| Literature DB >> 25487432 |
Jonathan R Linder, Harald M Stauss, Holly Gindes, Gary L Pierce, Nicholas H Von Bergen, William G Haynes, Jess G Fiedorowicz1.
Abstract
BACKGROUND: We sought to determine whether heart rate variability (HRV), blood pressure (BP) variability, and baroreceptor-heart rate reflex sensitivity can be reliably assessed using finger volume pulse waveforms obtained from the commercially available EndoPAT device.Entities:
Mesh:
Year: 2014 PMID: 25487432 PMCID: PMC4269858 DOI: 10.1186/1471-2261-14-180
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Sociodemographic and clinical characteristics (N = 65)
| Mean (SD) | |
|---|---|
| Age | 37 (14) |
| Pack years (Smoking) | |
| Entire sample | 2.6 (7.6) |
| Smokers only | 12.4 (12.5) |
| Body mass index (kg/m2) | 27.9 (6.5) |
| Heart rate (bpm) | 70.9 (13) |
| Systolic blood pressure (mmHg) | 119.5 (13.0) |
| Diastolic blood pressure (mmHg) | 74.9 (8.6) |
| Respiratory rate (min-1) | 13.4 (3) |
| N (%) | |
| Female gender | 39 (60%) |
| White, not hispanic | 61 (93.8%) |
| Unmarried | 32 (48%) |
| Unemployed | 3 (4.6%) |
| Alcohol misuse | 31 (47%) |
| History of tobacco use | 14 (22%) |
| Heart attack | 1 (1.5%) |
| High blood pressure | 9 (13.6%) |
| Diabetes or high blood sugar | 4 (6%) |
Transfer function analysis between finometer- and EndoPAT-derived time series
| Low frequency | High frequency | |||
|---|---|---|---|---|
| Mean(SD) | Coherence | Gain | Coherence | Gain |
|
| 0.33 (0.13) | 0.90 (0.45) | 0.27 (0.08) | 0.55 (0.34) |
|
| 0.81 (0.17) | 0.94 (0.14) | 0.80 (0.17) | 1.03 (0.22) |
Intraclass correlation coefficients for primary outcomes
| Measures | Intraclass correlation coefficients between EndoPAT and finometer | ||
|---|---|---|---|
| 95% Confidence interval | |||
| ICC | Lower bound | Upper bound | |
| SDNN | 0.99 | 0.98 | 0.99 |
| RMSSD | 0.96 | 0.94 | 0.98 |
| Absolute VLF of HRV | 0.94 | 0.90 | 0.96 |
| Relative VLF of HRV | 0.94 | 0.90 | 0.96 |
| Absolute LF of HRV | 0.94 | 0.90 | 0.96 |
| Relative LF of HRV | 0.92 | 0.87 | 0.95 |
| Absolute HF of HRV | 0.97 | 0.95 | 0.98 |
| Relative HF of HRV | 0.77 | 0.65 | 0.86 |
| Absolute VLF of BPV | 0.34 | 0.11 | 0.54 |
| Relative VLF of BPV | 0.37 | 0.14 | 0.56 |
| Absolute LV of BPV | 0.43 | 0.20 | 0.61 |
| Relative LF of BPV | 0.64 | 0.47 | 0.77 |
| Absolute HF of BPV | 0.57 | 0.38 | 0.71 |
| Relative HF of BPV | 0.51 | 0.31 | 0.67 |
| Baroreflex gain | 0.27 | 0.02 | 0.49 |
Figure 1Bland-Altman plots of time domain HRV parameters derived from EndoPAT and Finometer devices. Bland-Altman comparison between EndoPAT- and Finometer-derived heart rate time series shows a high agreement between the two devices for the measurement of SDNN and RMSSD.
Figure 2Bland-Altman plots of frequency domain HRV parameters derived from EndoPAT and Finometer devices. Bland-Altman comparison of frequency domain HRV parameters between EndoPAT- and Finometer-derived time series shows a high agreement between the two devices — across VLF, LF, and HF spectral ranges for both absolute and relative HRV.
Figure 3Bland-Altman plots of spectral domain BPV parameters derived from EndoPAT and Finometer devices. Bland-Altman comparison of frequency-domain BPV parameters between EndoPAT- and Finometer-derived time series shows low agreement between the two devices — across VLF, LF, and HF spectral ranges for both absolute and relative BPV. However, relative LF and absolute HF BPV demonstrate greater accuracy relative to other BPV measures.