A comparative planning study of step-and-shoot IMRT versus helical tomotherapy for whole-pelvis irradiation in cervical cancer.

Imjai Chitapanarux1, Ekkasit Tharavichitkul2, Wannapa Nobnop2, Somsak Wanwilairat2, Roy Vongtama3, Patrinee Traisathit4.   

Abstract

The aim of this study was to compare the dosimetric parameters of whole-pelvis radiotherapy (WPRT) for cervical cancer between step-and-shoot IMRT (SaS-IMRT) and Helical Tomotherapy™ (HT). Retrospective analysis was performed on 20 cervical cancer patients who received WPRT in our center between January 2011 and January 2014. SaS-IMRT and HT treatment plans were generated for each patient. The dosimetric values for target coverage and organ-at-risk (OAR) sparing were compared according to the criteria of the International Commission on Radiation Units and Measurements 83 (ICRU 83) guidelines. Differences in beam-on time (BOT) were also compared. All the PTV dosimetric parameters (D5%, D50% and D95%) for the HT plan were (statistically significantly) of better quality than those for the SaS-IMRT plan (P-value < 0.001 in all respects). HT was also significantly more accurate than SaS-IMRT with respect to the D98% and Dmean of the CTV (P-values of 0.008 and <0.001, respectively). The median Conformity Index (CI) did not differ between the two plans (P-value = 0.057). However, the Uniformity Index for HT was significantly better than that for SaS-IMRT (P-value < 0.001). The median of D50% for the bladder, rectum and small bowel were significantly lower in HT planning than SaS-IMRT (P-value < 0.001). For D2%, we found that HT provided better sparing to the rectum and bladder (P-value < 0.001). However, the median of D2% for the small bowel was comparable for both plans. The median of Dmax of the head of the left femur was significantly lower in the HT plan, but this did not apply for the head of the right femur. BOT for HT was significantly shorter than for SaS-IMRT (P-value < 0.001). HT provided highly accurate plans, with more homogeneous PTV coverage and superior sparing of OARs than SaS-IMRT. In addition, HT enabled a shorter delivery time than SaS-IMRT.
© The Author 2015. Published by Oxford University Press on behalf of The Japan Radiation Research Society and Japanese Society for Radiation Oncology.

Entities:  

Keywords:  IMRT; cervical cancer; step-and-shoot; tomotherapy

Mesh:

Year:  2015        PMID: 25720771      PMCID: PMC4426928          DOI: 10.1093/jrr/rrv004

Source DB:  PubMed          Journal:  J Radiat Res        ISSN: 0449-3060            Impact factor:   2.724


INTRODUCTION

Cervical cancer is the second most common cancer among women in Thailand; 9999 new cases were diagnosed in 2008 [1]. A significant survival benefit has been demonstrated for the combined approach of concurrent chemoradiotherapy (CCRT) [2-4]. The radiation therapy component consists of external beam irradiation to the primary tumor and regional lymph nodes, followed by a brachytherapy boost to the gross tumor in the cervix. However, acute Grade 3 or 4 hematological and gastrointestinal toxicities were found to be significantly higher in the CCRT group than in the radiotherapy (RT) alone group. Tan et al. [5] also demonstrated increased late toxicity after CCRT for locally advanced cervical cancer. To combat this increase in toxicity, whole-pelvic intensity-modulated radiotherapy (WP-IMRT) has been applied to gynecologic malignancies with excellent planning target volume (PTV) coverage and this has been associated with less acute gastrointestinal sequelae than conventional whole-pelvic radiotherapy (WPRT), as reported by Mundt et al. [6]. It has been demonstrated that, with the same target coverage, IMRT is superior to conventional techniques in normal tissue sparing for the treatment of cervical cancer, with lower gastrointestinal, genitourinary, and bone marrow toxicity [7-11]. Whole-pelvic step-and-shoot (SaS) IMRT for cervical cancer patients has been used in our center since 2007. A newer technique for delivering IMRT, helical tomotherapy™ (HT) (using a continuously rotating beam), can provide highly conformal dose distributions and simultaneous critical organ-sparing ability [12, 13]. It is being studied for use in gynecologic malignancies and thus far has provided promising results with respect to reliable set-up accuracy and a reduction in planning margin [14]. In our center, SaS-IMRT started in the year 2007, and HT (TomoTherapy Hi-Art with treatment planning system (TPS) Hi- Art version 4.2.1) commenced in 2012. We use both techniques in our practice. This is the first dosimetric comparison study of the two techniques in WPRT for locally advanced cervical cancer in our country.

MATERIALS AND METHODS

A total of 20 cervical cancer patients who received whole-pelvis irradiation with the SaS-IMRT technique at the Division of Therapeutic Radiology and Oncology, Chiang Mai University, between January 2011 and January 2014 were retrospectively identified. For IMRT planning, a computed tomography (CT) simulation was performed. Vac-Loc was used to immobilize patients in position. To prepare the bladder, patients were advised to urinate 20 min before scanning and to drink 200 ml of water after voiding. Patients were then set up in CT. For rectal preparation, a laxative was administered in case of rectal dilatation. This protocol was used during irradiation to maintain bladder and rectal volume. With the patients in supine treatment position and legs relaxed on the table, the pelvic region from the L1–L2 interspace and covering the whole vagina was scanned without intravenous contrast to obtain appropriate images. The CT slice thickness was 5 mm (no interslice gap). In our center, we have only one 12-year-old CT-simulator unit, which has a single slice thickness for IMRT planning (Toshiba: Asteion). To preserve the machine, a 5-mm slice thickness was selected (for a large volume site such as for pelvic irradiation). After imaging was completed, image data was sent for contouring (Oncentra Masterplan®, Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) and planning (KonRad treatment planning software®, Siemens, Concord, CA, USA). The RTOG/JCOG recommendations were used in combination as a guide for contouring the clinical target volume (CTV) [15-18]. The CTV was composed of the cervix, uterus, adnexae, upper half of the vagina, and pelvic lymph nodes (common iliac lymph nodes (LNs), external iliac LNs, internal iliac LNs, obturator LNs and presacral LNs). The PTV was defined as the CTV plus a 0.7-cm margin (for the pelvic lymph nodes) and a 1–1.5-cm margin (for the primary cervical tumor). The bladder, rectum, sigmoid colon, small bowel and heads of femurs were contoured as organs at risk (OARs). For the PTV, the dose of 1.8–2 Gy per fraction at five fractions per week was prescribed (i.e. a total dose of 45–46 Gy). The D95 (dose to 95% of the volume) of the PTV was calculated. Additional to the doses to 5% of the bladder, rectum and small bowel, the maximal doses to the heads of the femurs were also evaluated. As described earlier, patients were advised to prepare the bladder and rectum for treatment as for the CT simulation. An electronic portal imaging device (EPID) was used weekly to evaluate and make corrections to positioning. An SaS-IMRT plan was generated for each patient using KonRad, Siemens TPS, with seven coplanar step-and-shoot beams at fixed gantry angles G1 to G7. The G1 to G7 for most cases were 0, 40, 80, 125, 235, 280 and 320 degrees, respectively. For only six of the patients, G4 and G5 were 5-degrees modified to 120 and 240 degrees, respectively. All patients received SaS-IMRT whole-pelvis irradiation with our Siemens Primus. A HT plan was then created using the Tomotherapy treatment-planning system (HiArt, Tomotherapy, Tomotherapy Inc., Madison, WI) for every patient (by the same medical physicist who planned the SaS-IMRT). The optimization parameters of dose constraint and overlap priority were given the same consideration in both the SaS-IMRT plan and the HT plan. We used the same contouring as for SaS-IMRT for plan generation. Both plans were required to use identical planning objectives with respect to optimal PTV coverage and OAR sparing. Plans were designed using a field width of 5 cm, a pitch of 0.287 and modulation factor of 2.2 to 2.8. The mean setting and actual modulation factor were 2.235 and 2.032, respectively. Dosimetric parameters were analyzed for each technique. D50%, D95% and D5% of the PTV were selected. The priorities were prescribed as follows: PTV, bladder, rectum, sigmoid colon, small bowel, and head of femur. For organs-at-risk, D50% and D2% to the small bowel, rectum and bladder were considered. The Dmax of the bilateral heads of the femurs were also analyzed. To assess the uniformity of dose distribution in the PTV, we calculated a uniformity index (UI) from the formula UI = D5/D95, where D5 and D95 were the minimum doses delivered to 5% and 95% of the PTVs. The ideal value is 1, and it increases as the plan becomes less homogeneous. We also calculated the quality of coverage, with a conformity index (CI) that was defined as the ratio between the target volume and the target volume that was covered by the reference isodose. A CI = 1 corresponds to ideal conformation. A CI > 1 indicates that the irradiated volume is greater than the target volume. A CI < 1 indicates that the target volume is only partially irradiated. A direct comparison of the dosimetric parameters between SaS-IMRT and HT was also performed (using the Wilcoxon's matched pairs test) to determine if there was a significant difference for any of the parameters examined. Differences were considered statistically significant at P < 0.05. Additionally, beam-on timtes (BOTs) (defined as the time from first beam on until the last beam was turned off) of the two delivery systems were also evaluated and compared. The study was approved by our Institutional Review Board. All analyses were performed using SPSS version 17.

RESULTS

Patient characteristics

A total of 20 women were included, with a median age of 55.5 years (IQR, 52–63 years). All patients belonged to FIGO Stage IIB and IIIB and received concurrent chemoradiotherapy with weekly cisplatin at 40 mg/m2. All of the patients were treated with WP SaS-IMRT followed by brachytherapy or external beam radiotherapy according to the tumor status after WPRT. Baseline patient characteristics were shown in Table 1. The dosimetric data for SaS-IMRT and HT are shown in Tables 2 and 3, respectively.
Table 1.

Patient characteristics

No.AgeStageDose prescription
151IIB45 Gy/25 Fx
251IIB45 Gy/25 Fx
362IIB45 Gy/25 Fx
463IIB45 Gy/25 Fx
553IIB45 Gy/25 Fx
646IIIB45 Gy/25 Fx
752IIB45 Gy/25 Fx
857IIB45 Gy/25 Fx
952IIB45 Gy/25 Fx
1054IIB45 Gy/25 Fx
1152IIB45 Gy/25 Fx
1253IIIB45 Gy/25 Fx
1361IIB45 Gy/25 Fx
1457IIB45 Gy/25 Fx
1563IIB45 Gy/25 Fx
1673IIB46 Gy/23 Fx
1770IIB46 Gy /23 Fx
1867IIIB46 Gy /23 Fx
1953IIB46 Gy /23 Fx
2072IIB46 Gy /23 Fx
Table 2.

The data for SaS-IMRT

No.Bladder
Rectum
Head of femur
Bowel
CTV
PTV
D50 (Gy)D2 (Gy)D50 (Gy)D2 (Gy)Dmax Lt (Gy)Dmax Rt (Gy)D50 (Gy)D2 (Gy)Dmean (Gy)D98 (Gy)CIUID5 (Gy)D50 (Gy)D95 (Gy)CIUI
141.248.537.249.246.945.128.546.248.045.20.9811.10350.147.543.40.9251.155
244.947.541.347.545.245.430.544.847.845.40.9861.08449.647.744.30.9411.119
341.448.238.349.047.443.621.937.647.845.10.9811.10750.147.543.90.9381.141
442.948.937.048.844.841.928.941.548.545.90.9891.09350.348.144.80.9441.124
540.548.036.648.638.339.026.340.048.145.80.9861.09450.147.744.50.9421.128
641.148.843.550.342.340.931.246.548.746.20.9941.08450.548.344.30.9411.139
741.350.335.952.047.447.728.641.849.847.00.9991.09551.949.644.90.9491.156
840.251.040.650.848.048.024.941.849.546.20.9981.10851.849.145.00.9511.152
948.050.834.652.045.945.327.441.049.747.20.9991.09651.849.144.90.9501.153
1040.549.537.948.847.347.125.239.649.246.70.9961.09151.248.745.10.9521.136
1137.549.940.450.048.846.528.241.849.746.80.9941.10151.949.444.90.9491.155
1238.148.936.246.445.543.726.240.849.947.00.9981.10052.049.445.00.9501.156
1340.450.836.750.944.241.129.740.549.346.20.9921.11351.648.744.60.9421.157
1437.149.740.249.844.343.828.039.249.346.50.9961.09251.248.845.00.9521.137
1538.550.435.950.846.143.923.538.949.947.01.0001.09552.049.645.40.9601.146
1641.952.841.150.744.443.029.649.050.748.00.9941.10052.650.145.70.9451.152
1742.451.144.252.242.041.831.149.249.046.30.9821.11151.448.745.00.9441.142
1839.750.142.049.238.841.022.147.548.945.80.9761.09950.948.645.00.9421.131
1944.449.941.348.842.342.622.948.649.045.20.9671.11351.348.744.90.9381.144
2042.250.337.948.844.142.029.449.248.946.50.9821.09350.958.545.60.9471.117
Table 3.

The data for HT

No.Bladder
Rectum
Head of femur
Bowel
CTV
PTV
D50 (Gy)D2 (Gy)D50 (Gy)D2 (Gy)Dmax Lt (Gy)Dmax Rt (Gy)D50 (Gy)D2 (Gy)Dmean (Gy)D98 (Gy)CIUID5 (Gy)D50 (Gy)D95 (Gy)CIUI
134.345.333.946.541.141.524.145.146.145.30.9951.02846.746.044.80.9421.042
243.946.039.646.744.745.827.245.346.345.40.9921.02846.846.244.90.9491.042
336.646.839.246.744.343.521.040.446.645.40.9871.02947.346.544.80.9421.056
438.747.132.746.143.042.826.642.346.545.50.9901.02947.046.444.90.9501.047
538.246.834.045.743.842.025.941.447.246.00.9941.03447.847.145.00.9501.062
638.846.241.447.343.743.423.745.446.445.30.9891.03247.046.344.60.9401.054
741.346.833.846.745.144.426.341.146.445.70.9991.02647.046.345.10.9561.042
837.146.438.246.444.744.724.140.346.245.20.9921.02846.846.145.00.9501.040
944.846.633.746.944.743.926.040.646.445.60.9921.02747.046.345.10.9511.042
1036.746.531.546.643.743.522.741.846.445.60.9991.02546.946.244.90.9461.045
1138.347.038.647.644.344.226.640.946.745.90.9971.02747.246.644.90.9461.051
1236.146.635.946.242.042.623.744.346.345.40.9941.02746.846.144.90.9501.042
1338.746.835.346.344.043.127.241.946.545.70.9991.02647.046.344.80.9471.049
1438.446.336.246.643.642.826.041.146.345.40.9961.02746.946.244.90.9491.045
1538.146.634.146.243.043.321.040.746.045.20.9951.02546.646.044.90.9441.038
1635.547.738.84942.044.128.647.547.046.30.9911.03548.047.245.80.9481.048
1730.147.940.648.139.940.230.147.547.446.71.0001.02448.147.546.00.9511.046
1833.347.135.447.542.040.818.646.147.046.40.9971.02547.846.945.70.9421.046
1942.747.837.247.841.041.720.346.746.946.00.9851.03247.846.945.90.9491.041
2031.747.430.845.444.743.624.347.347.046.21.0001.02747.646.945.90.9461.037
Patient characteristics The data for SaS-IMRT The data for HT Dose distribution for SAS-IMRT (left) and HT (right) plan.

Target coverage

The dosimetric parameters for the HT and IMRT plans were analyzed (Table 4). Dose distribution in all SaS-IMRT and HT plans for all 20 patients satisfied clinical requirements. The dose distributions for SaS-IMRT and HT planned for one patient are shown in Fig 1. HT planning demonstrated a Dmean and D98% closer to the prescription dose than SaS-IMRT planning, with a P-value of < 0.001 and 0.008, respectively. All the PTV dosimetric parameters (D5%, D50% and D95%) of the HT plan were of significantly better quality than for the IMRT plan. The dose conformity and the CI did not differ between the plans, but the target dose UI of the HT plan was significantly better than that of the SaS-IMRT plan.
Table 4.

Wilcoxon matched pairs test between the two treatment plans

StructureDVH criteriaTomotherapy median (IQR)SaS-IMRT median (IQR)P-value
Age55.50 (52.00–63.00)
BladderD5038.15 (35.80–38.75)41.15 (39.95–42.30)<0.001
D246.80 (46.45–47.10)49.90 (48.85–50.60)<0.001
RectumD5035.65 (33.85–38.70)38.10 (36.65–41.20)<0.001
D246.65 (46.25–47.40)49.50 (48.80–50.80)<0.001
Small bowelD5025.10 (23.20–26.60)28.10 (25.05–29.50)<0.001
D242.10 (41.00–45.75)41.80 (40.25–47.00)0.823
Head of femurDmax Left43.70 (42.00–44.60)45.00 (42.75–47.20)0.021
Dmax Right43.35 (42.15–44.05)43.65 (41.82–45.38)0.225
CTVDmean46.45 (46.30–46.95)49.10 (48.60–49.70)<0.001
D9845.60 (45.40–46.00)46.25 (45.80–46.90)0.008
CI0.99 (0.99–1.00)0.99 (0.98–1.00)0.057
UI1.03 (1.03–1.03)1.10 (1.09–1.10)<0.001
PTVD547.0 (46.8–47.75)51.25 (50.35–51.88)<0.001
D5046.30 (46.20–46.90)48.70 (48.15–49.40)<0.001
D9544.90 (44.90–45.55)44.90 (44.53–45.00)<0.001
CI0.95 (9.94–0.95)0.94 (0.94–9.95)0.093
UI1.05 (1.04–1.05)1.14 (1.13–1.15)<0.001
Beam-on time230.6 (217.4–240.1)900.0 (840.0–904.0<0.001
Fig. 1.

Dose distribution for SAS-IMRT (left) and HT (right) plan.

Wilcoxon matched pairs test between the two treatment plans

OARs

HT showed better OAR dose sparing than SaS-IMRT. For the radiation dose to the bladder and rectum, the D50% and D2% of HT planning were both significantly less than for the SaS-IMRT plan (P-value < 0.001). Regarding the dose to the small bowel, D50% for the HT plan was also statistically less than for the SaS-IMRT plan (P-value < 0.001). However, we did not find a statistically significant difference in D2% for the small bowel when comparing the two techniques. The Dmax of the head of the left femur was (statistically significantly) lower in the HT plan, whereas there was no difference for the head of the right femur between the two techniques. The dose statistics of the OARs are also listed in Table 4.

Beam-on time

The median BOT was significantly shorter in HT delivery (230.6 s) than in SaS-IMRT delivery (900.0 s), with a P-value of < 0.001.

DISCUSSION

For the treatment of cervical cancer, WPRT has been used for more than five decades. The conventional two- or four-field box techniques were applied to the target, which was composed of the primary site and the pelvic LNs. Although this simple technique yielded good treatment results and acceptable toxicity, the OARs still received significant dosages, even with good blocking. With new techniques of imaging (CT or MRI) and planning, the possibility of keeping the target dose high and reducing the dose to the OARs has become a reality. After the first publication concerning the use of IMRT in gynecological cancers by Portelance et al. [7], the use of high-technology radiation therapy soon became more prevalent. Most studies have reported the benefits of IMRT in reducing the dose to OARs in radical and postoperative settings. [6, 7, 19–22]. Moreover, the supported study from Cancer Care Ontario reported comparative clinical outcomes for three dimension conformal radiotherapy (3DCRT) and IMRT. This report supported the use of IMRT in the treatment of gynecologic cancers in order to reduce the toxicity of the treatment. However, this group warned that clinicians should be aware of the uncertainties involved and be judicious in the use of gynecologic IMRT with the primary goal of reducing toxicity [23]. The development of intensity-modulated arc therapy (IMAT) was the next area of advancement in technique. The new variation of dynamic-IMRT allowed 360° treatment. HT, volumetric-modulated arc therapy (VMAT™) and RAPID arc™ are all IMAT technologies that were developed to improve treatment quality compared with conventional IMRT. Indeed, a promising trend has been observed, demonstrating IMAT provides better dose distributions and CIs with shorter BOTs when compared with conventional IMRT [24-26]. This study provides a dosimetric analysis of SaS-IMRT and HT to assess optimal treatment planning for pelvic irradiation in locally advanced cervical cancer. We performed the comparative planning studies by independently optimizing planning in the same patient. Our results indicate that HT had better conformal CTV and PTV coverage and sparing of OARs than did SaS-IMRT, as shown in Table 4. HT demonstrated a 3 Gy lower median dose at D2% of the bladder and rectum with statistically significant differences. HT did demonstrate a 1 Gy higher dose Dmean of the small bowel, which was concerning, although it was not significant statistically. HT was of clear benefit in lowering the cumulative bladder and rectal dose when compared with brachytherapy. All the PTV dosimetric parameters (D5%, D50% and D95%) for the HT plan were of significantly better quality than for the IMRT plan. Although the D98 of the CTV had a 1-Gy difference (45.6 versus 46.27; P = 0.008), the UI of HT was significantly closer to 1 than SaS-IMRT (P < 0.001) with the same CI (0.99). This demonstrates a lower overdose region for treatment by HT. After much discussion between radiation oncologists and medical physicists in our center, we have just this year changed the plan evaluation from D95 to D50 of the PTV, as per ICRU 83. So, D95 of the PTV was the main evaluation for the target, but we evaluated additional parameters as per ICRU 83 [27] recommendations. Although the median dose at D95 of the PTV had statistical significance, all plans achieved the target dose to D95 of the PTV, and HT planning showed better UI and less data deviation than SaS-IMRT. Moreover, our findings suggest that using HT will have a favorable impact on BOT. Another limitation in using pelvic SaS-IMRT for cervical cancer is organ motion, which is not evaluated prior to treatment delivery. This would be another advantage of using HT: using megavoltage CT before each treatment session, thus allowing the plan to have smaller margins as well as lower radiation exposure to nearby OARs because of improved set-up accuracy and reproducibility.

CONCLUSION

When treating cervical cancer with IMRT, HT provided a distinct advantage compared with SaS-IMRT with respect to target coverage and OARs. Moreover, it had shorter delivery treatment time.

CONFLICT OF INTEREST

The authors have declared that there are no conflicts of interest.

FUNDING

Funding by Faculty of Medicine, Chiang Mai University. Funding to pay the Open Access publication charges for this article was provided by Faculty of Medicine, Chiang Mai University.
  27 in total

1.  Consensus guidelines for delineation of clinical target volume for intensity-modulated pelvic radiotherapy for the definitive treatment of cervix cancer.

Authors:  Karen Lim; William Small; Lorraine Portelance; Carien Creutzberg; Ina M Jürgenliemk-Schulz; Arno Mundt; Loren K Mell; Nina Mayr; Akila Viswanathan; Anuja Jhingran; Beth Erickson; Jennifer De los Santos; David Gaffney; Catheryn Yashar; Sushil Beriwal; Aaron Wolfson; Alexandra Taylor; Walter Bosch; Issam El Naqa; Anthony Fyles
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-05-14       Impact factor: 7.038

2.  A consensus-based guideline defining the clinical target volume for pelvic lymph nodes in external beam radiotherapy for uterine cervical cancer.

Authors:  Takafumi Toita; Tatsuya Ohno; Yuko Kaneyasu; Takashi Uno; Ryouichi Yoshimura; Takeshi Kodaira; Kazuhisa Furutani; Goro Kasuya; Satoshi Ishikura; Toshiharu Kamura; Masahiro Hiraoka
Journal:  Jpn J Clin Oncol       Date:  2010-02-04       Impact factor: 3.019

3.  Estimates of worldwide burden of cancer in 2008: GLOBOCAN 2008.

Authors:  Jacques Ferlay; Hai-Rim Shin; Freddie Bray; David Forman; Colin Mathers; Donald Maxwell Parkin
Journal:  Int J Cancer       Date:  2010-12-15       Impact factor: 7.396

4.  Clinical outcomes of intensity-modulated pelvic radiation therapy for carcinoma of the cervix.

Authors:  Michael D Hasselle; Brent S Rose; Joel D Kochanski; Sameer K Nath; Rounak Bafana; Catheryn M Yashar; Yasmin Hasan; John C Roeske; Arno J Mundt; Loren K Mell
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-08-12       Impact factor: 7.038

5.  The emerging use of IMRT for treatment of cervical cancer.

Authors:  Christopher Loiselle; Wui-Jin Koh
Journal:  J Natl Compr Canc Netw       Date:  2010-12       Impact factor: 11.908

6.  Concurrent chemoradiotherapy with helical tomotherapy for oropharyngeal cancer: a preliminary result.

Authors:  Pei-Wei Shueng; Le-Jung Wu; Shiou-Yi Chen; Chi-Huang Hsiao; Hui-Ju Tien; Po-Wen Cheng; Ying-Shiung Kuo; Yu-Jen Chen; Chien-An Chen; Pei-Ying Hsieh; Chen-Hsi Hsieh
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-10-29       Impact factor: 7.038

7.  Clinical outcome in posthysterectomy cervical cancer patients treated with concurrent Cisplatin and intensity-modulated pelvic radiotherapy: comparison with conventional radiotherapy.

Authors:  Miao-Fen Chen; Chih-Jen Tseng; Ching-Cheng Tseng; Yuen-Chun Kuo; Chun-Yen Yu; Wen-Cheng Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-04-01       Impact factor: 7.038

8.  Long-term survival and late toxicity after chemoradiotherapy for cervical cancer--the Addenbrooke's experience.

Authors:  L T Tan; M Zahra
Journal:  Clin Oncol (R Coll Radiol)       Date:  2008-04-18       Impact factor: 4.126

9.  Total marrow irradiation with helical tomotherapy for bone marrow transplantation of multiple myeloma: first experience in Asia.

Authors:  Pei-Wei Shueng; Shih-Chiang Lin; Ngot-Swan Chong; Hsing-Yi Lee; Hui-Ju Tien; Le-Jung Wu; Chien-An Chen; Jason J S Lee; Chen-Hsi Hsieh
Journal:  Technol Cancer Res Treat       Date:  2009-02

10.  Whole pelvic helical tomotherapy for locally advanced cervical cancer: technical implementation of IMRT with helical tomotherapy.

Authors:  Chen-Hsi Hsieh; Ming-Chow Wei; Hsing-Yi Lee; Sheng-Mou Hsiao; Chien-An Chen; Li-Ying Wang; Yen-Ping Hsieh; Tung-Hu Tsai; Yu-Jen Chen; Pei-Wei Shueng
Journal:  Radiat Oncol       Date:  2009-12-10       Impact factor: 3.481

View more
  3 in total

1.  Dosimetric predictors and Lyman normal tissue complication probability model of hematological toxicity in cervical cancer patients with treated with pelvic irradiation.

Authors:  Dandan Wang; Yueju Yin; Qichao Zhou; Zirong Li; Xingmin Ma; Yong Yin; Baosheng Li; Tong Bai; Dapeng Li; Jian Zhu
Journal:  Med Phys       Date:  2021-12-10       Impact factor: 4.506

2.  Comparing efficacy of high-dose rate brachytherapy versus helical tomotherapy in the treatment of cervical cancer.

Authors:  Seongmin Kim; Sanghoon Lee; Jin Hwa Hong; Young Je Park; Jae Yun Song; Jae Kwan Lee; Nak Woo Lee
Journal:  J Gynecol Oncol       Date:  2019-12-20       Impact factor: 4.401

3.  Whole-pelvic radiotherapy with spot-scanning proton beams for uterine cervical cancer: a planning study.

Authors:  Shingo Hashimoto; Yuta Shibamoto; Hiromitsu Iwata; Hiroyuki Ogino; Hiroki Shibata; Toshiyuki Toshito; Chikao Sugie; Jun-Etsu Mizoe
Journal:  J Radiat Res       Date:  2016-07-05       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.