| Literature DB >> 25574426 |
Nathan D Shemonski1, Shawn S Ahn2, Yuan-Zhi Liu1, Fredrick A South1, P Scott Carney1, Stephen A Boppart3.
Abstract
Over the years, many computed optical interferometric techniques have been developed to perform high-resolution volumetric tomography. By utilizing the phase and amplitude information provided with interferometric detection, post-acquisition corrections for defocus and optical aberrations can be performed. The introduction of the phase, though, can dramatically increase the sensitivity to motion (most prominently along the optical axis). In this paper, we present two algorithms which, together, can correct for motion in all three dimensions with enough accuracy for defocus and aberration correction in computed optical interferometric tomography. The first algorithm utilizes phase differences within the acquired data to correct for motion along the optical axis. The second algorithm utilizes the addition of a speckle tracking system using temporally- and spatially-coherent illumination to measure motion orthogonal to the optical axis. The use of coherent illumination allows for high-contrast speckle patterns even when imaging apparently uniform samples or when highly aberrated beams cannot be avoided.Keywords: (100.5090) Phase-only filters; (110.3010) Image reconstruction techniques; (110.3175) Interferometric imaging; (110.3200) Inverse scattering; (110.4280) Noise in imaging systems; (110.4500) Optical coherence tomography
Year: 2014 PMID: 25574426 PMCID: PMC4285593 DOI: 10.1364/BOE.5.004131
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732