Case study on the orientation of phaco hand pieces during steam sterilization processes.

J P C M van Doornmalen Gomez Hoyos1, R A C van Wezel2, H W J M van Doornmalen3.   

Abstract

BACKGROUND: Steam sterilization is an essential part of infection prevention. The literature shows that sterilization of medical instruments containing channels is not trivial. Phaco hand pieces have a simple configuration: a device contains a channel with a constant radius. No literature was found indicating whether the sterilization conditions on the inner surface of a phaco hand piece are influenced by the orientation of the hand piece. AIM: To determine whether the orientation of a phaco hand piece influences the results of a sterilization process of this device.
METHODS: A qualitative case study, including experiments, is performed with a protocolled combination of steam sterilizer, process, phaco hand piece, orientation of the phaco hand piece, and wrapping.
FINDINGS: In this specific case, the orientation of the hand piece influenced the result of the steam sterilization process; in vertically (upright) oriented phaco hand pieces with free water drainage, sterilization conditions are reproducibly established. In the same process, in horizontally oriented or vertically oriented hand pieces without free drainage, these conditions are not established in a reproducible way.
CONCLUSION: In the investigated combination of sterilizer, process, load, loading pattern and wrapping, phaco hand pieces have to be oriented vertically (upright) with free water drainage to obtain steam sterilization conditions on the inner surface. It is likely that instruments with comparable configuration and dimensions will yield comparable results. It is therefore recommended that this issue is considered during the development of medical instruments and during performance qualifications of such instruments.
Copyright © 2015 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Hand pieces; Loading; Orientation; Phaco; Steam sterilization; Sterilization

Mesh:

Substances:

Year:  2015        PMID: 25726434     DOI: 10.1016/j.jhin.2015.01.013

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  2 in total

1.  Compliance of primary and secondary care public hospitals with standard practices for reprocessing and steam sterilization of reusable medical devices in Nepal: findings from nation-wide multicenter clustered audits.

Authors:  Gopal Panta; Ann K Richardson; Ian C Shaw; Patricia A Coope
Journal:  BMC Health Serv Res       Date:  2020-10-07       Impact factor: 2.655

2.  Monitoring Steam Penetration in Channeled Instruments: An Evidence-Based Worst-Case for Practical Situations.

Authors:  Francesco Tessarolo; Michela Masè; Andrea Visonà; Josephus P C M van Doornmalen Gomez Hoyos
Journal:  Front Med Technol       Date:  2020-11-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.