| Literature DB >> 25623220 |
Jens-Uwe Voigt1, Gianni Pedrizzetti2, Peter Lysyansky3, Tom H Marwick4, Hélène Houle5, Rolf Baumann6, Stefano Pedri7, Yasuhiro Ito8, Yasuhiko Abe9, Stephen Metz10, Joo Hyun Song11, Jamie Hamilton12, Partho P Sengupta13, Theodore J Kolias14, Jan d'Hooge1, Gerard P Aurigemma15, James D Thomas16, Luigi Paolo Badano17.
Abstract
Recognizing the critical need for standardization in strain imaging, in 2010, the European Association of Echocardiography (now the European Association of Cardiovascular Imaging, EACVI) and the American Society of Echocardiography (ASE) invited technical representatives from all interested vendors to participate in a concerted effort to reduce intervendor variability of strain measurement. As an initial product of the work of the EACVI/ASE/Industry initiative to standardize deformation imaging, we prepared this technical document which is intended to provide definitions, names, abbreviations, formulas, and procedures for calculation of physical quantities derived from speckle tracking echocardiography and thus create a common standard.Keywords: Definitions; Deformation imaging; Echocardiography; Left ventricle; Myocardial; Speckle tracking; Standard; Strain; Strain rate; Two-dimensional
Mesh:
Year: 2015 PMID: 25623220 DOI: 10.1016/j.echo.2014.11.003