Ivalina Hristova1, Ronald Boellaard, Wouter Vogel, Felix Mottaghy, Sandrine Marreaud, Sandra Collette, Patrick Schöffski, Roberta Sanfilippo, Raz Dewji, Winette van der Graaf, Wim J G Oyen.
Abstract
PURPOSE: (18)F-Labelled fluorodeoxyglucose (FDG) can detect early changes in tumour metabolism and may be a useful quantitative imaging biomarker (QIB) for prediction of disease stabilization, response and duration of progression-free survival (PFS). Standardization of imaging procedures is a prerequisite, especially in multicentre clinical trials. In this study we reviewed the quality of FDG scans and compliance with the imaging guideline (IG) in a phase III clinical trial.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25711176 PMCID: PMC4382532 DOI: 10.1007/s00259-015-3002-0
Source DB: PubMed Journal: Eur J Nucl Med Mol Imaging ISSN: 1619-7070 Impact factor: 9.236
Fig. 1Representative artefacts
Fig. 2a UT between FDG injection and start of the PET scan, based on DICOM tags b UT between FDG injection and start of the PET scan, based on ITFs
Reconstruction parameters
| Site code | PET slice thickness (mm) | Scan direction | Scanner | FDG dose (MBq) | Time per bed (s) | Reconstruction method |
|---|---|---|---|---|---|---|
| A | 2 | Cranial | Siemens Biograph 16 | 272–378 | 180 | OSEM2D 5i8s |
| B | 4 | Cranial | Philips Allegro Body© | 312–400 | 180 | 3D-RAMLA |
| B | 4 | Cranial | Philips GEMINI TF TOF 16 | 192–195 | 120 | 3D-RAMLA |
| C | 3.27 | Cranial | GE Discovery ST | 221–258 | 180 | 3D FORE IR |
| D | 5 | Cranial | Philips GEMINI TF© | 191 | 90 | BLOB-OS-TF |
| D | 5 | Cranial | Philips GEMINI TF TOF 16 | 164–238 | 60–90 | BLOB-OS-TF |
| E | 3.38 | Cranial | Siemens Biograph Duo | 202–370 | 240 | OSEM2D 2i8s |
| F | 4 | Cranial | Philips Allegro Body(C) | 349–369 | 120 | 3D-RAMLA |
| F | 4 | Cranial | Philips GEMINI TF© | 349–371 | 120 | BLOB-OS-TF |
| F | 4 | Cranial | Philips GEMINI TF TOF 16 | 369–407 | 120 | BLOB-OS-TF |
| G | 2.43 | Cranial | Siemens Biograph Duo | 275–343 | 210 | OSEM 2i8s |
| H | 3.27 | Cranial | GE Discovery STE | 379 | 150 | 3D IR |
| I | 4.25 | Cranial | GE Discovery LS | 323–381 | 86–240 | OSEM |
| I | 4.25 | Caudal | GE Advance | 331–381 | 240 | Unknown |
| J | 4 | Caudal/cranial | Philips GEMINI TF© | 235–291 | 120 | BLOB-OS-TF |
| K | 2 | Cranial | Siemens Biograph 16 | 706–812 | 240 | OSEM2D 4i8s |
| K | 3.27 | Cranial | GE Discovery ST | 645–813 | 300 | OSEM |
Fig. 3Liver SUVmean in normal liver tissue a UT was according to the IG for both FDG scans: 60 ± 5 min (11 patients). b UT of FU FDG scan was within the UT of the actual BL scan ± 10 min, irrespective of the absolute UT of the actual BL scan (12 patients). c UT outside the specifications of the IG and more than 10 min difference between BL and FU scan (11 patients)