Accuracy and reproducibility of a quantitative magnetic resonance imaging method for concurrent measurements of tissue relaxation times and proton density.

Wolfgang Krauss1, Martin Gunnarsson2, Torbjörn Andersson3, Per Thunberg4.   

Abstract

PURPOSE: To evaluate the accuracy and reproducibility of a quantitative magnetic resonance (qMR) imaging method (QRAPMASTER) for simultaneous measurements of T1 and T2 relaxation times, and proton density (PD).
MATERIALS AND METHODS: Measurements of T1, T2, and PD with qMR were performed using phantoms with different relaxation times and concentrations of heavy water. Healthy volunteers were examined with different head coils. Regional measurements were performed in normal-appearing white and gray matter from the healthy control subjects, and in multiple sclerosis (MS) patients.
RESULTS: In phantom measurements, QRAPMASTER slightly underestimated T1, and T2 variations between repeated measurements were modest. PD was generally overestimated. The overall relative difference was -1.2±5.3% (T1), -6.6±1.9% (T2), and 0.7±5.1% (PD). In healthy volunteers, there were no statistically significant differences of T1, T2 or PD using different head coils. Values of T1, T2, and PD obtained in healthy controls and MS patients were within reference ranges. However, significant differences were found in normal-appearing gray and white matter.
CONCLUSION: QRAPMASTER can be considered a sufficiently accurate and reproducible method for use in clinical practice. Neuropathology in normal-appearing brain tissue may be revealed using this MR method, with putative implications for quantification of tissue damage in neurological diseases.
Copyright © 2015. Published by Elsevier Inc.

Entities:  

Keywords:  Multiple sclerosis; PD mapping; Quantitative MRI; Synthetic MRI; T1 mapping; T2 mapping

Mesh:

Substances:

Year:  2015        PMID: 25708264     DOI: 10.1016/j.mri.2015.02.013

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  25 in total

1.  Synthetic MRI for Clinical Neuroimaging: Results of the Magnetic Resonance Image Compilation (MAGiC) Prospective, Multicenter, Multireader Trial.

Authors:  L N Tanenbaum; A J Tsiouris; A N Johnson; T P Naidich; M C DeLano; E R Melhem; P Quarterman; S X Parameswaran; A Shankaranarayanan; M Goyen; A S Field
Journal:  AJNR Am J Neuroradiol       Date:  2017-04-27       Impact factor: 3.825

2.  Image quality at synthetic brain magnetic resonance imaging in children.

Authors:  So Mi Lee; Young Hun Choi; Jung-Eun Cheon; In-One Kim; Seung Hyun Cho; Won Hwa Kim; Hye Jung Kim; Hyun-Hae Cho; Sun-Kyoung You; Sook-Hyun Park; Moon Jung Hwang
Journal:  Pediatr Radiol       Date:  2017-06-22

3.  Brain tissue and myelin volumetric analysis in multiple sclerosis at 3T MRI with various in-plane resolutions using synthetic MRI.

Authors:  Laetitia Saccenti; Christina Andica; Akifumi Hagiwara; Kazumasa Yokoyama; Mariko Yoshida Takemura; Shohei Fujita; Tomoko Maekawa; Koji Kamagata; Alice Le Berre; Masaaki Hori; Nobutaka Hattori; Shigeki Aoki
Journal:  Neuroradiology       Date:  2019-06-18       Impact factor: 2.804

4.  Changes in the ADC of diffusion-weighted MRI with the oscillating gradient spin-echo (OGSE) sequence due to differences in substrate viscosities.

Authors:  Tomoko Maekawa; Masaaki Hori; Katsutoshi Murata; Thorsten Feiweier; Issei Fukunaga; Christina Andica; Akifumi Hagiwara; Koji Kamagata; Saori Koshino; Osamu Abe; Shigeki Aoki
Journal:  Jpn J Radiol       Date:  2018-04-26       Impact factor: 2.374

5.  Rapid multicontrast brain imaging on a 0.35T MR-linac.

Authors:  Siamak P Nejad-Davarani; Niloufar Zakariaei; Yongsheng Chen; E Mark Haacke; Newton J Hurst; M Salim Siddiqui; Lonni R Schultz; James M Snyder; Tobias Walbert; Carri K Glide-Hurst
Journal:  Med Phys       Date:  2020-07-06       Impact factor: 4.071

6.  Initial experience with synthetic MRI of the knee at 3T: comparison with conventional T1 weighted imaging and T2 mapping.

Authors:  Sunghoon Park; Kyu-Sung Kwack; Young Ju Lee; Sung-Min Gho; Hyun Young Lee
Journal:  Br J Radiol       Date:  2017-11-16       Impact factor: 3.039

7.  Synthetic MRI in the Detection of Multiple Sclerosis Plaques.

Authors:  A Hagiwara; M Hori; K Yokoyama; M Y Takemura; C Andica; T Tabata; K Kamagata; M Suzuki; K K Kumamaru; M Nakazawa; N Takano; H Kawasaki; N Hamasaki; A Kunimatsu; S Aoki
Journal:  AJNR Am J Neuroradiol       Date:  2016-12-08       Impact factor: 3.825

8.  Quantification of brown adipose tissue in vivo using synthetic magnetic resonance imaging: an experimental study with mice model.

Authors:  Mengjuan Huo; Junzhao Ye; Zhi Dong; Huasong Cai; Meng Wang; Guoping Yin; Long Qian; Zi-Ping Li; Bihui Zhong; Shi-Ting Feng
Journal:  Quant Imaging Med Surg       Date:  2022-01

9.  Clinical Feasibility of Synthetic MRI in Multiple Sclerosis: A Diagnostic and Volumetric Validation Study.

Authors:  T Granberg; M Uppman; F Hashim; C Cananau; L E Nordin; S Shams; J Berglund; Y Forslin; P Aspelin; S Fredrikson; M Kristoffersen-Wiberg
Journal:  AJNR Am J Neuroradiol       Date:  2016-01-21       Impact factor: 3.825

10.  Whole brain 3D MR fingerprinting in multiple sclerosis: a pilot study.

Authors:  Thomaz R Mostardeiro; Ananya Panda; Norbert G Campeau; Robert J Witte; Nicholas B Larson; Yi Sui; Aiming Lu; Kiaran P McGee
Journal:  BMC Med Imaging       Date:  2021-05-22       Impact factor: 1.930

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.