Fast, automatic, and accurate catheter reconstruction in HDR brachytherapy using an electromagnetic 3D tracking system.

Eric Poulin1, Emmanuel Racine1, Dirk Binnekamp2, Luc Beaulieu1.   

Abstract

PURPOSE: In high dose rate brachytherapy (HDR-B), current catheter reconstruction protocols are relatively slow and error prone. The purpose of this technical note is to evaluate the accuracy and the robustness of an electromagnetic (EM) tracking system for automated and real-time catheter reconstruction.
METHODS: For this preclinical study, a total of ten catheters were inserted in gelatin phantoms with different trajectories. Catheters were reconstructed using a 18G biopsy needle, used as an EM stylet and equipped with a miniaturized sensor, and the second generation Aurora(®) Planar Field Generator from Northern Digital Inc. The Aurora EM system provides position and orientation value with precisions of 0.7 mm and 0.2°, respectively. Phantoms were also scanned using a μCT (GE Healthcare) and Philips Big Bore clinical computed tomography (CT) system with a spatial resolution of 89 μm and 2 mm, respectively. Reconstructions using the EM stylet were compared to μCT and CT. To assess the robustness of the EM reconstruction, five catheters were reconstructed twice and compared.
RESULTS: Reconstruction time for one catheter was 10 s, leading to a total reconstruction time inferior to 3 min for a typical 17-catheter implant. When compared to the μCT, the mean EM tip identification error was 0.69 ± 0.29 mm while the CT error was 1.08 ± 0.67 mm. The mean 3D distance error was found to be 0.66 ± 0.33 mm and 1.08 ± 0.72 mm for the EM and CT, respectively. EM 3D catheter trajectories were found to be more accurate. A maximum difference of less than 0.6 mm was found between successive EM reconstructions.
CONCLUSIONS: The EM reconstruction was found to be more accurate and precise than the conventional methods used for catheter reconstruction in HDR-B. This approach can be applied to any type of catheters and applicators.

Mesh:

Year:  2015        PMID: 25735278     DOI: 10.1118/1.4908011

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

Review 1.  Review of strategies for MRI based reconstruction of endocavitary and interstitial applicators in brachytherapy of cervical cancer.

Authors:  José Richart; Vicente Carmona-Meseguer; Teresa García-Martínez; Antonio Herreros; Antonio Otal; Santiago Pellejero; Ana Tornero-López; José Pérez-Calatayud
Journal:  Rep Pract Oncol Radiother       Date:  2018-07-23

2.  Improved electromagnetic tracking for catheter path reconstruction with application in high-dose-rate brachytherapy.

Authors:  Elodie Lugez; Hossein Sadjadi; Chandra P Joshi; Selim G Akl; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-02-18       Impact factor: 2.924

3.  Evaluation of an active magnetic resonance tracking system for interstitial brachytherapy.

Authors:  Wei Wang; Akila N Viswanathan; Antonio L Damato; Yue Chen; Zion Tse; Li Pan; Junichi Tokuda; Ravi T Seethamraju; Charles L Dumoulin; Ehud J Schmidt; Robert A Cormack
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

Review 4.  Review of advanced catheter technologies in radiation oncology brachytherapy procedures.

Authors:  Jun Zhou; Leonid Zamdborg; Evelyn Sebastian
Journal:  Cancer Manag Res       Date:  2015-07-16       Impact factor: 3.989

Review 5.  Magnetic resonance imaging-guided brachytherapy for cervical cancer: initiating a program.

Authors:  Amir M Owrangi; Joann I Prisciandaro; Abraam Soliman; Ananth Ravi; William Y Song
Journal:  J Contemp Brachytherapy       Date:  2015-10-30

Review 6.  Performance and suitability assessment of a real-time 3D electromagnetic needle tracking system for interstitial brachytherapy.

Authors:  Samir Boutaleb; Emmanuel Racine; Olivier Fillion; Antonio Bonillas; Gilion Hautvast; Dirk Binnekamp; Luc Beaulieu
Journal:  J Contemp Brachytherapy       Date:  2015-09-14

Review 7.  Electromagnetic tracking for treatment verification in interstitial brachytherapy.

Authors:  Christoph Bert; Markus Kellermeier; Kari Tanderup
Journal:  J Contemp Brachytherapy       Date:  2016-11-02

8.  HDR brachytherapy in vivo source position verification using a 2D diode array: A Monte Carlo study.

Authors:  Joel Poder; Dean Cutajar; Susanna Guatelli; Marco Petasecca; Andrew Howie; Joseph Bucci; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-06-01       Impact factor: 2.102

9.  Electromagnetic tracking (EMT) technology for improved treatment quality assurance in interstitial brachytherapy.

Authors:  Markus Kellermeier; Jens Herbolzheimer; Stephan Kreppner; Michael Lotter; Vratislav Strnad; Christoph Bert
Journal:  J Appl Clin Med Phys       Date:  2017-01       Impact factor: 2.102

10.  DRRGenerator: A Three-dimensional Slicer Extension for the Rapid and Easy Development of Digitally Reconstructed Radiographs.

Authors:  Lance Levine; Marc Levine
Journal:  J Clin Imaging Sci       Date:  2020-10-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.