Jelle J Halma1, Rienk Eshuis1, H Charles Vogely2, Steven M van Gaalen1, Arthur de Gast1.
Abstract
Little is known about the clinical application of highly cross-linked polyethylene (HXLPE) blended with vitamin E. This study evaluates an uncemented iso-elastic monoblock cup with vitamin E blended HXLPE. 112 patients were followed up for 2years. 95.5% completed the follow-up. The mean VAS score for patient satisfaction was 8.8 and the mean Harris Hip Score was 94.2. In 7 cases initial gaps behind the cup were observed, which disappeared completely during follow-up in 6 cases. The mean femoral head penetration rate was 0.055mm/year. No adverse reactions or abnormal mechanical behavior was observed with the short term use of vitamin E blended HXLPE. This study shows the promising performance of this cup and confirms the potential of vitamin E blended HXLPE.Entities:
Keywords: femoral head penetration; highly-crosslinked polyethylene; iso-elastic; total hip arthroplasty; uncemented monoblock cup; vitamin E blended
Mesh:
Substances:
Year: 2014 PMID: 25496930 DOI: 10.1016/j.arth.2014.11.017
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757