Literature DB >> 25612257

Failed calibration of anesthesia machine due to look-alike oxygen sensors.

Andrea P Dutoit1, Susanna R Berger, Mohanad Shukry, Amir L Butt, Alberto J de Armendi.   

Abstract

Labels and medications with similar appearances have the potential to harm patients and cause delays in hospital services. We report a problem involving the Maxtec MAX-1 and MAX-11 oxygen sensors which are commonly used on anesthesia machines. These oxygen sensors have nearly identical labels which resulted in inadvertent interchanging of the sensors. The incident required the replacement of a MAX-11 sensor with a MAX-1 sensor to ensure proper functioning of the anesthesia machine. Identification of these cases can educate health care professionals of potential sources of labeling errors and safety issues and can also bring about Food and Drug Administration policy changes.

Entities:  

Year:  2014        PMID: 25612257     DOI: 10.1097/ACC.0b013e31829c3c34

Source DB:  PubMed          Journal:  A A Case Rep        ISSN: 2325-7237


  1 in total

1.  Evaluation of the Approaches of the Anaesthesiologists on Maintenance of Anaesthesia Machines.

Authors:  Uğur Altınışık; Hatice Betül Altınışık; Tuncer Şimşek; Hasan Ali Kiraz; Tuğba Doğu; Hüseyin Toman; Hasan Şahin; Mesut Erbaş
Journal:  Turk J Anaesthesiol Reanim       Date:  2016-06-01
  1 in total