Variable Radio Frequency Proton-Electron Double-Resonance Imaging (VRF PEDRI) enables extracting a functional map from a limited number of images acquired at pre-selected EPR frequencies using specifically designed paramagnetic probes with high quality spatial resolution and short acquisition times. In this work we explored potential of VRF PEDRI for pH mapping of aqueous samples using recently synthesized pH-sensitive phosphonated trityl radical, pTR. The ratio of Overhauser enhancements measured at each pixel at two different excitation frequencies corresponding to the resonances of protonated and deprotonated forms of pTR probe allows for a pH map extraction. Long relaxation times of pTR allow for pH mapping at EPR irradiation power as low as 1.25 W during 130 s acquisition time with spatial resolution of about 1 mm. This is particularly important for in vivo applications enabling one to avoid sample overheating by reducing RF power deposition.Entities:
Keywords: Overhauser-enhanced magnetic resonance imaging; PH mapping; Proton-Electron Double-Resonance Imaging; Trityl radical
Year: 2014 PMID: 25530673 PMCID: PMC4268155 DOI: 10.1007/s00723-014-0570-2
Source DB: PubMed Journal: Appl Magn Reson ISSN: 0937-9347 Impact factor: 0.831