Validation of the A&D BP UA-651 device with a wide-range cuff for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010.

Elisabetta Benetti1, Claudio Fania, Paolo Palatini.   

Abstract

The aim of this study was to determine the accuracy of the A&D BP UA-651 device coupled to a wide-range cuff for home blood pressure (BP) measurement according to the International Protocol of the European Society of Hypertension. The device was evaluated in 33 patients. The mean age of the patients was 56.5±15.1 years. The mean systolic BP was 144.3±23.8 mmHg (range 88 : 196), the mean diastolic BP was 87.5±15.8 mmHg (range 38 : 132), and the mean arm circumference was 29.0±3.4 cm (range 22 : 36). The protocol requirements were followed precisely. The device passed all requirements, fulfilling the standards of the protocol. On average, the device overestimated the systolic BP by 0.7±3.4 mmHg and underestimated the diastolic BP by 0.8±3.6 mmHg. The measurement error was unrelated to the patient's arm circumference. These data show that the A&D BP UA-651 device coupled to a wide-range cuff fulfilled the requirements for validation by the International Protocol over a wide range of arm circumferences and can be recommended for clinical use in the adult population.

Entities:  

Mesh:

Year:  2015        PMID: 25536400     DOI: 10.1097/MBP.0000000000000100

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Self-Monitoring of Blood Pressure and Feed-back Using APP in TReatment of UnconTrolled Hypertension (SMART-BP): A Randomized Clinical Trial.

Authors:  Jin Joo Park; Minjae Yoon; Dong-Ju Choi; Sung-Ji Park; Sang-Ho Jo; Eung Ju Kim; Soo-Joong Kim; Sungyoung Lee
Journal:  Korean Circ J       Date:  2022-10       Impact factor: 3.101

2.  Association of a Smartphone Application With Medication Adherence and Blood Pressure Control: The MedISAFE-BP Randomized Clinical Trial.

Authors:  Kyle Morawski; Roya Ghazinouri; Alexis Krumme; Julie C Lauffenburger; Zhigang Lu; Erin Durfee; Leslie Oley; Jessica Lee; Namita Mohta; Nancy Haff; Jessie L Juusola; Niteesh K Choudhry
Journal:  JAMA Intern Med       Date:  2018-06-01       Impact factor: 21.873

3.  Validation of the RisingSun RS-651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060-2:2013 Standard.

Authors:  Jin She; Xizhou Guan; Yanyong Liu; Haiyan Xiang
Journal:  J Clin Hypertens (Greenwich)       Date:  2016-06-07       Impact factor: 3.738

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.