| Literature DB >> 25569959 |
Scott Almquist, Joshua de Bever, Robb Merrill, Dennis Parker, Douglas Christensen.
Abstract
Transcranial high-intensity focused ultrasound has recently been used to noninvasively treat several types of brain disorders. However, due to the large differences in acoustic properties of skulls and the surrounding soft tissue, it can be a challenge to adequately focus an ultrasonic beam through the skull. We present a novel, fast, full-wave method of correcting the aberrations caused by the skull by phasing the elements of a phased-array transducer to create constructive interference at the target. Because the method is full-wave, it also allows for trajectory planning by determining the phases required for multiple target points with negligible additional computational costs. Experimental hydrophone scans with an ex vivo skull sample using a 256-element 1-MHz transducer show an improvement of 6.2% in peak pressure at the focus and a reduction of side-lobe pressure by a factor of 2.31. Additionally, mispositioning of the peak pressure from the intended treatment location is reduced from 2.3 to 0.5 mm.Entities:
Mesh:
Year: 2014 PMID: 25569959 DOI: 10.1109/EMBC.2014.6943591
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X