Mark A Griswold; Dan Ma
Journal:  Proc Natl Acad Sci U S A       Date:  2021-09-30       Impact factor: 11.205

2.  Reducing slab boundary artifacts in three-dimensional multislab diffusion MRI using nonlinear inversion for slab profile encoding (NPEN).

Authors:  Wenchuan Wu; Peter J Koopmans; Robert Frost; Karla L Miller
Journal:  Magn Reson Med       Date:  2015-10-28       Impact factor: 4.668

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.