Neumann H1, Schulz A P2, Breer S1, Unger A3, Kienast B4.
Abstract
BACKGROUND: Osteochondral injuries, if not treated appropriately, often lead to severe osteoarthritis of the affected joint. Without refixation of the osteochondral fragment, human cartilage only repairs these defects imperfectly. All existing refixation systems for chondral defects have disadvantages, for instance bad MRI quality in the postoperative follow-up or low anchoring forces. To address the problem of reduced stability in resorbable implants, ultrasound-activated pins were developed. By ultrasound-activated melting of the tip of these implants a higher anchoring is assumed. Aim of the study was to investigate, if ultrasound-activated pins can provide a secure refixation of osteochondral fractures comparing to conventional screw and conventional, resorbable pin osteosynthesis. CT scans and scanning electron microscopy should proovegood refixation results with no further tissue damage by the melting of the ultrasound-activated pins in comparison to conventional osteosynthesis.Entities:
Keywords: Bioresorbable implant; CT; osteochondral fracture; scanning electron microscope; sheep study; ultrasound-activated pin.
Year: 2015 PMID: 25674184 PMCID: PMC4321208 DOI: 10.2174/1874325001509010007
Source DB: PubMed Journal: Open Orthop J ISSN: 1874-3250
Arithmetic mean margin of lysis and the osteotomy gap of different refixation systems by CT scan including standard deviation.
| Implant | Margin of Lysis Axial (mm) | Margin of Lysis Multiplanar (mm) | Osteotomy Gap Axial (mm) | Osteotomy Gap Multiplanar (mm) |
|---|---|---|---|---|
| SonicPin™ | 1.1 ± 1.1 | 0.8 ± 0.49 | 0.5 ± 0.6 | 0.2 ± 0.5 |
| Ethipin® | 0.2 ± 0.27 | 0.2 ± 0.3 | 1.1 ± 1.4 | 0.8 ± 2.0 |
| Screws | 1.5 ± 1.19 | 1.7 ±1.19 | 0.1 ± 0.4 | 0.0 ± 0.0 |
Arithmetic mean p-Values of statistical comparison of the margin of lysis by CT-scans including standard deviation.
| SonicPin™ | Ethipin® | Asnis™-Screws | |
|---|---|---|---|
| SonicPin™ | -- | p=0.0064 (axial) | p=0.12 (axial) |
| Ethipin® | p = 0.0064 (axial) | -- | p = 0.0075 (axial) |
| Asnis™-Screws | p = 0.12 (axial) | p = 0.0075 (axial) | -- |
Arithmetic mean fragment position and tibial surface rated in points of different refixation systems by CT scan including standard deviation.
| Implant | Fragment Position Axial | Fragment Position Multiplanar | Tibial Surface Axial | Tibial Surface Multiplanar |
|---|---|---|---|---|
| SonicPin™ | 3.21 ± 1.8 | 3.21 ± 1.7 | 4.21 ± 1.3 | 3.93 ± 1.5 |
| Etipin® | 3.00 ± 2.2 | 2.67 ± 1.9 | 4.33 ± 1.6 | 4.17 ± 1.6 |
| Screws | 4.13 ± 1.4 | 3.75 ± 1.3 | 3.5 ± 1.7 | 3.25 ± 1.6 |
Arithmetic mean bone densitiy measured in Hounsfield-units.
| Overall | Overall | SonicPin™ | SonicPin™ | Etipin® | Etipin® | Screw | Screw | |
|---|---|---|---|---|---|---|---|---|
| HU | 820 | 636 | 829 | 625 | 653 | 547 | 931 | 674 |
| + 28% | + 32% | + 19% | + 38% |