W H Ma1, Y J Liu1, W Wang1, Y Z Zhang1. 1. The Provincial Key Laboratory for Orthopedic Biomechanics of Hebei, The Third Hospital of Hebei Medical University, Shijiazhuang, Hebei Province, China.
Abstract
Magnesium and its alloys have recently been used in the development of lightweight, biodegradable implant materials. However, the corrosion properties of magnesium limit its clinical application. The purpose of this study was to comprehensively evaluate the degradation behavior and biomechanical properties of magnesium materials treated with micro-arc oxidation (MAO), which is a new promising surface treatment for developing corrosion resistance in magnesium, and to provide a theoretical basis for its further optimization and clinical application. The degradation behavior of MAO-treated magnesium was studied systematically by immersion and electrochemical tests, and its biomechanical performance when exposed to simulated body fluids was evaluated by tensile tests. In addition, the cell toxicity of MAO-treated magnesium samples during the corrosion process was evaluated, and its biocompatibility was investigated under in vivo conditions. The results of this study showed that the oxide coating layers could elevate the corrosion potential of magnesium and reduce its degradation rate. In addition, the MAO-coated sample showed no cytotoxicity and more new bone was formed around it during in vivo degradation. MAO treatment could effectively enhance the corrosion resistance of the magnesium specimen and help to keep its original mechanical properties. The MAO-coated magnesium material had good cytocompatibility and biocompatibility. This technique has an advantage for developing novel implant materials and may potentially be used for future clinical applications.
Magnesium and its alloys have recently been used in the development of lightweight, biodegradable implant materials. However, the corrosion properties of magnesium limit its clinical application. The purpose of this study was to comprehensively evaluate the degradation behavior and biomechanical properties of magnesium materials treated with micro-arc oxidation (MAO), which is a new promising surface treatment for developing corrosion resistance in magnesium, and to provide a theoretical basis for its further optimization and clinical application. The degradation behavior of MAO-treated magnesium was studied systematically by immersion and electrochemical tests, and its biomechanical performance when exposed to simulated body fluids was evaluated by tensile tests. In addition, the cell toxicity of MAO-treated magnesium samples during the corrosion process was evaluated, and its biocompatibility was investigated under in vivo conditions. The results of this study showed that the oxide coating layers could elevate the corrosion potential of magnesium and reduce its degradation rate. In addition, the MAO-coated sample showed no cytotoxicity and more new bone was formed around it during in vivo degradation. MAO treatment could effectively enhance the corrosion resistance of the magnesium specimen and help to keep its original mechanical properties. The MAO-coated magnesium materialhad good cytocompatibility and biocompatibility. This technique has an advantage for developing novel implant materials and may potentially be used for future clinical applications.
One desirable characteristic of an implant for bone fracture fixation is its ability to
degrade after the bone has healed. Otherwise, a second surgery is usually conducted to
remove the implant. Long-term adverse effects or even an increased risk of localinflammation may occur after long-term implantation, because the metallic implant is a
foreign body to human tissues. However, repeated surgeries not only increase the
morbidity rate of patients, but also result in increased health-care costs and longer
hospitalizations (1). To reduce such
complications, biomaterial engineers have become interested in developing biodegradable
metallic devices (2-4).Owing to their degradation properties, magnesium and its alloys are being actively
investigated as potential load-bearing orthopedic implant materials (5-8). Various
magnesium alloys have been developed, some of which have shown good biocompatibility.
However, the major obstacles of the clinical use of magnesium-based materials are its
rapid corrosion rate and the release of hydrogen gas upon degradation (9). It has been reported that the mechanical
integrity of magnesium alloy was only maintained for 6-8 weeks, with the release of
hydrogen during the corrosion process (10).
Therefore, in order to develop the corrosion resistance of magnesium, different
modification methods, such as alloying and various surface treatments, have been
introduced (9,11). For example, Witte et al. (12)
reported that magnesium alloys AZ31 and AZ91 could enhance the osteogenesis response and
increase newly formed bone. Xu et al. (13) also
reported that the Mg-Mn-Znalloy demonstrated good in vivo degradation
behavior with bone implants. However, it should be noted that most of the reported
biomedicalmagnesium alloys contain aluminum and/or rare earth elements. It is well
known that aluminum is harmful to neurons and osteoblasts, and is also associated with
dementia and Alzheimer's disease (14,15). The addition of rare earth metals such as
zirconium and cerium into the magnesium substrate may potentially be toxic to cells
(16,17)
and may lead to hepatotoxicity or have adverse effects on DNA transcription factors
(18,19). Consequently, aluminum and rare earth elements are unsuitable alloying
elements for biomedicalmagnesium materials, particularly when they are above their
normal levels. Several authors have pointed out that Ca, Mn, and Zn could be appropriate
candidates. Further research has demonstrated that Mg-Ca (4), Mg-Zn (20), and Mg-Mn-Zn
(13) alloys gradually degraded within bone and
had good biocompatibility both in vitro and in vivo;
however, the changes in the mechanical properties of these alloys during degradation
were not addressed.Apart from alloying, surface modifications to improve the corrosion properties of
magnesium alloys, such as micro-arc oxidation (MAO), ion implantation, and plasma
anodization, have been investigated (21). Of all
of these methods, MAO is a promising new surface treatment method that can improve
corrosion resistance, wear resistance, and the micro-hardness of magnesium alloys (22).Thus, in this study, oxide coating layers were synthesized on a commercially available
pure magnesium substrate using the MAO method, so as to improve biocompatibility and
reduce the degradation rate. The corrosion properties of MAO-treated magnesium were
investigated under both in vitro and in vivo
conditions and addressed the cytocompatibility and mechanical integrity of treated
samples during degradation.
Material and Methods
Sample preparation
Commercially available pure magnesium (Institute of Metal, Chinese Academy of
Sciences, China) was used in this study. Disc samples (11.3 mm in diameter and 3 mm
thick) were prepared for the electrochemical corrosion test, the immersion test, and
in vitro studies. Cylindrical rods for mechanical testing were 5
mm in diameter and 25 mm in gauge length, and, for the in vivo
animal study, the tube samples were 6 mm in external diameter, 4 mm in internal
diameter, and 5 mm in length. Prior to the MAO process, all samples were first
mechanically polished with waterproof abrasive paper (up to 1200 grits) to remove the
oxide, degreased with acetone, and then ultrasonically cleaned with ethanol and
distilled water, sequentially. Before the in vitro cytotoxicity and
in vivo degradation experiments, all samples were sterilized with
29 kGy of 60Co radiation.A MAO coating was prepared on the magnesium surface using the MAO procedure, which
was carried out with a constant current density for 10-40 min. After treatment, the
surface of the specimen appeared to be uniformly oxidized. The specimens were then
washed with distilled water and air-dried at room temperature.
In vitro degradation tests
In order to evaluate the in vitro degradation properties,
electrochemical measurements and immersion tests were performed in a standard
simulated body fluid (SBF) (23) at pH 7.4, and
the temperature was maintained at 37±0.5°C.
Electrochemical measurements
Electrochemical measurements were performed with a three-electrode system
(PARSTAT-2273; Princeton Applied Research, USA). A saturated calomel electrode was
used as reference. Potentiodynamic polarization curves were measured at a scan
rate of 1 mV/s. Electrochemical impedance spectroscopy (EIS) analysis was also
performed at open-circuit potential with a perturbing signal of 5 mV. The
frequency varied from 100 to 1 MHz, and all the EIS results were fitted and
analyzed using the Powersuit software (Agilent, USA).
Immersion tests
Immersion tests were carried out to conform with ASTM-G31-72 (24) (the ratio of surface area to solution
volume was 1 cm2:30 mL). Every 24 h, the SBF was changed to ensure that
the pH remained near physiological values. Samples were removed after 6 h, and 1,
3, 7, 14, and 30 days of immersion, rinsed with distilled water, and dried at room
temperature.After the samples had been immersed for 3 and 14 days, surface morphology was
observed using a scanning electronic microscope (Hitachi S-4800, Japan) with an
energy dispersive spectrometer (EDS; Inca-356, England), and X-ray diffraction
analysis (XRD; Bruker AXS-D8, Germany) was used to examine the composition of the
corrosion products.Finally, the samples were cleaned with chromic acid to remove the corrosion
products, and the degradation rates (in units of mm/year) were obtained according
to ASTM-G31-72. The corrosion rate is given by the equation: Corrosion
rate=KW/ADT, where the coefficient
K=8.76×104, W is the weight loss
(g), A is the sample area exposed to solution (cm2),
T is the exposure time (h), and D is the
density of the material (g/cm3) (24).The pH value of the solution was recorded during immersion tests (PHS-3C pH meter,
Leici, China), and the release of hydrogen gas during degradation was also
measured.
Mechanical properties
Tension tests were carried out with a CMT5105 universal testing machine (Shengzhen,
China), according to GB/T 228-2002 (China). The tensile samples had a gauge length of
25 mm. The samples were immersed in SBF using the same protocol as described for the
immersion test, and mechanical properties were monitored at 6 different times (6 h,
and 1, 3, 7, 14, and 30 days). A testing speed of 0.5 mm/min was used, and the yield
strengths of the coated and uncoated samples during degradation were determined and
compared.
Cytotoxicity assessments
Saos-2 cells were cultured in Dulbecco's modified Eagle's medium (DMEM; Gibco, USA),
supplemented with 10% fetalbovine serum in a humidified incubator at 95% relative
humidity and 5% CO2 at 37°C. Cytotoxicity was determined by indirect
contact. Extracts of the coated and uncoated samples were prepared according to GB/T
16886.5. The extraction media were serially diluted to 50% and 25% concentrations
after 72 h of incubation in a humidified atmosphere with 5% CO2 at
37°C.
Cell proliferation and viability
Cell proliferation and viability were measured with the
3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide (MTT) assay. Cells
were incubated in 96-well flat-bottomed cell culture plates at 2.5×104
cells/mL medium in each well and incubated for 24 h to allow cell attachment. The
medium was then replaced by 100 µL extraction medium of a different concentration.
After 24, 48, and 72 h, 20 µL of 5 mg/mL MTT solution (Sigma) was added to each
well and incubated in a humidified atmosphere of 5% CO2 at 37°C for 4
h. Subsequently, 150 µL DMSO was added to each well and absorbance measurements
were conducted at 490 nm using a Synergy H4 Hybrid Microplate Reader (Bio-Tek,
USA). DMEM was used as the negative control, and 5-fluorouracil (Qilu
Pharmaceutical Co., Ltd., China) was used as the positive control.
Fluorescent staining of cells
Cells (5×104) were seeded onto 12-well plates, incubated for 24 h, and
treated with extraction media of different concentrations. DMEM was used as the
negative control, and 5% DMSO (Sigma, USA) was used as the positive control. After
72 h, the cells were stained with Hoechst 33258 staining solution according to the
manufacturer's instructions (Beyotime, China) and then observed with a
fluorescence microscope (Imager A2, Carl Zeiss, Germany).
In vivo degradation and biocompatibility experiments
Surgery
Animal tests were approved by the Ethics Committee of the Third Hospital of Hebei
Medical University, and the in vivo degradation experiments were
performed in the animal laboratory at the hospital. A total of 12 six-month-old
male New Zealand rabbits (Laboratory Animal Center of Hebei Medical University)
with an average body weight of 2.5-2.8 kg were used. The rabbits were randomly
divided into 2 groups, and the chosen operation site was the right femoral
condyle. In the experimental group, an MAO-coated magnesium stent was used, and an
uncoated magnesium stent was implanted in the control group. Rabbits were
anesthetized with 30 mg/kg sodium pentobarbital (Shanghai Xinya Pharmaceutics
Ltd., China) by intravenous injection. The surgical site was cleaned with 0.9%
saline and 75% ethanol. After anesthesia, a tiny incision was made at the surgical
site of each rabbit. A hole 6 mm in depth and 6 mm in diameter was drilled into
the right femoral condyle, and the stent was implanted into the prepared hole. The
wound was rinsed with saline and sutured layer by layer, and then an aseptic
dressing was applied over the incision. All animals received a subcutaneous
injection of 1 mg/kg penicillin as an anti-inflammatory drug. The rabbits were
sacrificed 3 months postoperatively.
Biochemical tests
During the experiments, 5-mL blood samples were taken from the helix vessel of the
rabbits before surgery and at 1, 2, 4, 8, and 12 weeks after surgery. The blood
biochemical tests, including serum magnesium, serum creatinine (CREA), blood ureanitrogen (BUN), and alanine aminotransferase (ALT), were performed with an
automatic biochemical analyzer (Olympus AU 5400, Japan).
Radiographic and histological evaluation
Radiographs or computed tomography (CT) images were used to observe the in
vivo degradation process at 8 and 12 weeks. Meanwhile, heart, kidney,
spleen, and liver tissues from the rabbits were also inspected with hematoxylin
and eosin staining (Beyotime) 12 weeks after surgery to verify whether degradation
of the magnesiumhad harmed these important visceral organs. In addition, the bone
samples with implants were harvested and fixed in 10% formaldehyde for 3 days,
dehydrated with 70, 95, and 100% ethanol, and then embedded in methyl methacrylate
(Merck, Germany). Whole embedded samples were scanned in a micro-CT device (Inveon
Micro-CT, Siemens, Germany) to view the extent of degradation of the stents and
new bone growth. Afterwards, the embedded samples were cut into sections and then
micro-ground down to 50-70-µm thickness. The sectioned samples were stained with
methylene blue and fuchsin (Merck), and the interaction between bone and implants
was observed under a light microscope (Axio Lab. A1, Carl Zeiss).
Statistical analysis
The two-sample t-test was used to determine whether any significant
difference existed in the cytotoxicity and blood biochemical experiments. The
software package SAS (version 9.1, SAS Institute Inc., USA) was used for the
statistical analysis. Statistical significance was assigned to
P<0.05.
Results
Figure 1A shows typical polarization curves
for MAO-coated magnesium and pure magnesium. It shows that the corrosion current
(I
corr) value of MAO-coated magnesium was clearly less than that of pure
magnesium, which indicated that coated magnesium was less susceptible to
corrosion. Nyquist plots and the polarization resistance (Rp) of
samples are shown in Figure 1B: the higher
the Rp, the lower the corrosion rate (25). The Rp of MAO-coated magnesium was
higher than that of pure magnesium, which was consistent with the
I
corr data, showing that the rate of electrochemical degradation of the
coated magnesium was slower than that of pure magnesium. These results of
electrochemical tests suggested that the MAO coating could effectively protect the
magnesium metal from corrosion.
Figure 1
Electrochemical measurements. A, Polarization curves.
B. Nyquist plots of Mg and MAO-treated Mg. MAO:
micro-arc oxidation; Zre: real part of impedance; Zim: Imaginary part of
impedance.
Immersion experiments
Surface morphologies of pure magnesium and MAO-coated magnesium after 3 and 14
days of immersion are shown in Figure 2.
MAO-treated magnesium degraded more slowly than pure magnesium, which was in good
agreement with the electrochemical results. As shown in Figure 2C and D, pure magnesium samples experienced pitting
corrosion and were covered with partially protective corrosion products, but
obvious corrosion was not found on the MAO-coated magnesium samples, and there
were very small quantities of corrosion product deposits on the surface of those
samples. EDS results for the surface corrosion products on those metal samples are
illustrated in Figure 3. Also, a number of
cracks were observed on the surface of the pure magnesium samples after 14 days of
immersion (Figure 3A). The EDS results
revealed that the surface corrosion products (marked area in Figure 3) were rich in O, Mg, P, and Ca. Further XRD results
suggested that magnesium hydroxide [Mg(OH)2] and hydroxyapatite (HA)
precipitated onto the magnesium surface (Figure
4A), and magnesium oxide [MgO] precipitated onto the MAO-coated
magnesium surface (Figure 4B). Furthermore,
a strong background and broadened peaks can be observed in Figure 4A and B, which might be due to the presence of
amorphous corrosion products as indicated by Kuwahara et al. (26).
Figure 2
Microstructures of the surface of MAO-coated and uncoated magnesium
alloy after immersion for 3 and 14 days, A,
B, Uncoated and MAO-coated samples after immersion for 3
days, respectively. C, D, Uncoated and
MAO-coated samples after immersion for 14 days, respectively. MAO: micro-arc
oxidation.
Figure 3
Energy dispersive spectrometry of the corrosion products of untreated Mg
(A), and MAO-treated Mg (B) after
immersion in simulated body fluid for 14 days. MAO: micro-arc
oxidation.
Figure 4
X-ray diffraction analysis pattern of the corrosion products of
untreated Mg (A) and MAO-treated Mg (B)
after immersion in simulated body fluid for 14 days. MAO: micro-arc
oxidation.
Figure 5A shows the pH variation of
immersion tests; the pH rose rapidly (from 7.4 to 8.7) in the first 24 h and
basically stabilized after 72 h. At the end of the immersion tests (after 30
days), the pH was 9.04 for pure magnesium and 8.15 for MAO-coated magnesium.
Hydrogen evolution (Figure 5B) had a similar
trend to the pH value. In the early stage of immersion, pure magnesium reacted
with SBF acutely and a rapid generation of bubbles was observed, indicating a fast
rate of hydrogen evolution. After 48 h immersion, however, fewer bubbles appeared,
suggesting that the reaction had slowed down and that the rate of hydrogen
evolution had decreased. This could be due to corrosion films, including HA and
other phosphates, which had a protective effect and hence retarded further
degradation. However, hydrogen evolution was not found in any of the stages of
immersion in the MAO-coated samples. Figure
5C shows the corrosion rates of MAO-coated and uncoated magnesium over
time. The corrosion rate of magnesium was high in the first 24 h and subsequently
dropped rapidly, then stabilized after 7 days of immersion. However, the corrosion
rate of MAO-coated magnesium remained stable during the immersion tests. The
results of the immersion tests were in accordance with those of the
electrochemical measurements.
Figure 5
A, pH of simulated body fluid during 30 days of immersion
for Mg and MAO-coated Mg. B, Hydrogen evolution volumes of
the two implant materials over time. C, Corrosion rate of
the two implant materials over time. D, Tensile strength of
the two implant materials over time. MAO: micro-arc oxidation.
In vitro loss of mechanical integrity
The influence of in vitro degradation on the tensile strength of
both samples is demonstrated in Figure 5D. The
tensile strengths of both MAO-coated and uncoated samples were similar before
immersion (i.e., time point 0). However, the strength of the uncoated samples
decreased rapidly in the early stage of degradation, and descended slightly as the
immersion period increased. This may be due to the protective effect of the corrosion
products formed on the surface of samples during degradation. Meanwhile, the strength
of MAO-coated samples stayed constant between the 6-h and 30-day time points. The
tensile strength of the coated sample remained at approximately 165 MPa after 30 days
of immersion; however, that of the uncoated sample dropped to 140 MPa. Therefore, the
MAO coating kept magnesium at a constant tensile strength for a period of time.The viability of Saos-2 cells after 24, 48, and 72 h incubation is shown in Figure 6. There was no significant difference
between the absorbance values of cells in the extracts and those in the negative
control. As a positive control, the viability of cells incubated in 5-fluorouracil
was significantly reduced. Otherwise, cell proliferation was determined by
fluorescent staining. Figure 7A-G shows that
the morphology of Saos-2 cells cultured in different extracts after 72 h incubation
was normal and healthy, similar to that of the negative control, but the positive
control revealed many apoptotic cells, manifested as nuclear pyknosis (Figure 7H). According to ISO 10993-5:1999 (27), the cytotoxicity of these extracts was Grade
0-1. In other words, MAO-coated magnesium and uncoated magnesium samples have a level
of biosafety that is suitable for cellular applications.
Figure 6
Saos-2 cells viability after 24, 48, and 72 h incubation in the extracts of
MAO-coated and uncoated magnesium. *P<0.05, 5-fluorouracil (5-Fu) compared
to all other groups. MAO: micro-arc oxidation.
Figure 7
Fluorescent staining of Saos-2 cells after culture in the extracts of
MAO-coated and uncoated magnesium for 72 h. Panels A-G show
that the morphology of Saos-2 cells cultured in different extracts for 72 h was
normal and healthy, similar to that of the negative control, but the 5% DMSO
positive control revealed many apoptotic cells, manifested as nuclear pyknosis
(panel H). A, Negative control;
B, 100% Mg; C, 50% Mg; D,
25% Mg; E, 100% MAO-coated Mg; F, 50%
MAO-coated Mg; G, 25% MAO-coated Mg; H, 5%
DMSO positive control. MAO: micro-arc oxidation.
After implantation of MAO-treated and untreated samples, no rabbits displayed
inflammation and there were no unexpected deaths. There were no significant
differences (P>0.05) in the biochemical indicators of serum magnesium, CREA,
BUN, and ALT before and after the operation, which indicated that degradation of
the two implants also did not affect kidney and liver functions (Figure 8). These tests demonstrated good
biocompatibility of MAO-treated samples in vivo.
Figure 8
Changes in blood biochemical indicators before and after implantation.
A, Serum magnesium; B, creatinine
(CREA); C, blood urea nitrogen (BUN); D,
alanine aminotransferase (ALT). MAO: micro-arc oxidation.
Histology evaluation
Histological staining of heart, liver, kidney, and spleen showed that these
tissues were normal, which was in good agreement with the results of the
biochemical tests and also suggested good biocompatibility of the MAO-treated
implant in vivo.
Radiographic evaluation
Radiographs and CT images were taken after 8 and 12 weeks, respectively . The
untreated implant started to degrade in the first 8 weeks, as was evident from the
observation that the edge of the implant became fuzzy (Figure 9B). No adverse effects due to degradation of the metal
implant were observed in the rabbits, which was in agreement with the literature
(12). For a clearer image of the
implants, CT analysis was used to observe the degradation of the implant. The
untreated implant became blurry 12 weeks postoperatively (Figure 9D), so the radiograph and the CT image offered
evidence that the untreated implant gradually degraded within the bone. However,
all MAO-coated implants were intact throughout the entire implantation period
(Figure 9A and C), and no obvious
degradation was observed 12 weeks postoperatively.
Figure 9
Radiographs of MAO-coated magnesium (A) and uncoated
magnesium (B) implants after 8 weeks post-operation; and CT
images of MAO-coated magnesium (C) and uncoated magnesium
(D) implants after 12 weeks post-operation. MAO:
micro-arc oxidation.
Bone formation evaluation
a) Bone histology. Figure 10A and
B shows the tissue response to both MAO-coated and uncoated magnesium
implants 3 months after implantation, where new bone tissue was observed to form
around the implant. All samples showed direct contact with the newly formed bone,
and more bone was formed around the MAO-coated implants (Figure 10A) compared to the uncoated sample (Figure 10B).
Figure 10
Histological photographs of methylene blue and magenta staining of bone
tissue formed around the MAO-coated magnesium implant (A)
and the uncoated magnesium (B) implant 3 months after
implantation in the greater trochanter where arrows indicate newly formed
bone. Micro-CT 3-D reconstruction models of newly formed bone (green in
color) on both coated magnesium (C) and uncoated magnesium
(D) implants 3 months post-operation are shown. MAO:
micro-arc oxidation.
b) Micro-CT analysis. The in vivo new bone
formation during corrosion of the implant in rabbits was studied using micro-CT.
Figure 10C and D shows 3-dimensional
models of the newly formed bone on both MAO-coated and uncoated implants. The
MAO-coated magnesium sample showed more new bone growth after 3 months of
implantation (Figure 10C), which was in
accordance with the result of bone tissue histology.
Discussion
Magnesium and its alloys have recently been used in the development of lightweight,
biodegradable implant materials. However, the major obstacles to the clinical use of
magnesium alloys are its rapid degradation rate and the release of hydrogen gas upon
degradation. Therefore, different modifications have been performed on magnesium alloys,
one of which is surface modification (28-31). By conducting a suitable surface modification,
the corrosion resistance properties of magnesium alloys may be enhanced. MAO, also
called plasma electrolytic oxidation or anodic spark oxidation, is a useful anodic
oxidation technique for depositing a ceramic coating on the surface of valve metals,
such as Al, Ti, Zr, and their alloys (21,22). MAO processes are typically characterized by
the phenomenon of electrical discharge on the anode in aqueous solution. In this study,
oxide coating layers were formed on a pure magnesium substrate using MAO, in order to
enhance biocompatibility and reduce the degradation rate.A uniform MgO coating layer was fabricated on the surface of pure magnesium, and the
corrosion properties were evaluated with electrochemical and immersion tests in an SBF.
Electrochemical tests showed a significant increase in corrosion resistance for the MAO
coating. Immersion tests in SBF solution also revealed an effective reduction in
corrosion rate for the MAO-coated samples, and the pH values of the coated samples
always remained at a lower level. The coatings showed mild degradation, whereas the
uncoated sample showed relatively obvious degradation, and appreciable quantities of Ca
and P were observed on the surface of the sample after immersion in SBF.The dissolution of magnesium in aqueous solution (for instance SBF) includes anodic and
cathodic reactions (32). The existence of
chloride ions (Cl−) transforms Mg(OH)2 into soluble
MgCl2, resulting in excess OH− in the solution (3). Eventually the pH will rise. In fact, even
though the bulk solution has a pH as low as 4, the local pH near the surface of the
magnesium could be greater than 10 (33). As a
result, if the solution contains ions such as PO4
3−, Ca2+, etc., HA
[Ca10(PO4)6(OH)2] is likely to nucleate
and grow on the magnesium surface owing to the supersaturated condition at high pH
(34). This phenomenon explains the detection
of HA by XRD in this study (Figure 4). Moreover,
when magnesium ions dissolve into the solution, phosphates containing Mg/Ca form and
attach tightly to the matrix. Taking the excess OH− into consideration, some
complicated compounds [represented by
MgxCay(PO4)z(OH)] might precipitate on
the surface. Kuwahara et al. (26) also pointed
out that the corrosion products on the surface of magnesium immersed in Hank's solution
might be amorphous
(Ca0.86Mg0.14)10(PO4)6(OH)2,
a rather complicated compound. In view of the ion concentrations in the SBF used in this
study, which are similar to those in Hank's solution, there might be some amorphous
phosphates containing magnesium/calcium, as Kuwahara (26) indicated. In fact, a strong background and broadened peaks can be
observed in Figure 4, which possibly resulted from
the presence of amorphous corrosion products.The in vitro cytotoxicity of these two metal samples indicated that
they are safe to be used as implantable material. Furthermore, bone formation in the
in vivo environment was studied using micro-CT. The newly formed
bone was found around the implants of both coated and uncoated samples and no adverse
effects were found after implantation, which also proved their good biocompatibility.
Several studies (12,13) have shown that the corrosion layer containing such
magnesium-substituted calcium phosphate compounds on magnesium could promote
osteoinductivity and osteoconductivity, predicting good biocompatibility of magnesium.
However, in this study a greater amount of new bone formation was found around the
MAO-coated samples than around the uncoated samples. This may be due to the rapid
degradation and rapidly elevated ambient pH in the uncoated sample. As a consequence,
osteoblasts were unable to proliferate and adhere very well. In addition, Zreiqat (35) reported that magnesium ions could enhance the
adhesion of human bone-derived cells and increase the levels of α5β1- and β1-integrin
receptors. Nevertheless, large amounts of magnesium ions released during corrosion of
the uncoated sample possibly inactivated new bone formation (36), thereby resulting in less new bone formation around the
uncoated sample than around the MAO-coated sample. The coating on the MAO-coated sample
reduced the rate of corrosion and the level of magnesium ions, which has been reported
to enhance osteoblastic activity and to stimulate the growth of new bone tissue (1,35).
Histological analysis also revealed an area of bone formation around these two implants,
and there was an absence of inflammation and necrosis, suggesting that there were no
toxic effects in the surrounding tissues. The excess magnesium produced by degradation
could be excreted by the kidneys, without the kidney, liver, or heart disorders observed
during degradation. This correlated with serum biochemical measurements, where no
significant differences were observed between serum magnesium levels after implantation
for either the coated or the uncoated samples, which are most likely due to homeostatic
regulation by the kidney (37). This was a good
indication that the coated sample would be safe for in vivo use,
considering that, once the coating degraded, the remaining uncoated magnesium alloy
would also degrade and not induce adverse effects into the localized tissues. These
results also implied good in vivo biocompatibility of the MAO-coated
implant, which reduced the rate of magnesium ion release and allowed for homeostatic
maintenance of physiologicalmagnesium levels. Nevertheless, the mechanism of magnesium
ion absorption is still unknown, and whether magnesium ions can be metabolized or will
accumulate in certain organs also requires further detailed investigation.In this study, it was found that degradation rates of MAO-coated samples were slower
than those of pure magnesium. For both MAO-coated and uncoated samples, degradation
rates for samples immersed for 30 days were lower than for samples immersed for 3 days,
owing to the protective layer on the surface. During the early stage of immersion in
SBF, the sample degraded quickly, accompanied by rapid formation of an insoluble
protective corrosion layer that retarded degradation, and therefore the corrosion rate
of the sample slowed down over time. Under in vivo conditions, slight
corrosion occurred on the MAO-coated implant but not on the uncoated implant, which
correlated with the in vitro data. This indicated that the MAO membrane
was able to prevent metal from having direct contact with body fluid, thereby
controlling the degradation rate of magnesium metal in the in vivo
environment. However, it is difficult to determine the in vivo
degradation rate precisely, because the circumstances in vivo are quite
complicated. It should be noted that there were a number of organic components in the
in vivo environment, such as proteins and cells. However, the
in vitro testing solution was SBF, which mainly contained inorganic
ions such as Cl−, H2PO4
−, and Ca2+. Rettig and Virtanen (38) found that albumin influenced the corrosion process of magnesium alloys
in SBF. As a consequence, the in vivo degradation products are quite
complex. The different compositions of in vitro and in
vivo degradation products suggest that degradation of the magnesium alloy
will be greatly influenced by the surrounding environment. Thus, if the magnesium alloy
is used under different conditions, possible different degradation behaviors must be
taken into consideration, otherwise the rate of degradation may be estimated
incorrectly. In addition, no gas bubbles were observed on the radiographs throughout the
entire experimental period for both MAO-coated and uncoated samples, which was different
from other studies (3,20). Possible reasons included the different animal model used, the
size of the implant, and the implantation site. The lower surface-to-volume ratio of the
sample used in our study would have resulted in a low surface area exposed to the body,
thereby reducing the amount of corrosion of the implant. This could have resulted in a
decrease in the hydrogen gas release rate, and thus the gas would have been absorbed
quickly. As reported by Witte et al. (39), if the
degradation rate was slow enough, bubble formation would likely be avoided, because the
hydrogen could diffuse into the surrounding tissues.With regard to the mechanical properties of the samples, MAO treatment would contribute
to controlling the degradation process so that mechanical integrity would meet the
requirements before the bone has healed. It was found that the tensile strength of both
the coated and uncoated samples at the 0 h time point were similar, since the MAO
membrane did not affect the bulk mechanical properties of the magnesium metal. Although
the tensile strength of the uncoated sample dropped significantly due to fast corrosion
and magnesium ion release during the early time points, the rate of decrease in tensile
strength was reduced at later time points owing to the formation of a corrosion layer
(40). However, the tensile strength of the
MAO-coated samples remained at least 30 MPa higher than that of the uncoated samples
after 30 days of immersion, which resulted mainly from the slower corrosion rate of the
implant. The mechanical integrity of an orthopedic implant is very important, because
when it is used to fix fractured bones it must provide enough mechanical support to the
bone throughout the healing process. It was noteworthy that MAO treatment slowed down
the rate of corrosion of the implant and maintained its strength at about 90% for 1
month, which is suitable for orthopedic implants, because the strength of the implant
was maintained within the first month, allowing for a longer healing period for
fractures. However, further long-term studies are needed to confirm this.In summary, this study demonstrated that the MAO process is an effective method of
surface modification for magnesium alloys. The formation of an MAO coating on the
implant was shown to reduce the corrosion rate of the implant. In addition, the
mechanical properties of the MAO-coated samples were maintained during the immersion
test, in contrast to uncoated samples, which is suitable for implant applications
because of the protection of the MAO coating layer against corrosion. This is a great
advantage for the application of MAO-coated implants in orthopedic procedures, because
the slower degradation rate and the retained mechanical strength of the coated implants
could allow sufficient time for bone healing. Good cell biocompatibility was found in
in vitro cytotoxicity assessments. Moreover, animal implant
experiments indicated that there were no disorders of the heart, kidney, liver, and
spleen, and no negative effects of magnesium release were observed, indicating that
these two implants had good biocompatibility in vivo. Also, larger
amounts of new bone formation were found around the MAO-coated samples compared to the
uncoated samples. More importantly, the study showed that serum magnesium levels after
implantation remained within a normal physiological range for both the MAO-coated
implants and the uncoated implants, which indicated that, after degradation of the
polymer coating on the implant, further corrosion of the implant would not result in
cell toxicity. Nevertheless, further studies are needed to improve the properties of the
MAO membrane, such as porosity, hardness, and adhesion to the implant, and additional
long-term in vivo studies are required to further validate the use of
MAO-coated implants for orthopedic use.
Authors: Frank Witte; Frank Feyerabend; Petra Maier; Jens Fischer; Michael Störmer; Carsten Blawert; Wolfgang Dietzel; Norbert Hort Journal: Biomaterials Date: 2007-01-05 Impact factor: 12.479
Authors: Hong-Zhi Zhou; Ya-da Li; Lin Liu; Xiao-Dong Chen; Wei-Qiang Wang; Guo-Wu Ma; Yu-Cheng Su; Min Qi; Bin Shi Journal: J Huazhong Univ Sci Technolog Med Sci Date: 2017-02-22