Literature DB >> 25729734

Efficacy of the measurement of 25-hydroxyvitamin D2 and D3 levels by using PerkinElmer liquid chromatography-tandem mass spectrometry vitamin D kit compared with DiaSorin radioimmunoassay kit and Elecsys vitamin D total assay.

Ho-Seok Kwak1, Hee-Jung Chung1, Dong-Hee Cho1, Mi-Hyun Park1, Eun-Suk Ku1, Eun Jung Park2, Han Jin Oh2.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25729734      PMCID: PMC4330182          DOI: 10.3343/alm.2015.35.2.263

Source DB:  PubMed          Journal:  Ann Lab Med        ISSN: 2234-3806            Impact factor:   3.464


× No keyword cloud information.
Vitamin D2 (ergocalciferol) and D3 (cholecalciferol) can be procured from exogenous sources. These are then metabolized to 25-hydroxyvitamin D (25OHD2 and 25OHD3) in the liver. Measuring the levels of both 25OHD2 and 25OHD3 is imperative in assessing clinical nutritional status [1]. Vitamin D2 or D3 is provided as a vitamin D supplement in many countries. Serum 25OHD levels can be measured by competitive binding assay, RIA, automated immunoassay, HPLC, and by the recently developed liquid chromatography-tandem mass spectrometry (LC-MS/MS) technique. LC-MS/MS is considered as the "gold standard" for the detection and quantification of 25OHD2 and 25OHD3. The MS/MS Vitamin D kit from PerkinElmer (PerkinElmer, Waltham, MA, USA) is a commercial reagent kit, intended for the quantitative determination of 25OHD2 and 25OHD3. The MS/MS Vitamin D kit protocol was compared with the following assays: RIA from DiaSorin (DiaSorin, Stillwater, MN, USA) and automated electro-chemiluminescence immunoassay (ECLIA) from Roche (Roche Diagnostics GmbH, Mannheim, Germany). After receiving approval by the Ethics Review Board of the Cheil General Hospital and Women's Healthcare Centre (Seoul, Korea), consecutive samples (n=50) sent for routine 25OHD analysis were used. The MS/MS Vitamin D kit was used along with the MS/MS Vitamin D Derivatization Box (PerkinElmer) on an LC-MS/MS system that included ACQUITY TQD tandem mass spectrometer (Waters, Milford, MA, USA). The MS/MS Vitamin D kit was compared with 25OHD 125I-based RIA kit and Elecsys Vitamin D Total assay. The MS/MS Vitamin D kit, RIA kit, and Elecsys Vitamin D Total assay were run according to the manufacturers' specifications. All three assays were compared by linear regression and Bland-Altman plot. The correlation between the methods was compared by using Pearson's correlation coefficient. Agreement in the assessment of the vitamin D status between methods was evaluated by using Cohen's kappa [2]. Statistical analysis was performed by SPSS software (version 18.0.0, SPSS Inc. Chicago, IL, USA). Precision of the LC-MS/MS method was evaluated by inter-assay CV (n=20) of quality control materials supplied by the manufacturer. At the three levels of 25OHD2, CV was <4.0%. At the three levels of 25OHD3, CV was <5.3%. Inter-assay CV for RIA and ECLIA were <13.0% and <9.8%, respectively. A comparison of LC-MS/MS with ECLIA yielded the following regression equation: ECLIA=1.1325×LC-MS/MS+0.52. The corresponding equation for RIA was: RIA=1.0546×LC-MS/MS-0.8733. In comparison with LC-MS/MS, the ECLIA demonstrated an R2 value of 0.8741 (Fig. 1A), with an average bias of +8.4 ng/mL (15.4%) (Fig. 1C), and the RIA demonstrated an R2 value of 0.8976 (Fig. 1B), with an average bias of +0.6 ng/mL (1.9%) (Fig. 1D). This trend was also demonstrated in previous reports, with ECLIA showing positive bias compared with LC-MS/MS [2, 3]. The distribution of results for 25OHD2 and 25OHD3 is shown in Fig. 2. The 25OHD3 levels showed no significant difference (Fig. 2A), while the 25OHD2 levels were biased towards the lower end (Fig. 2B). Compared to LC-MS/MS, having a cutoff of 20 ng/mL (insufficiency vs. normal), 4% (1/25) of the samples were misclassified as normal with RIA and 12% (3/25) of the samples were misclassified as normal with ECLIA. Relatively, agreement of RIA was better (kappa=0.96) than that of ECLIA (kappa= 0.88). RIA and ECLIA, which are currently employed in clinical laboratories for total 25OHD concentration measurement, showed an acceptable correlation with LC-MS/MS in the analytical range.
Fig. 1

Comparison between immunometric assays and LC-MS/MS (PerkinElmer MS/MS Vitamin D kit) for 25-hydroxyvitamin D quantification: (A, B) Linear regression between LC-MS/MS and ECLIA (Elecsys Vitamin D total assay), and LC-MS/MS and RIA (DiaSorin RIA kit), respectively. (C, D) Bland-Altman plot between LC-MS/MS and ECLIA, and LC-MS/MS and RIA, respectively. Open circles represent samples containing relatively low concentrations of 25-hydroxyvitamin D2 (<1 ng/mL), and black circles represent samples containing relatively high concentrations of 25-hydroxyvitamin D2 (≥1 ng/mL).

Abbreviations: ECLIA, electrochemiluminescence immunoassay; LC-MS/MS, liquid chromatography-tandem mass spectrometry.

Fig. 2

Distribution for 25-hydroxyvitamin D3 (25OHD3) (A) and 25-hydroxyvitamin D2 (25OHD2) (B). Results were obtained by analyzing serum samples provided by 50 volunteers.

The MS/MS Vitamin D kit allows for the quantitative determination of the most clinically relevant metabolite forms of vitamin D (25OHD2 and 25OHD3). The 25OHD levels determined by MS/MS Vitamin D kit were in overall agreement with the levels determined by DiaSorin RIA and Roche ECLIA.
  3 in total

1.  Comparison of four current 25-hydroxyvitamin D assays.

Authors:  Hee-Won Moon; Jin-Hee Cho; Mina Hur; Junghan Song; Gwi Young Oh; Chul Min Park; Yeo-Min Yun; Jin Q Kim
Journal:  Clin Biochem       Date:  2012-01-08       Impact factor: 3.281

2.  Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.

Authors:  Yu Chen; Lois Kinney; Andrea Božović; Hilary Smith; Heather Tarr; Eleftherios P Diamandis; Adrien LeBlanc
Journal:  Clin Biochem       Date:  2012-06-13       Impact factor: 3.281

3.  Quantification of serum 25-hydroxyvitamin D(2) and D(3) using HPLC-tandem mass spectrometry and examination of reference intervals for diagnosis of vitamin D deficiency.

Authors:  Amy K Saenger; Thomas J Laha; Deborah E Bremner; Sayed M H Sadrzadeh
Journal:  Am J Clin Pathol       Date:  2006-06       Impact factor: 2.493

  3 in total
  2 in total

1.  The association of serum levels of zinc and vitamin D with wasting among Iranian pre-school children.

Authors:  Pegah Nasiri-Babadi; Mehdi Sadeghian; Omid Sadeghi; Fereydoun Siassi; Ahmadreza Dorosty; Ahmad Esmaillzadeh; Hamed Pouraram
Journal:  Eat Weight Disord       Date:  2020-01-03       Impact factor: 4.652

2.  Discrepancy between Vitamin D Total Immunoassays due to Various Cross-reactivities.

Authors:  Jun Hyung Lee; Jee-Hye Choi; Oh Joo Kweon; Ae Ja Park
Journal:  J Bone Metab       Date:  2015-08-31
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.