False detection of atrial fibrillation in children by a blood pressure monitor with atrial fibrillation detection function.

Adrian Justin Cheung1, Bernard Man Yung Cheung2.   

Abstract

A sphygmomanometer that can detect atrial fibrillation may help to identify asymptomatic patients who might benefit from anticoagulation. Its performance in young people has not been reported. In a school project measuring blood pressure in 60 normal healthy male teenagers (age range 13-18 years; mean±SD 15.0±1.5 years), a Microlife BPA200 blood pressure monitor indicated atrial fibrillation in 11 participants (18%). These participants did not have any personal or family history of heart disease. They had sinus arrhythmia and had significantly lower systolic and diastolic blood pressure, body mass index and waist circumference (all p values <0.05). In young people, atrial fibrillation is very uncommon and false positives are likely. To avoid unnecessary alarm and referrals, this function is best turned off in young patients. 2015 BMJ Publishing Group Ltd.

Entities:  

Mesh:

Year:  2015        PMID: 25612757      PMCID: PMC4307065          DOI: 10.1136/bcr-2014-207974

Source DB:  PubMed          Journal:  BMJ Case Rep        ISSN: 1757-790X


  6 in total

Review 1.  Respiratory sinus arrhythmia: why does the heartbeat synchronize with respiratory rhythm?

Authors:  Fumihiko Yasuma; Jun-Ichiro Hayano
Journal:  Chest       Date:  2004-02       Impact factor: 9.410

2.  Prevalence of diagnosed atrial fibrillation in adults: national implications for rhythm management and stroke prevention: the AnTicoagulation and Risk Factors in Atrial Fibrillation (ATRIA) Study.

Authors:  A S Go; E M Hylek; K A Phillips; Y Chang; L E Henault; J V Selby; D E Singer
Journal:  JAMA       Date:  2001-05-09       Impact factor: 56.272

Review 3.  Accuracy of oscillometric blood pressure monitors for the detection of atrial fibrillation: a systematic review.

Authors:  Willem J Verberk; Peter W de Leeuw
Journal:  Expert Rev Med Devices       Date:  2012-11       Impact factor: 3.166

4.  Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation.

Authors:  G S Stergiou; N Karpettas; A Protogerou; E G Nasothimiou; M Kyriakidis
Journal:  J Hum Hypertens       Date:  2009-02-12       Impact factor: 3.012

5.  Detection of atrial fibrillation using a modified microlife blood pressure monitor.

Authors:  Joseph Wiesel; Lorenzo Fitzig; Yehuda Herschman; Frank C Messineo
Journal:  Am J Hypertens       Date:  2009-05-28       Impact factor: 2.689

6.  Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors.

Authors:  Karen Kearley; Mary Selwood; Ann Van den Bruel; Matthew Thompson; David Mant; Fd Richard Hobbs; David Fitzmaurice; Carl Heneghan
Journal:  BMJ Open       Date:  2014-05-02       Impact factor: 2.692

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.