Insight into the mechanism of failure of the Riata lead under advisory.

Ratika Parkash1, Stanley Tung2, Jean Champagne3, Jeffrey S Healey4, Bernard Thibault5, Douglas Cameron6, Anthony Tang7, Sean Connors8, Marianne Beardsall9, Iqwal Mangat10, Felix Ayala-Paredes11, Satish Toal12, Derek Exner13, Raymond Yee7, Andrew D Krahn14.   

Abstract

BACKGROUND: Cable externalization and insulation abrasion are known to occur with the St Jude Medical Riata leads under advisory. The distribution of these abnormalities and how they relate to clinical presentation have not been well described.
OBJECTIVE: In this study, we sought to determine the relationship between structural lead failure and clinical presentation by using the analysis of returned Riata products in Canada.
METHODS: The analyses of returned Riata products in Canada were obtained from St Jude Medical, Sylmar, CA. These data were correlated with the clinical presentation of patients just before lead removal from service.
RESULTS: As of May 1, 2013, there were 263 returned Riata leads in Canada. Of these, 43 (16.8%) were found to have insulation abrasion that was due to either lead-can or lead-other device interaction (70%) or inside-out abrasion (27.9%). The predilection of lead-to-can abrasion was seen in the Riata 7-F leads (84.2% vs 58.4%; P = .07), while inside-out abrasion was more common in the Riata 8-F leads (37.5% vs 15.8%; P = .12). Electrical abnormalities were frequent (20 of 31 [65.4%]) and most often due to electrical noise (45.2%), although inappropriate shocks were present (25.8%). Death occurred in 1 of 43 (2.3%) of those patients with an insulation defect in the lead-can abrasion group.
CONCLUSION: Lead-can abrasion is the most common form of insulation defect in the Riata group of leads under advisory. Management of this group of leads under advisory should not neglect the issue of lead-can abrasion, in addition to detection of cable externalization.
Copyright © 2015 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Device advisory; Implantable cardioverter-defibrillator; Lead failure

Mesh:

Year:  2014        PMID: 25485777     DOI: 10.1016/j.hrthm.2014.12.002

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  3 in total

1.  Cable externalization at the proximal portion of the superior vena cava coil in Riata implantable cardioverter defibrillator leads.

Authors:  Tomonori Katsuki; Hiroshi Furusho; Takashi Kusayama; Shinichiro Takashima; Takeshi Kato; Hisayoshi Murai; Soichiro Usui; Shuichi Kaneko; Masayuki Takamura
Journal:  J Cardiol Cases       Date:  2016-09-28

Review 2.  [ICD leads].

Authors:  Carsten W Israel; Mohamed Karim Sheta
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06

3.  Riata silicone defibrillation lead with normal electrical measures at routine ambulatory check: The role of high-voltage shock testing.

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Cardiol       Date:  2016-11-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.