PURPOSE: To investigate proton density (PD)-weighted zero TE (ZT) imaging for morphological depiction and segmentation of cranial bone structures.
METHODS: A rotating ultra-fast imaging sequence (RUFIS) type ZT pulse sequence was developed and optimized for 1) efficient capture of short T2 bone signals and 2) flat PD response for soft-tissues. An inverse logarithmic image scaling (i.e., -log(image)) was used to highlight bone and differentiate it from surrounding soft-tissue and air. Furthermore, a histogram-based bias-correction method was developed for subsequent threshold-based air, soft-tissue, and bone segmentation.
RESULTS: PD-weighted ZT imaging in combination with an inverse logarithmic scaling was found to provide excellent depiction of cranial bone structures. In combination with bias correction, also excellent segmentation results were achieved. A two-dimensional histogram analysis demonstrates a strong, approximately linear correlation between inverse log-scaled ZT and low-dose CT for Hounsfield units (HU) between -300 HU and 1,500 HU (corresponding to soft-tissue and bone).
CONCLUSIONS: PD-weighted ZT imaging provides robust and efficient depiction of bone structures in the head, with an excellent contrast between air, soft-tissue, and bone. Besides structural bone imaging, the presented method is expected to be of relevance for attenuation correction in positron emission tomography (PET)/MR and MR-based radiation therapy planning.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  PET/MR; RUFIS; UTE; ZTE; attenuation correction; bone; segmentation; zero TE

Mesh:

Year:  2015        PMID: 25639956     DOI: 10.1002/mrm.25545

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  60 in total

1.  Zero-Echo-Time and Dixon Deep Pseudo-CT (ZeDD CT): Direct Generation of Pseudo-CT Images for Pelvic PET/MRI Attenuation Correction Using Deep Convolutional Neural Networks with Multiparametric MRI.

Authors:  Andrew P Leynes; Jaewon Yang; Florian Wiesinger; Sandeep S Kaushik; Dattesh D Shanbhag; Youngho Seo; Thomas A Hope; Peder E Z Larson
Journal:  J Nucl Med       Date:  2017-10-30       Impact factor: 10.057

2.  Rapid dual-RF, dual-echo, 3D ultrashort echo time craniofacial imaging: A feasibility study.

Authors:  Hyunyeol Lee; Xia Zhao; Hee Kwon Song; Rosaline Zhang; Scott P Bartlett; Felix W Wehrli
Journal:  Magn Reson Med       Date:  2018-12-18       Impact factor: 4.668

3.  Evaluation of Sinus/Edge-Corrected Zero-Echo-Time-Based Attenuation Correction in Brain PET/MRI.

Authors:  Jaewon Yang; Florian Wiesinger; Sandeep Kaushik; Dattesh Shanbhag; Thomas A Hope; Peder E Z Larson; Youngho Seo
Journal:  J Nucl Med       Date:  2017-05-04       Impact factor: 10.057

4.  In vivo prediction of temporomandibular joint disc thickness and position changes for different jaw positions.

Authors:  Benedikt Sagl; Martina Schmid-Schwap; Eva Piehslinger; Claudia Kronnerwetter; Michael Kundi; Siegfried Trattnig; Ian Stavness
Journal:  J Anat       Date:  2019-02-20       Impact factor: 2.610

5.  A Quantitative Evaluation of Joint Activity and Attenuation Reconstruction in TOF PET/MR Brain Imaging.

Authors:  Ahmadreza Rezaei; Georg Schramm; Stefanie M A Willekens; Gaspar Delso; Koen Van Laere; Johan Nuyts
Journal:  J Nucl Med       Date:  2019-04-12       Impact factor: 10.057

Review 6.  Machine learning in quantitative PET: A review of attenuation correction and low-count image reconstruction methods.

Authors:  Tonghe Wang; Yang Lei; Yabo Fu; Walter J Curran; Tian Liu; Jonathon A Nye; Xiaofeng Yang
Journal:  Phys Med       Date:  2020-07-29       Impact factor: 2.685

7.  Ramped hybrid encoding for improved ultrashort echo time imaging.

Authors:  Hyungseok Jang; Curtis N Wiens; Alan B McMillan
Journal:  Magn Reson Med       Date:  2015-09-18       Impact factor: 4.668

8.  A rapid and robust gradient measurement technique using dynamic single-point imaging.

Authors:  Hyungseok Jang; Alan B McMillan
Journal:  Magn Reson Med       Date:  2016-10-03       Impact factor: 4.668

Review 9.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26

Review 10.  PET-MRI for the Study of Metabolic Bone Disease.

Authors:  James S Yoder; Feliks Kogan; Garry E Gold
Journal:  Curr Osteoporos Rep       Date:  2018-12       Impact factor: 5.096

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.