Raquel Annoni1, Ruy Camargo Pires-Neto1.
Abstract
OBJECTIVE: To test the effectiveness of using a cuff pressure relief valve technique to maintain cuff pressure levels within the normal in vitro range (Phase 1) in patients admitted to the intensive care unit (Phase 2) and to test the reproducibility of the technique using different syringes.Entities:
Mesh:
Year: 2014 PMID: 25607265 PMCID: PMC4304464 DOI: 10.5935/0103-507X.20140056
Source DB: PubMed Journal: Rev Bras Ter Intensiva ISSN: 0103-507X
Cuff pressures observed after inflating the pilot balloon in Phase 1 of the study
| P10 | 45.2±7.5 | 54.8±10.1 | 48.2±4.3 | 47±6.1 | 59.6±9.2 | p=0.043 |
| P20 | 84±16.5 | 53.2±5 | 48.8±4.6 | 51.6±14.5 | 35±2.4 | p<0.001 |
Values are expressed as the mean±standard deviation of the five measurements performed with each syringe. P10 - maneuver performed with a 10mL syringe; P20 - maneuver performed with a 20mL syringe.
comparisons between 10mL syringes;
comparisons between 20mL syringes.
Intraclass correlation coefficient between the five tests performed with each syringe in Phase 1.
| 1 | 0.03 (-8.2-0.89) | 0.4 | 0.3 (-5.6-0.92) | 0.3 |
| 2 | -0.33 (-11.7-0.86) | 0.6 | 0.49 (-3.8-0.94) | 0.2 |
| 3 | 0.12 (-7.3-0.9) | 0.4 | 0.63 (-2.5-0.96) | 0.2 |
| 4 | 0.15 (-7-0.91) | 0.4 | 0.33 (-5.3-0.93) | 0.3 |
| 5 | 0.16 (-6.9-0.91) | 0.4 | 0.8 (-0.6-0.98) | 0.06 |
P10 - maneuver performed with a 10mL syringe; P20 - maneuver performed with a 20mL syringe. Values for the intraclass correlation coefficients and the 95% confidence intervals (95%CI).
Figure 1Comparisons between the initial cuff pressures (cmH2O) (P-initial) and the cuff pressures obtained during maneuvers with 5mL (P5), 10mL (P10), and 20mL (P20) syringes in Phase 2 of the study; in vivo experiments (individual data for all patients). * p<0.001 compared with the P-initial.
Figure 2Bland-Altman plot of the differences and means between the initial cuff pressures (P-initial) and the cuff pressures obtained during maneuvers with the 5mL (P5) (A), 10mL (P10) (B), and 20mL (P20) (C) syringes. The differences are listed on the y-axis, and the mean of the two values are on the x-axis. Bias is shown as a solid line, and the 95% limits of agreement are shown as dotted lines.
ULA - 95% upper limit of agreement (bias+1.96 * standard deviation); LLA - 95% lower limit of agreement (bias-1.96 * standard deviation).