Literature DB >> 25652476

Ultralow dose computed tomography attenuation correction for pediatric PET CT using adaptive statistical iterative reconstruction.

Samuel L Brady1, Barry L Shulkin2.   

Abstract

PURPOSE: To develop ultralow dose computed tomography (CT) attenuation correction (CTAC) acquisition protocols for pediatric positron emission tomography CT (PET CT).
METHODS: A GE Discovery 690 PET CT hybrid scanner was used to investigate the change to quantitative PET and CT measurements when operated at ultralow doses (10-35 mA s). CT quantitation: noise, low-contrast resolution, and CT numbers for 11 tissue substitutes were analyzed in-phantom. CT quantitation was analyzed to a reduction of 90% volume computed tomography dose index (0.39/3.64; mGy) from baseline. To minimize noise infiltration, 100% adaptive statistical iterative reconstruction (ASiR) was used for CT reconstruction. PET images were reconstructed with the lower-dose CTAC iterations and analyzed for: maximum body weight standardized uptake value (SUVbw) of various diameter targets (range 8-37 mm), background uniformity, and spatial resolution. Radiation dose and CTAC noise magnitude were compared for 140 patient examinations (76 post-ASiR implementation) to determine relative dose reduction and noise control.
RESULTS: CT numbers were constant to within 10% from the nondose reduced CTAC image for 90% dose reduction. No change in SUVbw, background percent uniformity, or spatial resolution for PET images reconstructed with CTAC protocols was found down to 90% dose reduction. Patient population effective dose analysis demonstrated relative CTAC dose reductions between 62% and 86% (3.2/8.3-0.9/6.2). Noise magnitude in dose-reduced patient images increased but was not statistically different from predose-reduced patient images.
CONCLUSIONS: Using ASiR allowed for aggressive reduction in CT dose with no change in PET reconstructed images while maintaining sufficient image quality for colocalization of hybrid CT anatomy and PET radioisotope uptake.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25652476     DOI: 10.1118/1.4905045

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Guidelines on nuclear medicine imaging in neuroblastoma.

Authors:  Zvi Bar-Sever; Lorenzo Biassoni; Barry Shulkin; Grace Kong; Michael S Hofman; Egesta Lopci; Irina Manea; Jacek Koziorowski; Rita Castellani; Ariane Boubaker; Bieke Lambert; Thomas Pfluger; Helen Nadel; Susan Sharp; Francesco Giammarile
Journal:  Eur J Nucl Med Mol Imaging       Date:  2018-10       Impact factor: 9.236

Review 2.  Operational and Dosimetric Aspects of Pediatric PET/CT.

Authors:  Frederic H Fahey; Alison Goodkind; Robert D MacDougall; Leah Oberg; Sonja I Ziniel; Richard Cappock; Michael J Callahan; Neha Kwatra; S Ted Treves; Stephan D Voss
Journal:  J Nucl Med       Date:  2017-07-07       Impact factor: 10.057

3.  A personalized deep learning denoising strategy for low-count PET images.

Authors:  Qiong Liu; Hui Liu; Niloufar Mirian; Sijin Ren; Varsha Viswanath; Joel Karp; Suleman Surti; Chi Liu
Journal:  Phys Med Biol       Date:  2022-07-13       Impact factor: 4.174

4.  Dynamic cardiac PET imaging: Technological improvements advancing future cardiac health.

Authors:  Grant T Gullberg; Uttam M Shrestha; Youngho Seo
Journal:  J Nucl Cardiol       Date:  2018-01-31       Impact factor: 5.952

  4 in total