Centrifugal multiplexing fixed-volume dispenser on a plastic lab-on-a-disk for parallel biochemical single-end-point assays.

Moonwoo La1, Sang Min Park1, Dong Sung Kim1.   

Abstract

In this study, a multiple sample dispenser for precisely metered fixed volumes was successfully designed, fabricated, and fully characterized on a plastic centrifugal lab-on-a-disk (LOD) for parallel biochemical single-end-point assays. The dispenser, namely, a centrifugal multiplexing fixed-volume dispenser (C-MUFID) was designed with microfluidic structures based on the theoretical modeling about a centrifugal circumferential filling flow. The designed LODs were fabricated with a polystyrene substrate through micromachining and they were thermally bonded with a flat substrate. Furthermore, six parallel metering and dispensing assays were conducted at the same fixed-volume (1.27 μl) with a relative variation of ±0.02 μl. Moreover, the samples were metered and dispensed at different sub-volumes. To visualize the metering and dispensing performances, the C-MUFID was integrated with a serpentine micromixer during parallel centrifugal mixing tests. Parallel biochemical single-end-point assays were successfully conducted on the developed LOD using a standard serum with albumin, glucose, and total protein reagents. The developed LOD could be widely applied to various biochemical single-end-point assays which require different volume ratios of the sample and reagent by controlling the design of the C-MUFID. The proposed LOD is feasible for point-of-care diagnostics because of its mass-producible structures, reliable metering/dispensing performance, and parallel biochemical single-end-point assays, which can identify numerous biochemical.

Entities:  

Year:  2015        PMID: 25610516      PMCID: PMC4297279          DOI: 10.1063/1.4905940

Source DB:  PubMed          Journal:  Biomicrofluidics        ISSN: 1932-1058            Impact factor:   2.800


  16 in total

1.  Rapid multi sample DNA amplification using rotary-linear polymerase chain reaction device (PCRDisc).

Authors:  D Sugumar; L X Kong; Asma Ismail; M Ravichandran; Lee Su Yin
Journal:  Biomicrofluidics       Date:  2012-03-14       Impact factor: 2.800

2.  Fully integrated lab-on-a-disc for simultaneous analysis of biochemistry and immunoassay from whole blood.

Authors:  Beom Seok Lee; Yang Ui Lee; Han-Sang Kim; Tae-Hyeong Kim; Jiwoon Park; Jeong-Gun Lee; Jintae Kim; Hanshin Kim; Wee Gyo Lee; Yoon-Kyoung Cho
Journal:  Lab Chip       Date:  2010-11-01       Impact factor: 6.799

3.  Integrated microfluidic compact disc device with potential use in both centralized and point-of-care laboratory settings.

Authors:  Mats Inganäs; Helene Dérand; Ann Eckersten; Gunnar Ekstrand; Ann-Kristin Honerud; Gerald Jesson; Gunnar Thorsén; Tobias Söderman; Per Andersson
Journal:  Clin Chem       Date:  2005-10       Impact factor: 8.327

4.  Capillary filling in patterned channels.

Authors:  H Kusumaatmaja; C M Pooley; S Girardo; D Pisignano; J M Yeomans
Journal:  Phys Rev E Stat Nonlin Soft Matter Phys       Date:  2008-06-17

5.  Towards plug and play filling of microfluidic devices by utilizing networks of capillary stop valves.

Authors:  B Hagmeyer; F Zechnall; M Stelzle
Journal:  Biomicrofluidics       Date:  2014-09-19       Impact factor: 2.800

6.  An enzyme-linked immunosorbent assay on a centrifugal platform using magnetic beads.

Authors:  Chih-Hsin Shih; Ho-Chin Wu; Chong-Yi Chang; Wen-Hong Huang; Yi-Feng Yang
Journal:  Biomicrofluidics       Date:  2014-09-19       Impact factor: 2.800

7.  Disposable integrated microfluidic biochip for blood typing by plastic microinjection moulding.

Authors:  Dong Sung Kim; Se Hwan Lee; Chong H Ahn; Jae Y Lee; Tai Hun Kwon
Journal:  Lab Chip       Date:  2006-04-04       Impact factor: 6.799

8.  Point-of-Care Test Equipment for Flexible Laboratory Automation.

Authors:  Won Suk You; Jae Jun Park; Sung Moon Jin; Sung Moo Ryew; Hyouk Ryeol Choi
Journal:  J Lab Autom       Date:  2014-02-04

9.  Euler force actuation mechanism for siphon valving in compact disk-like microfluidic chips.

Authors:  Yongbo Deng; Jianhua Fan; Song Zhou; Teng Zhou; Junfeng Wu; Yin Li; Zhenyu Liu; Ming Xuan; Yihui Wu
Journal:  Biomicrofluidics       Date:  2014-03-05       Impact factor: 2.800

Review 10.  Recent advances in enzyme assays.

Authors:  Jean-Philippe Goddard; Jean-Louis Reymond
Journal:  Trends Biotechnol       Date:  2004-07       Impact factor: 19.536

View more
  2 in total

1.  A portable rotating disc as blood rheometer.

Authors:  Rahul Agarwal; Arnab Sarkar; Subhechchha Paul; Suman Chakraborty
Journal:  Biomicrofluidics       Date:  2019-12-02       Impact factor: 2.800

Review 2.  Challenges in the Use of Compact Disc-Based Centrifugal Microfluidics for Healthcare Diagnostics at the Extreme Point of Care.

Authors:  Jordon Gilmore; Monsur Islam; Rodrigo Martinez-Duarte
Journal:  Micromachines (Basel)       Date:  2016-03-24       Impact factor: 2.891

  2 in total