Figure 1An abbreviated diagram of the process (boxes) and review (ovals) steps for intensity modulated radiation therapy planning for an individual patient. Each color (or shade) represents member of the treatment team. *Peer review will be addressed in detail in a report of the white paper series on patient safety.
Recommendations to guard against catastrophic failures for IMRT
| Recommended tests and procedures | Person who performs task | Primary review responsibility | Second review |
|---|---|---|---|
| Halt a procedure if the operator is unclear about what is being done. | All | All | All |
| Verify the patient information, treatment site, and prescription. | All | All | All |
| Verify correct positioning of the high dose region of isodose plan relative to targets. | Dosimetrist | Physician | Physicist |
| Verify the recording of reference and shift information from the planning scan in patient chart (electronic or paper). | Dosimetrist | Physicist | Therapist |
| Assess pretreatment localization/portal images with respect to corresponding reference images before first treatment; physician determines frequency of IGRT techniques. | Dosimetrist exports reference images from treatment planning system | Physician | Therapist |
| Verify that the correct version of the patient's treatment plan is approved, sent to treatment management system, and used for patient-specific QA. | Dosimetrist exports from the treatment planning system | Physicist | Therapists confirm against prescription for each treatment; physician prescription should specify the physician approved plan |
| Before the first treatment or for any change in treatment, perform patient-specific QA to guarantee that data transfer between systems is correct before patient treatment begins. | Physicist, dosimetrist, therapist or physics assistant | Physicist | Therapists confirm that only fully approved plans are used for treatment |
| Perform a complete chart check including review of information in treatment management system prior to the start of any treatment and after any change in treatment before changes are used for treatment. | Physicist | Therapist | |
| Perform a time out prior to treatment delivery. | Therapist | Second therapist | |
| Perform a check of treatment parameters before start of and during first treatment against a fixed version of the treatment plan. | Dosimetrist exported from TPS; verified by physicist | Therapist | Second therapist |
| Perform end-to-end testing to guarantee transfer of data among all systems involved in imaging, planning and dose delivery (periodically and after any software or hardware changes). | Physicist, therapist, or physics assistant | Physicist | Second physicist to review |
| Perform a time out prior to treatment delivery. | Therapist | Second therapist | |
| Perform a check of treatment parameters before start of and during first treatment against a fixed version of the treatment plan. | Dosimetrist exported from TPS; verified by physicist | Therapist | Second therapist |
| Perform end-to-end testing to guarantee transfer of data among all systems involved in imaging, planning and dose delivery (periodically and after any software or hardware changes). | Physicist, therapist, or physics assistant | Physicist | Second physicist to review |
IGRT, image-guided radiation therapy; IMRT, intensity modulated radiation therapy; MLC, multileaf collimator; QA, quality assurance; TPS, treatment planning system.