Literature DB >> 25715994

Comparison of Response of DRI Oxycodone Semiquantitative Immunoassay With True Oxycodone Values Determined by Liquid Chromatography Combined With Tandem Mass Spectrometry: Sensitivity of the DRI Assay at 100 ng/ml Cut-Off and Validity of Semiquantitative Value.

R Brent Dixon1, Bonnette Davis2, Amitava Dasgupta3.   

Abstract

OBJECTIVE: Oxycodone is a widely used opioid for pain management and patient's compliance with therapy is often monitored by using oxycodone immunoassay. The performance of the DRI oxycodone immunoassay was compared with liquid chromatography combined with tandem mass spectrometry (LC/MS/MS) assay.
MATERIALS AND METHODS: In 48 urine specimens collected from patients taking oxycodone, urinary oxycodone concentrations were determined using LC/MS/MS and the DRI oxycodone immunoassay for application on the Cobas c 501 analyzer (Roche Diagnostics, Indianapolis, IN).
RESULTS: Out of 48 specimens, 14 specimens showed oxycodone value less than 100 ng/ml, seven specimens had low positive values (between 101 and 165 ng/ml) and all other specimens had values 165 to 1789 ng/ml using the LC/MS/MS assay. The DRI oxycodone assay successfully identified all oxycodone specimens with oxycodone concentrations over the 100 ng/ml. In addition, the DRI assay also showed positive response in 11 out of 14 specimens with oxycodone values less than 100 ng/ml. However, semiquantitative values obtained by the DRI assay did not match with true oxycodone and metabolite oxymorphone concentrations combined obtained by using LC/MS/MS.
CONCLUSIONS: DRI oxycodone immunoassay at 100 ng/ml is a reliable immunoassay for analysis of oxycodone in urine.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  100 ng/ml cut-off; DRI oxycodone; LC/MS/MS

Mesh:

Substances:

Year:  2015        PMID: 25715994      PMCID: PMC6807119          DOI: 10.1002/jcla.21834

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  14 in total

Review 1.  Advanced urine toxicology testing.

Authors:  Peter L Tenore
Journal:  J Addict Dis       Date:  2010-10

Review 2.  Safety and efficacy of controlled-release oxycodone: a systematic literature review.

Authors:  D Gary Rischitelli; Sean H Karbowicz
Journal:  Pharmacotherapy       Date:  2002-07       Impact factor: 4.705

3.  Can an immunoassay become a standard technique in detecting oxycodone and its metabolites?

Authors:  Jude M Abadie; Kim H Allison; David A Black; James Garbin; Andrew J Saxon; Daniel D Bankson
Journal:  J Anal Toxicol       Date:  2005 Nov-Dec       Impact factor: 3.367

4.  Validation of an automated solid-phase extraction method for the analysis of 23 opioids, cocaine, and metabolites in urine with ultra-performance liquid chromatography-tandem mass spectrometry.

Authors:  María del Mar Ramírez Fernández; Filip Van Durme; Sarah M R Wille; Vincent di Fazio; Natalie Kummer; Nele Samyn
Journal:  J Anal Toxicol       Date:  2014-04-30       Impact factor: 3.367

5.  Vital signs: overdoses of prescription opioid pain relievers---United States, 1999--2008.

Authors: 
Journal:  MMWR Morb Mortal Wkly Rep       Date:  2011-11-04       Impact factor: 17.586

6.  Evaluation of the usefulness of an oxycodone immunoassay in combination with a traditional opiate immunoassay for the screening of opiates in urine.

Authors:  Marie Gingras; Marie-Hélène Laberge; Michel Lefebvre
Journal:  J Anal Toxicol       Date:  2010-03       Impact factor: 3.367

7.  American Society of Interventional Pain Physicians (ASIPP) guidelines for responsible opioid prescribing in chronic non-cancer pain: Part 2--guidance.

Authors:  Laxmaiah Manchikanti; Salahadin Abdi; Sairam Atluri; Carl C Balog; Ramsin M Benyamin; Mark V Boswell; Keith R Brown; Brian M Bruel; David A Bryce; Patricia A Burks; Allen W Burton; Aaron K Calodney; David L Caraway; Kimberly A Cash; Paul J Christo; Kim S Damron; Sukdeb Datta; Timothy R Deer; Sudhir Diwan; Ike Eriator; Frank J E Falco; Bert Fellows; Stephanie Geffert; Christopher G Gharibo; Scott E Glaser; Jay S Grider; Haroon Hameed; Mariam Hameed; Hans Hansen; Michael E Harned; Salim M Hayek; Standiford Helm; Joshua A Hirsch; Jeffrey W Janata; Alan D Kaye; Adam M Kaye; David S Kloth; Dhanalakshmi Koyyalagunta; Marion Lee; Yogesh Malla; Kavita N Manchikanti; Carla D McManus; Vidyasagar Pampati; Allan T Parr; Ramarao Pasupuleti; Vikram B Patel; Nalini Sehgal; Sanford M Silverman; Vijay Singh; Howard S Smith; Lee T Snook; Daneshvari R Solanki; Deborah H Tracy; Ricardo Vallejo; Bradley W Wargo
Journal:  Pain Physician       Date:  2012-07       Impact factor: 4.965

8.  Laboratory monitoring of OxyContin (oxycodone): clinical pitfalls.

Authors:  Randal L Von Seggern; Charles P Fitzgerald; Leon C Adelman; James U Adelman
Journal:  Headache       Date:  2004-01       Impact factor: 5.887

9.  Optimizing urine drug testing for monitoring medication compliance in pain management.

Authors:  Stacy E F Melanson; Adam S Ptolemy; Ajay D Wasan
Journal:  Pain Med       Date:  2013-07-30       Impact factor: 3.750

Review 10.  Oxycodone: new 'old' drug.

Authors:  Klaus T Olkkola; Nora M Hagelberg
Journal:  Curr Opin Anaesthesiol       Date:  2009-08       Impact factor: 2.706

View more